Determinants of Test Variability in Scotopic Microperimetry: Effects of Dark Adaptation and Test Indices

PURPOSE: To test the effect of different dark adaptation conditions and reliability indices on the variability of two color scotopic microperimetry. METHODS: We analyzed data from 22 consecutive visually healthy adults. Scotopic microperimetry was performed (Macular Integrity Assessment microperimet...

Descripción completa

Detalles Bibliográficos
Autores principales: Montesano, Giovanni, Naska, Timos K., Higgins, Bethany E., Wright, David M., Hogg, Ruth E., Crabb, David P.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Association for Research in Vision and Ophthalmology 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7814359/
https://www.ncbi.nlm.nih.gov/pubmed/33520422
http://dx.doi.org/10.1167/tvst.10.1.26
_version_ 1783638043356299264
author Montesano, Giovanni
Naska, Timos K.
Higgins, Bethany E.
Wright, David M.
Hogg, Ruth E.
Crabb, David P.
author_facet Montesano, Giovanni
Naska, Timos K.
Higgins, Bethany E.
Wright, David M.
Hogg, Ruth E.
Crabb, David P.
author_sort Montesano, Giovanni
collection PubMed
description PURPOSE: To test the effect of different dark adaptation conditions and reliability indices on the variability of two color scotopic microperimetry. METHODS: We analyzed data from 22 consecutive visually healthy adults. Scotopic microperimetry was performed (Macular Integrity Assessment microperimeter, CenterVue, Padua, Italy) with two wavelength stimuli, cyan (505 nm) and red (627 nm), after a dark adaptation time of 10, 20, or 30 minutes. All tests were repeated twice to measure test–retest variability with Bland–Altman plots. We also provide a method to more accurately quantify the false-positive (FP) responses based on response data (button pressing) from the device, similar to FP responses used in standard static perimetry. Data on fixation stability (95% bivariate contour ellipse area) and blind spot responses were also extracted. Their relationship with measured sensitivity (in decibels) and test–retest variability was quantified through linear mixed effect models. RESULTS: Dark adaptation had a significant effect on the sensitivity (dB) measured with the cyan stimulus (P < 0.001), but no effect on the red stimulus. Of the three metrics, the novel FP responses showed the best association with test–retest variability and was the only predictor consistently significant for all tests (P < 0.01). CONCLUSIONS: Dark adaptation protocols should be carefully standardized for scotopic testing, especially if a cyan stimulus is used. The proposed FP responses should be used to assess reliability of microperimetry examinations instead of other metrics. TRANSLATIONAL RELEVANCE: We developed a method to calculate a more accurate estimate of the FP responses using data available to all researchers, generalizable to all Macular Integrity Assessment microperimeter tests.
format Online
Article
Text
id pubmed-7814359
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher The Association for Research in Vision and Ophthalmology
record_format MEDLINE/PubMed
spelling pubmed-78143592021-01-29 Determinants of Test Variability in Scotopic Microperimetry: Effects of Dark Adaptation and Test Indices Montesano, Giovanni Naska, Timos K. Higgins, Bethany E. Wright, David M. Hogg, Ruth E. Crabb, David P. Transl Vis Sci Technol Article PURPOSE: To test the effect of different dark adaptation conditions and reliability indices on the variability of two color scotopic microperimetry. METHODS: We analyzed data from 22 consecutive visually healthy adults. Scotopic microperimetry was performed (Macular Integrity Assessment microperimeter, CenterVue, Padua, Italy) with two wavelength stimuli, cyan (505 nm) and red (627 nm), after a dark adaptation time of 10, 20, or 30 minutes. All tests were repeated twice to measure test–retest variability with Bland–Altman plots. We also provide a method to more accurately quantify the false-positive (FP) responses based on response data (button pressing) from the device, similar to FP responses used in standard static perimetry. Data on fixation stability (95% bivariate contour ellipse area) and blind spot responses were also extracted. Their relationship with measured sensitivity (in decibels) and test–retest variability was quantified through linear mixed effect models. RESULTS: Dark adaptation had a significant effect on the sensitivity (dB) measured with the cyan stimulus (P < 0.001), but no effect on the red stimulus. Of the three metrics, the novel FP responses showed the best association with test–retest variability and was the only predictor consistently significant for all tests (P < 0.01). CONCLUSIONS: Dark adaptation protocols should be carefully standardized for scotopic testing, especially if a cyan stimulus is used. The proposed FP responses should be used to assess reliability of microperimetry examinations instead of other metrics. TRANSLATIONAL RELEVANCE: We developed a method to calculate a more accurate estimate of the FP responses using data available to all researchers, generalizable to all Macular Integrity Assessment microperimeter tests. The Association for Research in Vision and Ophthalmology 2021-01-15 /pmc/articles/PMC7814359/ /pubmed/33520422 http://dx.doi.org/10.1167/tvst.10.1.26 Text en Copyright 2021 The Authors http://creativecommons.org/licenses/by/4.0/ This work is licensed under a Creative Commons Attribution 4.0 International License.
spellingShingle Article
Montesano, Giovanni
Naska, Timos K.
Higgins, Bethany E.
Wright, David M.
Hogg, Ruth E.
Crabb, David P.
Determinants of Test Variability in Scotopic Microperimetry: Effects of Dark Adaptation and Test Indices
title Determinants of Test Variability in Scotopic Microperimetry: Effects of Dark Adaptation and Test Indices
title_full Determinants of Test Variability in Scotopic Microperimetry: Effects of Dark Adaptation and Test Indices
title_fullStr Determinants of Test Variability in Scotopic Microperimetry: Effects of Dark Adaptation and Test Indices
title_full_unstemmed Determinants of Test Variability in Scotopic Microperimetry: Effects of Dark Adaptation and Test Indices
title_short Determinants of Test Variability in Scotopic Microperimetry: Effects of Dark Adaptation and Test Indices
title_sort determinants of test variability in scotopic microperimetry: effects of dark adaptation and test indices
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7814359/
https://www.ncbi.nlm.nih.gov/pubmed/33520422
http://dx.doi.org/10.1167/tvst.10.1.26
work_keys_str_mv AT montesanogiovanni determinantsoftestvariabilityinscotopicmicroperimetryeffectsofdarkadaptationandtestindices
AT naskatimosk determinantsoftestvariabilityinscotopicmicroperimetryeffectsofdarkadaptationandtestindices
AT higginsbethanye determinantsoftestvariabilityinscotopicmicroperimetryeffectsofdarkadaptationandtestindices
AT wrightdavidm determinantsoftestvariabilityinscotopicmicroperimetryeffectsofdarkadaptationandtestindices
AT hoggruthe determinantsoftestvariabilityinscotopicmicroperimetryeffectsofdarkadaptationandtestindices
AT crabbdavidp determinantsoftestvariabilityinscotopicmicroperimetryeffectsofdarkadaptationandtestindices