Magnetic Particle Imaging: In vitro Signal Analysis and Lumen Quantification of 21 Endovascular Stents

PURPOSE: Endovascular stents are medical devices, which are implanted in stenosed blood vessels to ensure sufficient blood flow. Due to a high rate of in-stent re-stenoses, there is the need of a noninvasive imaging method for the early detection of stent occlusion. The evaluation of the stent lumen...

Descripción completa

Detalles Bibliográficos
Autores principales: Wegner, Franz, von Gladiss, Anselm, Haegele, Julian, Grzyska, Ulrike, Sieren, Malte Maria, Stahlberg, Erik, Oechtering, Thekla Helene, Lüdtke-Buzug, Kerstin, Barkhausen, Joerg, Buzug, Thorsten M, Friedrich, Thomas
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7810673/
https://www.ncbi.nlm.nih.gov/pubmed/33469281
http://dx.doi.org/10.2147/IJN.S284694
Descripción
Sumario:PURPOSE: Endovascular stents are medical devices, which are implanted in stenosed blood vessels to ensure sufficient blood flow. Due to a high rate of in-stent re-stenoses, there is the need of a noninvasive imaging method for the early detection of stent occlusion. The evaluation of the stent lumen with computed tomography (CT) and magnetic resonance imaging (MRI) is limited by material-induced artifacts. The purpose of this work is to investigate the potential of the tracer-based modality magnetic particle imaging (MPI) for stent lumen visualization and quantification. METHODS: In this in vitro study, 21 endovascular stents were investigated in a preclinical MPI scanner. Therefore, the stents were implanted in vessel phantoms. For the signal analysis, the phantoms were scanned without tracer material, and the signal-to-noise-ratio was analyzed. For the evaluation of potential artifacts and the lumen quantification, the phantoms were filled with diluted tracer agent. To calculate the stent lumen diameter a calibrated threshold value was applied. RESULTS: We can show that it is possible to visualize the lumen of a variety of endovascular stents without material induced artifacts, as the stents do not generate sufficient signals in MPI. The stent lumen quantification showed a direct correlation between the calculated and nominal diameter (r = 0.98). CONCLUSION: In contrast to MRI and CT, MPI is able to visualize and quantify stent lumina very accurately.