Cargando…
Performance of two rapid point of care SARS-COV-2 antibody assays against laboratory-based automated chemiluminescent immunoassays for SARS-COV-2 IG-G, IG-M and total antibodies
INTRODUCTION: We evaluated two SARS-CoV-2 antibody point-of-care tests (POCTs) (Abbott Panbio COVID-19 IgG/IgM and Roche SARS-CoV-2 Rapid Antibody tests) and compared the results to their respective chemiluminescent immunoassays (CLIAs) (Abbott Architect IgM, Architect IgG, Roche Cobas total antibod...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7816621/ https://www.ncbi.nlm.nih.gov/pubmed/33501369 http://dx.doi.org/10.1016/j.plabm.2021.e00201 |
Sumario: | INTRODUCTION: We evaluated two SARS-CoV-2 antibody point-of-care tests (POCTs) (Abbott Panbio COVID-19 IgG/IgM and Roche SARS-CoV-2 Rapid Antibody tests) and compared the results to their respective chemiluminescent immunoassays (CLIAs) (Abbott Architect IgM, Architect IgG, Roche Cobas total antibody assays). METHOD: 200 pre-pandemic sera and 48 samples positive for various conditions (18 viral hepatitis, 18 dengue, 11 ANA and 1 dsDNA) were used as controls and to assess for cross-reactivity. Anonymised residual leftover sera positive for SARS-CoV-2 on RT-PCR were recruited as cases (n = 133). The sensitivity/specificity/cross-reactivity/positive predictive value (PPV)/negative predictive value (NPV) of the POCTs were assessed. Concordance between the POCTs and chemiluminescent immunoassays (CLIAs) were analysed. RESULTS: Abbott/Roche POCT specificity was 98.7%/100% (95% CI 96.5–99.8/98.5–100) and sensitivity was 97.2%/97.2% (95% CI 85.5–99.9/85.5–99.9) in cases ≥14 days post-first positive RT-PCR (POS), PPV was 68.7%/100% (95% CI 41.3–87.2/94.7–100.0), and NPV was 97.4%/97.6% (95% CI 97.0–97.8/97.2–98.0). In cases ≥14 days POS, concordance of Abbott/Roche POCT and CLIAs was 97.2%/100% (35/36 and 36/36 results). The sensitivity of individual IgM-band results on both POCTs did not increase >95% even after 14 days POS (Abbott 2.78%, Roche 44.4%). CONCLUSION: Both POCTs have good specificity, little cross-reactivity with other antibodies, and sensitivity >95% when used in subjects ≥14 days POS. Analysis of individual POCT IgG/IgM-bands did not provide any additional information. POCTs can substitute for CLIAs in cases ≥14 days POS. In low prevalence areas, POCTs would be especially useful when combined with antigen testing in an orthogonal format to increase the PPV of COVID-19 results. |
---|