Use of dried blood spot samples for SARS-CoV-2 antibody detection using the Roche Elecsys ® high throughput immunoassay
Dried blood spot samples (DBS) provide an alternative sample type to venous blood samples for antibody testing. DBS are used by NHS for diagnosing Hepatitis C and by Public Health England for large scale HIV and Hepatitis C serosurveillance; the applicability of DBS based approaches for SARS-CoV-2 a...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Published by Elsevier B.V.
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7817498/ https://www.ncbi.nlm.nih.gov/pubmed/33588354 http://dx.doi.org/10.1016/j.jcv.2021.104739 |
Sumario: | Dried blood spot samples (DBS) provide an alternative sample type to venous blood samples for antibody testing. DBS are used by NHS for diagnosing Hepatitis C and by Public Health England for large scale HIV and Hepatitis C serosurveillance; the applicability of DBS based approaches for SARS-CoV-2 antibody detection is uncertain. The study aimed to compare antibody detection in DBS eluates using the Roche Elecsys ® immunoassay with antibody detection in paired plasma samples, using the same assay. The study was in one Police and one Fire & Rescue facility in England; it comprised of 195 participants within a larger sample COVID-19 serodiagnostics study of keyworkers, EDSAB-HOME. Outcome measures were sensitivity and specificity of DBS (the index test) relative to plasma (the reference test), at an experimental cut-off; quality of DBS sample collected; estimates of relative sensitivity of DBS vs. plasma immunoassay in a larger population. 18/195 (9.2%) participants tested positive using plasma samples. DBS sample quality varied markedly by phlebotomist, and low sample volume significantly reduced immunoassay signals. Using an experimental cut-off, sensitivity and specificity of DBS were 89.0% (95% CI 67.2, 96.9%) and 100.0% (95% CI 97.9, 100%) respectively compared with using plasma. The limit of detection for DBS is about 30 times higher than for plasma. DBS use for SARS-CoV-2 serology, though feasible, is insensitive relative to immunoassays on plasma. Sample quality impacts on assay performance. Alternatives, including the collection of capillary blood samples, should be considered for screening programs. |
---|