Cardiac Output Evaluation on Septic Shock Patients: Comparison between Calibrated and Uncalibrated Devices during Vasopressor Therapy

There are no reliable, non-invasive methods to accurately measure cardiac output (CO) in septic patients. MostCare (Vytech Health™, Vygon, Padova, Italy), is a beat-to-beat, self calibrated method for CO measurement based on continuous analysis of reflected arterial pressure waveforms. We enrolled 4...

Descripción completa

Detalles Bibliográficos
Autores principales: Persona, Paolo, Valeri, Ilaria, Saraceni, Elisabetta, De Cassai, Alessandro, Calabrese, Fabrizia, Navalesi, Paolo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7826755/
https://www.ncbi.nlm.nih.gov/pubmed/33435270
http://dx.doi.org/10.3390/jcm10020213
_version_ 1783640596079968256
author Persona, Paolo
Valeri, Ilaria
Saraceni, Elisabetta
De Cassai, Alessandro
Calabrese, Fabrizia
Navalesi, Paolo
author_facet Persona, Paolo
Valeri, Ilaria
Saraceni, Elisabetta
De Cassai, Alessandro
Calabrese, Fabrizia
Navalesi, Paolo
author_sort Persona, Paolo
collection PubMed
description There are no reliable, non-invasive methods to accurately measure cardiac output (CO) in septic patients. MostCare (Vytech Health™, Vygon, Padova, Italy), is a beat-to-beat, self calibrated method for CO measurement based on continuous analysis of reflected arterial pressure waveforms. We enrolled 40 patients that were suffering from septic shock and requiring norepinephrine infusion to target blood pressure in order to to evaluate the level of agreement between a calibrated transpulmonary thermodilution device (PiCCO System, Pulsion Medical Systems, Feldkirchen, Germany) and the MostCare system in detecting and tracking changes in CO measurements related to norepinephrine reduction in septic shock patients,. PiCCO was connected to a 5 Fr femoral artery catheter and to a central venous catheter. System calibration was performed with 15 mL of cold saline injection over about 3 s. The MostCare device was connected to the artery catheter to analyze the arterial waveform. Before reducing norepinephrine infusion, the PiCCO system was calibrated, the MostCare waveform was optimized, and the values of the complete hemodynamic profile were recorded (T1). Norepinephrine infusion was then reduced by 0.03 mcg/Kg/min. After 30 min, a new calibration of PiCCO system and a new record on both monitors were performed (T2). Static measurements agreements were assessed using the Bland-Altman test, while trending ability was investigated using polar plot analysis. If volume expansion occurred, then related data were separately analyzed. At T1 mean the CO was 5.38 (SD 0.60) L/min, the mean difference was 0.176 L/min, the limits of agreement (LoA) was +1.39 and −1.04 L/min, and the percentage error (PE) was 22.6%; at T2 the mean CO was 5.44 (SD 0.73) L/min, the mean difference was 0.053 L/min, the LoA was +1.51 and −1.40, and the PE was 27%. After considering the volume expansion between T1 and T2, the mean CO at T1 was 5.39 L/min (SD 0.47), the LoA was +1.09 and −0.78 L/min, and the percentage error (PE) was 17%; at T2 the mean CO was 5.35 L/min (SD 0.81), the LoA was +1.73 and −1.52 L/min, and the PE was 30%. The polar plot diagram seems to confirm the trending ability of MostCare system versus the reference method. In septic patients, when the arterial waveform is accurate, MostCare and PiCCO transpulmonary thermodilution exhibit good agreement even after the reduction of norepinephrine and changes in vascular tone or volume expansion. MostCare could be a rapid to set, reliable, and useful tool to monitor hemodynamic variations in septic patients in emergency contexts where thermodilution methods or other advanced systems are not easily available.
format Online
Article
Text
id pubmed-7826755
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-78267552021-01-25 Cardiac Output Evaluation on Septic Shock Patients: Comparison between Calibrated and Uncalibrated Devices during Vasopressor Therapy Persona, Paolo Valeri, Ilaria Saraceni, Elisabetta De Cassai, Alessandro Calabrese, Fabrizia Navalesi, Paolo J Clin Med Article There are no reliable, non-invasive methods to accurately measure cardiac output (CO) in septic patients. MostCare (Vytech Health™, Vygon, Padova, Italy), is a beat-to-beat, self calibrated method for CO measurement based on continuous analysis of reflected arterial pressure waveforms. We enrolled 40 patients that were suffering from septic shock and requiring norepinephrine infusion to target blood pressure in order to to evaluate the level of agreement between a calibrated transpulmonary thermodilution device (PiCCO System, Pulsion Medical Systems, Feldkirchen, Germany) and the MostCare system in detecting and tracking changes in CO measurements related to norepinephrine reduction in septic shock patients,. PiCCO was connected to a 5 Fr femoral artery catheter and to a central venous catheter. System calibration was performed with 15 mL of cold saline injection over about 3 s. The MostCare device was connected to the artery catheter to analyze the arterial waveform. Before reducing norepinephrine infusion, the PiCCO system was calibrated, the MostCare waveform was optimized, and the values of the complete hemodynamic profile were recorded (T1). Norepinephrine infusion was then reduced by 0.03 mcg/Kg/min. After 30 min, a new calibration of PiCCO system and a new record on both monitors were performed (T2). Static measurements agreements were assessed using the Bland-Altman test, while trending ability was investigated using polar plot analysis. If volume expansion occurred, then related data were separately analyzed. At T1 mean the CO was 5.38 (SD 0.60) L/min, the mean difference was 0.176 L/min, the limits of agreement (LoA) was +1.39 and −1.04 L/min, and the percentage error (PE) was 22.6%; at T2 the mean CO was 5.44 (SD 0.73) L/min, the mean difference was 0.053 L/min, the LoA was +1.51 and −1.40, and the PE was 27%. After considering the volume expansion between T1 and T2, the mean CO at T1 was 5.39 L/min (SD 0.47), the LoA was +1.09 and −0.78 L/min, and the percentage error (PE) was 17%; at T2 the mean CO was 5.35 L/min (SD 0.81), the LoA was +1.73 and −1.52 L/min, and the PE was 30%. The polar plot diagram seems to confirm the trending ability of MostCare system versus the reference method. In septic patients, when the arterial waveform is accurate, MostCare and PiCCO transpulmonary thermodilution exhibit good agreement even after the reduction of norepinephrine and changes in vascular tone or volume expansion. MostCare could be a rapid to set, reliable, and useful tool to monitor hemodynamic variations in septic patients in emergency contexts where thermodilution methods or other advanced systems are not easily available. MDPI 2021-01-09 /pmc/articles/PMC7826755/ /pubmed/33435270 http://dx.doi.org/10.3390/jcm10020213 Text en © 2021 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Persona, Paolo
Valeri, Ilaria
Saraceni, Elisabetta
De Cassai, Alessandro
Calabrese, Fabrizia
Navalesi, Paolo
Cardiac Output Evaluation on Septic Shock Patients: Comparison between Calibrated and Uncalibrated Devices during Vasopressor Therapy
title Cardiac Output Evaluation on Septic Shock Patients: Comparison between Calibrated and Uncalibrated Devices during Vasopressor Therapy
title_full Cardiac Output Evaluation on Septic Shock Patients: Comparison between Calibrated and Uncalibrated Devices during Vasopressor Therapy
title_fullStr Cardiac Output Evaluation on Septic Shock Patients: Comparison between Calibrated and Uncalibrated Devices during Vasopressor Therapy
title_full_unstemmed Cardiac Output Evaluation on Septic Shock Patients: Comparison between Calibrated and Uncalibrated Devices during Vasopressor Therapy
title_short Cardiac Output Evaluation on Septic Shock Patients: Comparison between Calibrated and Uncalibrated Devices during Vasopressor Therapy
title_sort cardiac output evaluation on septic shock patients: comparison between calibrated and uncalibrated devices during vasopressor therapy
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7826755/
https://www.ncbi.nlm.nih.gov/pubmed/33435270
http://dx.doi.org/10.3390/jcm10020213
work_keys_str_mv AT personapaolo cardiacoutputevaluationonsepticshockpatientscomparisonbetweencalibratedanduncalibrateddevicesduringvasopressortherapy
AT valeriilaria cardiacoutputevaluationonsepticshockpatientscomparisonbetweencalibratedanduncalibrateddevicesduringvasopressortherapy
AT saracenielisabetta cardiacoutputevaluationonsepticshockpatientscomparisonbetweencalibratedanduncalibrateddevicesduringvasopressortherapy
AT decassaialessandro cardiacoutputevaluationonsepticshockpatientscomparisonbetweencalibratedanduncalibrateddevicesduringvasopressortherapy
AT calabresefabrizia cardiacoutputevaluationonsepticshockpatientscomparisonbetweencalibratedanduncalibrateddevicesduringvasopressortherapy
AT navalesipaolo cardiacoutputevaluationonsepticshockpatientscomparisonbetweencalibratedanduncalibrateddevicesduringvasopressortherapy