Comparative in-vitro Study of the Trachospray, a New Device for Topical Anaesthesia of the Upper Airway

BACKGROUND: Obtaining complete topical anaesthesia of the airway remains a clinical challenge. Particle size is one of the most important variables for the dose deposited and the distribution of aerosols in the airways. The mass median aerodynamic diameter of the particles should be in the range of...

Descripción completa

Detalles Bibliográficos
Autores principales: van Geffen, Geert-Jan, Markerink, Hielke, van Barneveld, Marc, Verhoeven, Frank, Scheffer, Gert Jan, Bruhn, Jörgen
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7837536/
https://www.ncbi.nlm.nih.gov/pubmed/33519246
http://dx.doi.org/10.2147/MDER.S292529
_version_ 1783642973066493952
author van Geffen, Geert-Jan
Markerink, Hielke
van Barneveld, Marc
Verhoeven, Frank
Scheffer, Gert Jan
Bruhn, Jörgen
author_facet van Geffen, Geert-Jan
Markerink, Hielke
van Barneveld, Marc
Verhoeven, Frank
Scheffer, Gert Jan
Bruhn, Jörgen
author_sort van Geffen, Geert-Jan
collection PubMed
description BACKGROUND: Obtaining complete topical anaesthesia of the airway remains a clinical challenge. Particle size is one of the most important variables for the dose deposited and the distribution of aerosols in the airways. The mass median aerodynamic diameter of the particles should be in the range of 5–20 µm. We developed the “Trachospray” as a soft mist spray device for local anaesthetics. This in-vitro comparative test was designed to compare the performance of the new Trachospray device with two existing medical devices. The performance was determined by comparing the spray deposition patterns in the mouth, throat, trachea and lungs. METHODS: The human airway was simulated with an artificial idealized mouth and throat model, connected to a Next Generation Impactor. Four measurements were taken for each device (Trachospray, jet nebulizer and a spray pump) with 5.85% NaCl. A fifth measurement was carried out with 0.5% fluorescein solution for a visual inspection of the deposition patterns. The mass median aerodynamic diameter and geometric standard deviation of the droplets were measured. RESULTS: The Trachospray produced an even coverage in the mouth, hypopharynx and vocal cords, with only a small lung fraction. The jet nebulizer produced a much thinner layer coverage of the tongue and surface around the vocal chords with a high lung deposition. The spray pump produced big droplets which deposited mainly at the hypopharynx. CONCLUSION: The Trachospray device deposits local anaesthetics in the targeted areas for topical anaesthesia of the airway and has promising characteristics for providing effective airway anaesthesia.
format Online
Article
Text
id pubmed-7837536
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Dove
record_format MEDLINE/PubMed
spelling pubmed-78375362021-01-28 Comparative in-vitro Study of the Trachospray, a New Device for Topical Anaesthesia of the Upper Airway van Geffen, Geert-Jan Markerink, Hielke van Barneveld, Marc Verhoeven, Frank Scheffer, Gert Jan Bruhn, Jörgen Med Devices (Auckl) Original Research BACKGROUND: Obtaining complete topical anaesthesia of the airway remains a clinical challenge. Particle size is one of the most important variables for the dose deposited and the distribution of aerosols in the airways. The mass median aerodynamic diameter of the particles should be in the range of 5–20 µm. We developed the “Trachospray” as a soft mist spray device for local anaesthetics. This in-vitro comparative test was designed to compare the performance of the new Trachospray device with two existing medical devices. The performance was determined by comparing the spray deposition patterns in the mouth, throat, trachea and lungs. METHODS: The human airway was simulated with an artificial idealized mouth and throat model, connected to a Next Generation Impactor. Four measurements were taken for each device (Trachospray, jet nebulizer and a spray pump) with 5.85% NaCl. A fifth measurement was carried out with 0.5% fluorescein solution for a visual inspection of the deposition patterns. The mass median aerodynamic diameter and geometric standard deviation of the droplets were measured. RESULTS: The Trachospray produced an even coverage in the mouth, hypopharynx and vocal cords, with only a small lung fraction. The jet nebulizer produced a much thinner layer coverage of the tongue and surface around the vocal chords with a high lung deposition. The spray pump produced big droplets which deposited mainly at the hypopharynx. CONCLUSION: The Trachospray device deposits local anaesthetics in the targeted areas for topical anaesthesia of the airway and has promising characteristics for providing effective airway anaesthesia. Dove 2021-01-22 /pmc/articles/PMC7837536/ /pubmed/33519246 http://dx.doi.org/10.2147/MDER.S292529 Text en © 2021 van Geffen et al. http://creativecommons.org/licenses/by-nc/3.0/ This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms (https://www.dovepress.com/terms.php).
spellingShingle Original Research
van Geffen, Geert-Jan
Markerink, Hielke
van Barneveld, Marc
Verhoeven, Frank
Scheffer, Gert Jan
Bruhn, Jörgen
Comparative in-vitro Study of the Trachospray, a New Device for Topical Anaesthesia of the Upper Airway
title Comparative in-vitro Study of the Trachospray, a New Device for Topical Anaesthesia of the Upper Airway
title_full Comparative in-vitro Study of the Trachospray, a New Device for Topical Anaesthesia of the Upper Airway
title_fullStr Comparative in-vitro Study of the Trachospray, a New Device for Topical Anaesthesia of the Upper Airway
title_full_unstemmed Comparative in-vitro Study of the Trachospray, a New Device for Topical Anaesthesia of the Upper Airway
title_short Comparative in-vitro Study of the Trachospray, a New Device for Topical Anaesthesia of the Upper Airway
title_sort comparative in-vitro study of the trachospray, a new device for topical anaesthesia of the upper airway
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7837536/
https://www.ncbi.nlm.nih.gov/pubmed/33519246
http://dx.doi.org/10.2147/MDER.S292529
work_keys_str_mv AT vangeffengeertjan comparativeinvitrostudyofthetrachosprayanewdevicefortopicalanaesthesiaoftheupperairway
AT markerinkhielke comparativeinvitrostudyofthetrachosprayanewdevicefortopicalanaesthesiaoftheupperairway
AT vanbarneveldmarc comparativeinvitrostudyofthetrachosprayanewdevicefortopicalanaesthesiaoftheupperairway
AT verhoevenfrank comparativeinvitrostudyofthetrachosprayanewdevicefortopicalanaesthesiaoftheupperairway
AT scheffergertjan comparativeinvitrostudyofthetrachosprayanewdevicefortopicalanaesthesiaoftheupperairway
AT bruhnjorgen comparativeinvitrostudyofthetrachosprayanewdevicefortopicalanaesthesiaoftheupperairway