Bending resistance and cyclic fatigue resistance of WaveOne Gold, Reciproc Blue, and HyFlex EDM instruments

BACKGROUND/PURPOSE: Several single-file systems manufactured using different heat treatment and operated by different kinematics have been released. This study compared the bending resistance and cyclic fatigue resistance of three NiTi files, and examined their phase-transformation behavior. MATERIA...

Descripción completa

Detalles Bibliográficos
Autores principales: Oh, Soram, Kum, Kee-Yeon, Kim, Hyun-Jung, Moon, Su-Young, Kim, Hyeon-Cheol, Chaniotis, Antonis, Perinpanayagam, Hiran, Pedullá, Eugenio, Chang, Seok Woo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Association for Dental Sciences of the Republic of China 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7816021/
https://www.ncbi.nlm.nih.gov/pubmed/33505619
http://dx.doi.org/10.1016/j.jds.2019.10.003
_version_ 1783638357079752704
author Oh, Soram
Kum, Kee-Yeon
Kim, Hyun-Jung
Moon, Su-Young
Kim, Hyeon-Cheol
Chaniotis, Antonis
Perinpanayagam, Hiran
Pedullá, Eugenio
Chang, Seok Woo
author_facet Oh, Soram
Kum, Kee-Yeon
Kim, Hyun-Jung
Moon, Su-Young
Kim, Hyeon-Cheol
Chaniotis, Antonis
Perinpanayagam, Hiran
Pedullá, Eugenio
Chang, Seok Woo
author_sort Oh, Soram
collection PubMed
description BACKGROUND/PURPOSE: Several single-file systems manufactured using different heat treatment and operated by different kinematics have been released. This study compared the bending resistance and cyclic fatigue resistance of three NiTi files, and examined their phase-transformation behavior. MATERIALS AND METHODS: WaveOne Gold Primary (WOG), Reciproc Blue R25 (RPB), and HyFlex EDM OneFile (HDM) were tested (N = 40/instrument). A vertical load was applied to a point 3 mm from the tip, and the stress was measured until a displacement of 3 mm. Tests were conducted at either room temperature (RT: 22 °C) or body temperature (BT: 37 °C) (n = 10). Cyclic fatigue resistance tests were performed in an artificial canal, which had a curvature angle of 40° and a 5-mm radius. Tests were conducted at either RT or BT (n = 10). Instruments were operated according to the manufacturers' instructions. Test results were analyzed using the Kruskal–Wallis and the Mann–Whitney tests. Additional three instruments of each brand were subjected to differential scanning calorimetry (DSC). RESULTS: At RT the bending resistance of three files were not significantly different. However, at BT the bending resistance of RPB was highest, followed by WOG, and HDM (P < 0.05). At RT, RPB demonstrated the longest fracture time, followed by HDM, and WOG (P < 0.05). At BT, HDM had the longest fracture time, followed by RPB, and WOG (P < 0.05). The WOG, RPB consisted of austenite in a considerable proportion, whereas HDM was mainly martensite state at BT. CONCLUSION: HDM presented superior flexibility and cyclic fatigue resistance at BT.
format Online
Article
Text
id pubmed-7816021
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Association for Dental Sciences of the Republic of China
record_format MEDLINE/PubMed
spelling pubmed-78160212021-01-26 Bending resistance and cyclic fatigue resistance of WaveOne Gold, Reciproc Blue, and HyFlex EDM instruments Oh, Soram Kum, Kee-Yeon Kim, Hyun-Jung Moon, Su-Young Kim, Hyeon-Cheol Chaniotis, Antonis Perinpanayagam, Hiran Pedullá, Eugenio Chang, Seok Woo J Dent Sci Original Article BACKGROUND/PURPOSE: Several single-file systems manufactured using different heat treatment and operated by different kinematics have been released. This study compared the bending resistance and cyclic fatigue resistance of three NiTi files, and examined their phase-transformation behavior. MATERIALS AND METHODS: WaveOne Gold Primary (WOG), Reciproc Blue R25 (RPB), and HyFlex EDM OneFile (HDM) were tested (N = 40/instrument). A vertical load was applied to a point 3 mm from the tip, and the stress was measured until a displacement of 3 mm. Tests were conducted at either room temperature (RT: 22 °C) or body temperature (BT: 37 °C) (n = 10). Cyclic fatigue resistance tests were performed in an artificial canal, which had a curvature angle of 40° and a 5-mm radius. Tests were conducted at either RT or BT (n = 10). Instruments were operated according to the manufacturers' instructions. Test results were analyzed using the Kruskal–Wallis and the Mann–Whitney tests. Additional three instruments of each brand were subjected to differential scanning calorimetry (DSC). RESULTS: At RT the bending resistance of three files were not significantly different. However, at BT the bending resistance of RPB was highest, followed by WOG, and HDM (P < 0.05). At RT, RPB demonstrated the longest fracture time, followed by HDM, and WOG (P < 0.05). At BT, HDM had the longest fracture time, followed by RPB, and WOG (P < 0.05). The WOG, RPB consisted of austenite in a considerable proportion, whereas HDM was mainly martensite state at BT. CONCLUSION: HDM presented superior flexibility and cyclic fatigue resistance at BT. Association for Dental Sciences of the Republic of China 2020-12 2019-12-30 /pmc/articles/PMC7816021/ /pubmed/33505619 http://dx.doi.org/10.1016/j.jds.2019.10.003 Text en © 2020 Association for Dental Sciences of the Republic of China. Publishing services by Elsevier B.V. http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Original Article
Oh, Soram
Kum, Kee-Yeon
Kim, Hyun-Jung
Moon, Su-Young
Kim, Hyeon-Cheol
Chaniotis, Antonis
Perinpanayagam, Hiran
Pedullá, Eugenio
Chang, Seok Woo
Bending resistance and cyclic fatigue resistance of WaveOne Gold, Reciproc Blue, and HyFlex EDM instruments
title Bending resistance and cyclic fatigue resistance of WaveOne Gold, Reciproc Blue, and HyFlex EDM instruments
title_full Bending resistance and cyclic fatigue resistance of WaveOne Gold, Reciproc Blue, and HyFlex EDM instruments
title_fullStr Bending resistance and cyclic fatigue resistance of WaveOne Gold, Reciproc Blue, and HyFlex EDM instruments
title_full_unstemmed Bending resistance and cyclic fatigue resistance of WaveOne Gold, Reciproc Blue, and HyFlex EDM instruments
title_short Bending resistance and cyclic fatigue resistance of WaveOne Gold, Reciproc Blue, and HyFlex EDM instruments
title_sort bending resistance and cyclic fatigue resistance of waveone gold, reciproc blue, and hyflex edm instruments
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7816021/
https://www.ncbi.nlm.nih.gov/pubmed/33505619
http://dx.doi.org/10.1016/j.jds.2019.10.003
work_keys_str_mv AT ohsoram bendingresistanceandcyclicfatigueresistanceofwaveonegoldreciprocblueandhyflexedminstruments
AT kumkeeyeon bendingresistanceandcyclicfatigueresistanceofwaveonegoldreciprocblueandhyflexedminstruments
AT kimhyunjung bendingresistanceandcyclicfatigueresistanceofwaveonegoldreciprocblueandhyflexedminstruments
AT moonsuyoung bendingresistanceandcyclicfatigueresistanceofwaveonegoldreciprocblueandhyflexedminstruments
AT kimhyeoncheol bendingresistanceandcyclicfatigueresistanceofwaveonegoldreciprocblueandhyflexedminstruments
AT chaniotisantonis bendingresistanceandcyclicfatigueresistanceofwaveonegoldreciprocblueandhyflexedminstruments
AT perinpanayagamhiran bendingresistanceandcyclicfatigueresistanceofwaveonegoldreciprocblueandhyflexedminstruments
AT pedullaeugenio bendingresistanceandcyclicfatigueresistanceofwaveonegoldreciprocblueandhyflexedminstruments
AT changseokwoo bendingresistanceandcyclicfatigueresistanceofwaveonegoldreciprocblueandhyflexedminstruments