Multicenter comparison of the Cobas 6800 system with the RealStar RT-PCR kit for the detection of SARS-CoV-2
BACKGROUND: RT-PCR testing is crucial in the diagnostic of SARS-CoV-2 infection. The use of reliable and comparable PCR assays is a cornerstone to allow use of different PCR assays depending on the local equipment. In this work, we provide a comparison of the Cobas® (Roche) and the RealStar® assay (...
Autores principales: | , , , , , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier B.V.
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7836314/ https://www.ncbi.nlm.nih.gov/pubmed/32769025 http://dx.doi.org/10.1016/j.jcv.2020.104573 |
_version_ | 1783642721940930560 |
---|---|
author | Wirden, Marc Feghoul, Linda Bertine, Mélanie Nere, Marie-Laure Le Hingrat, Quentin Abdi, Basma Boutolleau, David Ferre, Valentine Marie Jary, Aude Delaugerre, Constance Marcelin, Anne-Genevieve Descamps, Diane Legoff, Jérôme Visseaux, Benoit Chaix, Marie-Laure |
author_facet | Wirden, Marc Feghoul, Linda Bertine, Mélanie Nere, Marie-Laure Le Hingrat, Quentin Abdi, Basma Boutolleau, David Ferre, Valentine Marie Jary, Aude Delaugerre, Constance Marcelin, Anne-Genevieve Descamps, Diane Legoff, Jérôme Visseaux, Benoit Chaix, Marie-Laure |
author_sort | Wirden, Marc |
collection | PubMed |
description | BACKGROUND: RT-PCR testing is crucial in the diagnostic of SARS-CoV-2 infection. The use of reliable and comparable PCR assays is a cornerstone to allow use of different PCR assays depending on the local equipment. In this work, we provide a comparison of the Cobas® (Roche) and the RealStar® assay (Altona). METHODS: Assessment of the two assays was performed prospectively in three reference Parisians hospitals, using 170 clinical samples. They were tested with the Cobas® assay, selected to obtain a distribution of cycle threshold (Ct) as large as possible, and tested with the RealStar assay with three largely available extraction platforms: QIAsymphony (Qiagen), MagNAPure (Roche) and NucliSENS-easyMag (BioMérieux). RESULTS: Overall, the agreement (positive for at least one gene) was 76 %. This rate differed considerably depending on the Cobas Ct values for gene E: below 35 (n = 91), the concordance was 99 %. Regarding the positive Ct values, linear regression analysis showed a coefficient of determination (R(2)) of 0.88 and the Deming regression line revealed a strong correlation with a slope of 1.023 and an intercept of -3.9. Bland-Altman analysis showed that the mean difference (Cobas® minus RealStar®) was + 3.3 Ct, with a SD of + 2.3 Ct. CONCLUSIONS: In this comparison, both RealStar® and Cobas® assays provided comparable qualitative results and a high correlation when both tests were positive. Discrepancies exist after 35 Ct and varied depending on the extraction system used for the RealStar® assay, probably due to a low viral load close to the detection limit of both assays. |
format | Online Article Text |
id | pubmed-7836314 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Elsevier B.V. |
record_format | MEDLINE/PubMed |
spelling | pubmed-78363142021-01-26 Multicenter comparison of the Cobas 6800 system with the RealStar RT-PCR kit for the detection of SARS-CoV-2 Wirden, Marc Feghoul, Linda Bertine, Mélanie Nere, Marie-Laure Le Hingrat, Quentin Abdi, Basma Boutolleau, David Ferre, Valentine Marie Jary, Aude Delaugerre, Constance Marcelin, Anne-Genevieve Descamps, Diane Legoff, Jérôme Visseaux, Benoit Chaix, Marie-Laure J Clin Virol Article BACKGROUND: RT-PCR testing is crucial in the diagnostic of SARS-CoV-2 infection. The use of reliable and comparable PCR assays is a cornerstone to allow use of different PCR assays depending on the local equipment. In this work, we provide a comparison of the Cobas® (Roche) and the RealStar® assay (Altona). METHODS: Assessment of the two assays was performed prospectively in three reference Parisians hospitals, using 170 clinical samples. They were tested with the Cobas® assay, selected to obtain a distribution of cycle threshold (Ct) as large as possible, and tested with the RealStar assay with three largely available extraction platforms: QIAsymphony (Qiagen), MagNAPure (Roche) and NucliSENS-easyMag (BioMérieux). RESULTS: Overall, the agreement (positive for at least one gene) was 76 %. This rate differed considerably depending on the Cobas Ct values for gene E: below 35 (n = 91), the concordance was 99 %. Regarding the positive Ct values, linear regression analysis showed a coefficient of determination (R(2)) of 0.88 and the Deming regression line revealed a strong correlation with a slope of 1.023 and an intercept of -3.9. Bland-Altman analysis showed that the mean difference (Cobas® minus RealStar®) was + 3.3 Ct, with a SD of + 2.3 Ct. CONCLUSIONS: In this comparison, both RealStar® and Cobas® assays provided comparable qualitative results and a high correlation when both tests were positive. Discrepancies exist after 35 Ct and varied depending on the extraction system used for the RealStar® assay, probably due to a low viral load close to the detection limit of both assays. Elsevier B.V. 2020-09 2020-07-31 /pmc/articles/PMC7836314/ /pubmed/32769025 http://dx.doi.org/10.1016/j.jcv.2020.104573 Text en © 2020 Elsevier B.V. All rights reserved. Since January 2020 Elsevier has created a COVID-19 resource centre with free information in English and Mandarin on the novel coronavirus COVID-19. The COVID-19 resource centre is hosted on Elsevier Connect, the company's public news and information website. Elsevier hereby grants permission to make all its COVID-19-related research that is available on the COVID-19 resource centre - including this research content - immediately available in PubMed Central and other publicly funded repositories, such as the WHO COVID database with rights for unrestricted research re-use and analyses in any form or by any means with acknowledgement of the original source. These permissions are granted for free by Elsevier for as long as the COVID-19 resource centre remains active. |
spellingShingle | Article Wirden, Marc Feghoul, Linda Bertine, Mélanie Nere, Marie-Laure Le Hingrat, Quentin Abdi, Basma Boutolleau, David Ferre, Valentine Marie Jary, Aude Delaugerre, Constance Marcelin, Anne-Genevieve Descamps, Diane Legoff, Jérôme Visseaux, Benoit Chaix, Marie-Laure Multicenter comparison of the Cobas 6800 system with the RealStar RT-PCR kit for the detection of SARS-CoV-2 |
title | Multicenter comparison of the Cobas 6800 system with the RealStar RT-PCR kit for the detection of SARS-CoV-2 |
title_full | Multicenter comparison of the Cobas 6800 system with the RealStar RT-PCR kit for the detection of SARS-CoV-2 |
title_fullStr | Multicenter comparison of the Cobas 6800 system with the RealStar RT-PCR kit for the detection of SARS-CoV-2 |
title_full_unstemmed | Multicenter comparison of the Cobas 6800 system with the RealStar RT-PCR kit for the detection of SARS-CoV-2 |
title_short | Multicenter comparison of the Cobas 6800 system with the RealStar RT-PCR kit for the detection of SARS-CoV-2 |
title_sort | multicenter comparison of the cobas 6800 system with the realstar rt-pcr kit for the detection of sars-cov-2 |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7836314/ https://www.ncbi.nlm.nih.gov/pubmed/32769025 http://dx.doi.org/10.1016/j.jcv.2020.104573 |
work_keys_str_mv | AT wirdenmarc multicentercomparisonofthecobas6800systemwiththerealstarrtpcrkitforthedetectionofsarscov2 AT feghoullinda multicentercomparisonofthecobas6800systemwiththerealstarrtpcrkitforthedetectionofsarscov2 AT bertinemelanie multicentercomparisonofthecobas6800systemwiththerealstarrtpcrkitforthedetectionofsarscov2 AT neremarielaure multicentercomparisonofthecobas6800systemwiththerealstarrtpcrkitforthedetectionofsarscov2 AT lehingratquentin multicentercomparisonofthecobas6800systemwiththerealstarrtpcrkitforthedetectionofsarscov2 AT abdibasma multicentercomparisonofthecobas6800systemwiththerealstarrtpcrkitforthedetectionofsarscov2 AT boutolleaudavid multicentercomparisonofthecobas6800systemwiththerealstarrtpcrkitforthedetectionofsarscov2 AT ferrevalentinemarie multicentercomparisonofthecobas6800systemwiththerealstarrtpcrkitforthedetectionofsarscov2 AT jaryaude multicentercomparisonofthecobas6800systemwiththerealstarrtpcrkitforthedetectionofsarscov2 AT delaugerreconstance multicentercomparisonofthecobas6800systemwiththerealstarrtpcrkitforthedetectionofsarscov2 AT marcelinannegenevieve multicentercomparisonofthecobas6800systemwiththerealstarrtpcrkitforthedetectionofsarscov2 AT descampsdiane multicentercomparisonofthecobas6800systemwiththerealstarrtpcrkitforthedetectionofsarscov2 AT legoffjerome multicentercomparisonofthecobas6800systemwiththerealstarrtpcrkitforthedetectionofsarscov2 AT visseauxbenoit multicentercomparisonofthecobas6800systemwiththerealstarrtpcrkitforthedetectionofsarscov2 AT chaixmarielaure multicentercomparisonofthecobas6800systemwiththerealstarrtpcrkitforthedetectionofsarscov2 |