Whole-body x-ray dark-field radiography of a human cadaver

BACKGROUND: Grating-based x-ray dark-field and phase-contrast imaging allow extracting information about refraction and small-angle scatter, beyond conventional attenuation. A step towards clinical translation has recently been achieved, allowing further investigation on humans. METHODS: After the e...

Descripción completa

Detalles Bibliográficos
Autores principales: Andrejewski, Jana, De Marco, Fabio, Willer, Konstantin, Noichl, Wolfgang, Gustschin, Alex, Koehler, Thomas, Meyer, Pascal, Kriner, Fabian, Fischer, Florian, Braun, Christian, Fingerle, Alexander A., Herzen, Julia, Pfeiffer, Franz, Pfeiffer, Daniela
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7835263/
https://www.ncbi.nlm.nih.gov/pubmed/33495889
http://dx.doi.org/10.1186/s41747-020-00201-1
Descripción
Sumario:BACKGROUND: Grating-based x-ray dark-field and phase-contrast imaging allow extracting information about refraction and small-angle scatter, beyond conventional attenuation. A step towards clinical translation has recently been achieved, allowing further investigation on humans. METHODS: After the ethics committee approval, we scanned the full body of a human cadaver in anterior-posterior orientation. Six measurements were stitched together to form the whole-body image. All radiographs were taken at a three-grating large-object x-ray dark-field scanner, each lasting about 40 s. Signal intensities of different anatomical regions were assessed. The magnitude of visibility reduction caused by beam hardening instead of small-angle scatter was analysed using different phantom materials. Maximal effective dose was 0.3 mSv for the abdomen. RESULTS: Combined attenuation and dark-field radiography are technically possible throughout a whole human body. High signal levels were found in several bony structures, foreign materials, and the lung. Signal levels were 0.25 ± 0.13 (mean ± standard deviation) for the lungs, 0.08 ± 0.06 for the bones, 0.023 ± 0.019 for soft tissue, and 0.30 ± 0.02 for an antibiotic bead chain. We found that phantom materials, which do not produce small-angle scatter, can generate a strong visibility reduction signal. CONCLUSION: We acquired a whole-body x-ray dark-field radiograph of a human body in few minutes with an effective dose in a clinical acceptable range. Our findings suggest that the observed visibility reduction in the bone and metal is dominated by beam hardening and that the true dark-field signal in the lung is therefore much higher than that of the bone.