Cargando…
Comparative in-vitro Study of the Trachospray, a New Device for Topical Anaesthesia of the Upper Airway
BACKGROUND: Obtaining complete topical anaesthesia of the airway remains a clinical challenge. Particle size is one of the most important variables for the dose deposited and the distribution of aerosols in the airways. The mass median aerodynamic diameter of the particles should be in the range of...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Dove
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7837536/ https://www.ncbi.nlm.nih.gov/pubmed/33519246 http://dx.doi.org/10.2147/MDER.S292529 |
_version_ | 1783642973066493952 |
---|---|
author | van Geffen, Geert-Jan Markerink, Hielke van Barneveld, Marc Verhoeven, Frank Scheffer, Gert Jan Bruhn, Jörgen |
author_facet | van Geffen, Geert-Jan Markerink, Hielke van Barneveld, Marc Verhoeven, Frank Scheffer, Gert Jan Bruhn, Jörgen |
author_sort | van Geffen, Geert-Jan |
collection | PubMed |
description | BACKGROUND: Obtaining complete topical anaesthesia of the airway remains a clinical challenge. Particle size is one of the most important variables for the dose deposited and the distribution of aerosols in the airways. The mass median aerodynamic diameter of the particles should be in the range of 5–20 µm. We developed the “Trachospray” as a soft mist spray device for local anaesthetics. This in-vitro comparative test was designed to compare the performance of the new Trachospray device with two existing medical devices. The performance was determined by comparing the spray deposition patterns in the mouth, throat, trachea and lungs. METHODS: The human airway was simulated with an artificial idealized mouth and throat model, connected to a Next Generation Impactor. Four measurements were taken for each device (Trachospray, jet nebulizer and a spray pump) with 5.85% NaCl. A fifth measurement was carried out with 0.5% fluorescein solution for a visual inspection of the deposition patterns. The mass median aerodynamic diameter and geometric standard deviation of the droplets were measured. RESULTS: The Trachospray produced an even coverage in the mouth, hypopharynx and vocal cords, with only a small lung fraction. The jet nebulizer produced a much thinner layer coverage of the tongue and surface around the vocal chords with a high lung deposition. The spray pump produced big droplets which deposited mainly at the hypopharynx. CONCLUSION: The Trachospray device deposits local anaesthetics in the targeted areas for topical anaesthesia of the airway and has promising characteristics for providing effective airway anaesthesia. |
format | Online Article Text |
id | pubmed-7837536 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Dove |
record_format | MEDLINE/PubMed |
spelling | pubmed-78375362021-01-28 Comparative in-vitro Study of the Trachospray, a New Device for Topical Anaesthesia of the Upper Airway van Geffen, Geert-Jan Markerink, Hielke van Barneveld, Marc Verhoeven, Frank Scheffer, Gert Jan Bruhn, Jörgen Med Devices (Auckl) Original Research BACKGROUND: Obtaining complete topical anaesthesia of the airway remains a clinical challenge. Particle size is one of the most important variables for the dose deposited and the distribution of aerosols in the airways. The mass median aerodynamic diameter of the particles should be in the range of 5–20 µm. We developed the “Trachospray” as a soft mist spray device for local anaesthetics. This in-vitro comparative test was designed to compare the performance of the new Trachospray device with two existing medical devices. The performance was determined by comparing the spray deposition patterns in the mouth, throat, trachea and lungs. METHODS: The human airway was simulated with an artificial idealized mouth and throat model, connected to a Next Generation Impactor. Four measurements were taken for each device (Trachospray, jet nebulizer and a spray pump) with 5.85% NaCl. A fifth measurement was carried out with 0.5% fluorescein solution for a visual inspection of the deposition patterns. The mass median aerodynamic diameter and geometric standard deviation of the droplets were measured. RESULTS: The Trachospray produced an even coverage in the mouth, hypopharynx and vocal cords, with only a small lung fraction. The jet nebulizer produced a much thinner layer coverage of the tongue and surface around the vocal chords with a high lung deposition. The spray pump produced big droplets which deposited mainly at the hypopharynx. CONCLUSION: The Trachospray device deposits local anaesthetics in the targeted areas for topical anaesthesia of the airway and has promising characteristics for providing effective airway anaesthesia. Dove 2021-01-22 /pmc/articles/PMC7837536/ /pubmed/33519246 http://dx.doi.org/10.2147/MDER.S292529 Text en © 2021 van Geffen et al. http://creativecommons.org/licenses/by-nc/3.0/ This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms (https://www.dovepress.com/terms.php). |
spellingShingle | Original Research van Geffen, Geert-Jan Markerink, Hielke van Barneveld, Marc Verhoeven, Frank Scheffer, Gert Jan Bruhn, Jörgen Comparative in-vitro Study of the Trachospray, a New Device for Topical Anaesthesia of the Upper Airway |
title | Comparative in-vitro Study of the Trachospray, a New Device for Topical Anaesthesia of the Upper Airway |
title_full | Comparative in-vitro Study of the Trachospray, a New Device for Topical Anaesthesia of the Upper Airway |
title_fullStr | Comparative in-vitro Study of the Trachospray, a New Device for Topical Anaesthesia of the Upper Airway |
title_full_unstemmed | Comparative in-vitro Study of the Trachospray, a New Device for Topical Anaesthesia of the Upper Airway |
title_short | Comparative in-vitro Study of the Trachospray, a New Device for Topical Anaesthesia of the Upper Airway |
title_sort | comparative in-vitro study of the trachospray, a new device for topical anaesthesia of the upper airway |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7837536/ https://www.ncbi.nlm.nih.gov/pubmed/33519246 http://dx.doi.org/10.2147/MDER.S292529 |
work_keys_str_mv | AT vangeffengeertjan comparativeinvitrostudyofthetrachosprayanewdevicefortopicalanaesthesiaoftheupperairway AT markerinkhielke comparativeinvitrostudyofthetrachosprayanewdevicefortopicalanaesthesiaoftheupperairway AT vanbarneveldmarc comparativeinvitrostudyofthetrachosprayanewdevicefortopicalanaesthesiaoftheupperairway AT verhoevenfrank comparativeinvitrostudyofthetrachosprayanewdevicefortopicalanaesthesiaoftheupperairway AT scheffergertjan comparativeinvitrostudyofthetrachosprayanewdevicefortopicalanaesthesiaoftheupperairway AT bruhnjorgen comparativeinvitrostudyofthetrachosprayanewdevicefortopicalanaesthesiaoftheupperairway |