Simultaneous ventilation in the Covid-19 pandemic. A bench study
COVID-19 pandemic sets the healthcare system to a shortage of ventilators. We aimed at assessing tidal volume (V(T)) delivery and air recirculation during expiration when one ventilator is divided into 2 test-lungs. The study was performed in a research laboratory in a medical ICU of a University ho...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7815120/ https://www.ncbi.nlm.nih.gov/pubmed/33465155 http://dx.doi.org/10.1371/journal.pone.0245578 |
_version_ | 1783638173281157120 |
---|---|
author | Guérin, Claude Cour, Martin Stevic, Neven Degivry, Florian L’Her, Erwan Louis, Bruno Argaud, Laurent |
author_facet | Guérin, Claude Cour, Martin Stevic, Neven Degivry, Florian L’Her, Erwan Louis, Bruno Argaud, Laurent |
author_sort | Guérin, Claude |
collection | PubMed |
description | COVID-19 pandemic sets the healthcare system to a shortage of ventilators. We aimed at assessing tidal volume (V(T)) delivery and air recirculation during expiration when one ventilator is divided into 2 test-lungs. The study was performed in a research laboratory in a medical ICU of a University hospital. An ICU (V500) and a lower-level ventilator (Elisée 350) were attached to two test-lungs (QuickLung) through a dedicated flow-splitter. A 50 mL/cmH(2)O Compliance (C) and 5 cmH(2)O/L/s Resistance (R) were set in both A and B test-lungs (A C50R5 / B C50R5, step1), A C50-R20 / B C20-R20 (step 2), A C20-R20 / B C10-R20 (step 3), and A C50-R20 / B C20-R5 (step 4). Each ventilator was set in volume and pressure control mode to deliver 800mL V(T). We assessed V(T) from a pneumotachograph placed immediately before each lung, pendelluft air, and expiratory resistance (circuit and valve). Values are median (1(st)-3(rd) quartiles) and compared between ventilators by non-parametric tests. Between Elisée 350 and V500 in volume control V(T) in A/B test- lungs were 381/387 vs. 412/433 mL in step 1, 501/270 vs. 492/370 mL in step 2, 509/237 vs. 496/332 mL in step 3, and 496/281 vs. 480/329 mL in step 4. In pressure control the corresponding values were 373/336 vs. 430/414 mL, 416/185 vs. 322/234 mL, 193/108 vs. 176/ 92 mL and 422/201 vs. 481/329mL, respectively (P<0.001 between ventilators at each step for each volume). Pendelluft air volume ranged between 0.7 to 37.8 ml and negatively correlated with expiratory resistance in steps 2 and 3. The lower-level ventilator performed closely to the ICU ventilator. In the clinical setting, these findings suggest that, due to dependence of V(T) to C, pressure control should be preferred to maintain adequate V(T) at least in one patient when C and/or R changes abruptly and monitoring of V(T) should be done carefully. Increasing expiratory resistance should reduce pendelluft volume. |
format | Online Article Text |
id | pubmed-7815120 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-78151202021-01-27 Simultaneous ventilation in the Covid-19 pandemic. A bench study Guérin, Claude Cour, Martin Stevic, Neven Degivry, Florian L’Her, Erwan Louis, Bruno Argaud, Laurent PLoS One Research Article COVID-19 pandemic sets the healthcare system to a shortage of ventilators. We aimed at assessing tidal volume (V(T)) delivery and air recirculation during expiration when one ventilator is divided into 2 test-lungs. The study was performed in a research laboratory in a medical ICU of a University hospital. An ICU (V500) and a lower-level ventilator (Elisée 350) were attached to two test-lungs (QuickLung) through a dedicated flow-splitter. A 50 mL/cmH(2)O Compliance (C) and 5 cmH(2)O/L/s Resistance (R) were set in both A and B test-lungs (A C50R5 / B C50R5, step1), A C50-R20 / B C20-R20 (step 2), A C20-R20 / B C10-R20 (step 3), and A C50-R20 / B C20-R5 (step 4). Each ventilator was set in volume and pressure control mode to deliver 800mL V(T). We assessed V(T) from a pneumotachograph placed immediately before each lung, pendelluft air, and expiratory resistance (circuit and valve). Values are median (1(st)-3(rd) quartiles) and compared between ventilators by non-parametric tests. Between Elisée 350 and V500 in volume control V(T) in A/B test- lungs were 381/387 vs. 412/433 mL in step 1, 501/270 vs. 492/370 mL in step 2, 509/237 vs. 496/332 mL in step 3, and 496/281 vs. 480/329 mL in step 4. In pressure control the corresponding values were 373/336 vs. 430/414 mL, 416/185 vs. 322/234 mL, 193/108 vs. 176/ 92 mL and 422/201 vs. 481/329mL, respectively (P<0.001 between ventilators at each step for each volume). Pendelluft air volume ranged between 0.7 to 37.8 ml and negatively correlated with expiratory resistance in steps 2 and 3. The lower-level ventilator performed closely to the ICU ventilator. In the clinical setting, these findings suggest that, due to dependence of V(T) to C, pressure control should be preferred to maintain adequate V(T) at least in one patient when C and/or R changes abruptly and monitoring of V(T) should be done carefully. Increasing expiratory resistance should reduce pendelluft volume. Public Library of Science 2021-01-19 /pmc/articles/PMC7815120/ /pubmed/33465155 http://dx.doi.org/10.1371/journal.pone.0245578 Text en © 2021 Guérin et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Research Article Guérin, Claude Cour, Martin Stevic, Neven Degivry, Florian L’Her, Erwan Louis, Bruno Argaud, Laurent Simultaneous ventilation in the Covid-19 pandemic. A bench study |
title | Simultaneous ventilation in the Covid-19 pandemic. A bench study |
title_full | Simultaneous ventilation in the Covid-19 pandemic. A bench study |
title_fullStr | Simultaneous ventilation in the Covid-19 pandemic. A bench study |
title_full_unstemmed | Simultaneous ventilation in the Covid-19 pandemic. A bench study |
title_short | Simultaneous ventilation in the Covid-19 pandemic. A bench study |
title_sort | simultaneous ventilation in the covid-19 pandemic. a bench study |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7815120/ https://www.ncbi.nlm.nih.gov/pubmed/33465155 http://dx.doi.org/10.1371/journal.pone.0245578 |
work_keys_str_mv | AT guerinclaude simultaneousventilationinthecovid19pandemicabenchstudy AT courmartin simultaneousventilationinthecovid19pandemicabenchstudy AT stevicneven simultaneousventilationinthecovid19pandemicabenchstudy AT degivryflorian simultaneousventilationinthecovid19pandemicabenchstudy AT lhererwan simultaneousventilationinthecovid19pandemicabenchstudy AT louisbruno simultaneousventilationinthecovid19pandemicabenchstudy AT argaudlaurent simultaneousventilationinthecovid19pandemicabenchstudy |