Comparison of ocular biometric measurements in patients with cataract using three swept-source optical coherence tomography devices
BACKGROUND: Precise measurement of ocular biometry is critical for determining intraocular lens power. Newly developed swept-source optical coherence tomography (SS-OCT) - based ocular biometric devices, ANTERION and CASIA2 provide ocular biometric measurements as IOLMaster 700. This study aimed to...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7839224/ https://www.ncbi.nlm.nih.gov/pubmed/33504333 http://dx.doi.org/10.1186/s12886-021-01826-5 |
_version_ | 1783643352523079680 |
---|---|
author | Oh, Richul Oh, Joo Youn Choi, Hyuk Jin Kim, Mee Kum Yoon, Chang Ho |
author_facet | Oh, Richul Oh, Joo Youn Choi, Hyuk Jin Kim, Mee Kum Yoon, Chang Ho |
author_sort | Oh, Richul |
collection | PubMed |
description | BACKGROUND: Precise measurement of ocular biometry is critical for determining intraocular lens power. Newly developed swept-source optical coherence tomography (SS-OCT) - based ocular biometric devices, ANTERION and CASIA2 provide ocular biometric measurements as IOLMaster 700. This study aimed to assess agreement between three devices. METHODS: This retrospective comparative study includes patients with cataract who underwent ocular biometric measurements with three devices, ANTERION, CASIA2, and IOLMaster 700, at Seoul National University Hospital, in April 2020. Anterior keratometry, total keratometry, central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), and axial length (AL) were the main parameters for the comparison. To assess the agreement between the devices, intraclass coefficient (ICC) and Bland-Altman analysis with 95% limits of agreement (LoA) were used. RESULTS: A total of 47 eyes of 29 patients were measured with three devices. Average anterior keratometry showed excellent agreement (ICC ≥ 0.989), and the mean difference was less than 0.1 D. However, the ICC of the total average keratometry ranged from 0.808 to 0.952, and the difference was more than 0.43 D. The AL measured by ANTERION and IOLMaster 700 showed excellent agreement (ICC = 0.999), and the mean difference was 0.005 mm. The ANTERION and IOLMaster 700 did not obtain AL in six (12.8%) and three (6.4%) cases, respectively (P = 0.001 by Fisher’s exact test). The CCT, ACD, and LT also showed excellent agreement (ICC > 0.9). CONCLUSIONS: The new SS-OCT-based devices, ANTERION, and CASIA2 showed a good agreement with IOLMaster 700 in measuring ocular biometry except for the total keratometry. The AL of ANTERION and IOLMaster 700 showed excellent agreement. |
format | Online Article Text |
id | pubmed-7839224 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-78392242021-01-27 Comparison of ocular biometric measurements in patients with cataract using three swept-source optical coherence tomography devices Oh, Richul Oh, Joo Youn Choi, Hyuk Jin Kim, Mee Kum Yoon, Chang Ho BMC Ophthalmol Research Article BACKGROUND: Precise measurement of ocular biometry is critical for determining intraocular lens power. Newly developed swept-source optical coherence tomography (SS-OCT) - based ocular biometric devices, ANTERION and CASIA2 provide ocular biometric measurements as IOLMaster 700. This study aimed to assess agreement between three devices. METHODS: This retrospective comparative study includes patients with cataract who underwent ocular biometric measurements with three devices, ANTERION, CASIA2, and IOLMaster 700, at Seoul National University Hospital, in April 2020. Anterior keratometry, total keratometry, central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), and axial length (AL) were the main parameters for the comparison. To assess the agreement between the devices, intraclass coefficient (ICC) and Bland-Altman analysis with 95% limits of agreement (LoA) were used. RESULTS: A total of 47 eyes of 29 patients were measured with three devices. Average anterior keratometry showed excellent agreement (ICC ≥ 0.989), and the mean difference was less than 0.1 D. However, the ICC of the total average keratometry ranged from 0.808 to 0.952, and the difference was more than 0.43 D. The AL measured by ANTERION and IOLMaster 700 showed excellent agreement (ICC = 0.999), and the mean difference was 0.005 mm. The ANTERION and IOLMaster 700 did not obtain AL in six (12.8%) and three (6.4%) cases, respectively (P = 0.001 by Fisher’s exact test). The CCT, ACD, and LT also showed excellent agreement (ICC > 0.9). CONCLUSIONS: The new SS-OCT-based devices, ANTERION, and CASIA2 showed a good agreement with IOLMaster 700 in measuring ocular biometry except for the total keratometry. The AL of ANTERION and IOLMaster 700 showed excellent agreement. BioMed Central 2021-01-27 /pmc/articles/PMC7839224/ /pubmed/33504333 http://dx.doi.org/10.1186/s12886-021-01826-5 Text en © The Author(s) 2021 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Research Article Oh, Richul Oh, Joo Youn Choi, Hyuk Jin Kim, Mee Kum Yoon, Chang Ho Comparison of ocular biometric measurements in patients with cataract using three swept-source optical coherence tomography devices |
title | Comparison of ocular biometric measurements in patients with cataract using three swept-source optical coherence tomography devices |
title_full | Comparison of ocular biometric measurements in patients with cataract using three swept-source optical coherence tomography devices |
title_fullStr | Comparison of ocular biometric measurements in patients with cataract using three swept-source optical coherence tomography devices |
title_full_unstemmed | Comparison of ocular biometric measurements in patients with cataract using three swept-source optical coherence tomography devices |
title_short | Comparison of ocular biometric measurements in patients with cataract using three swept-source optical coherence tomography devices |
title_sort | comparison of ocular biometric measurements in patients with cataract using three swept-source optical coherence tomography devices |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7839224/ https://www.ncbi.nlm.nih.gov/pubmed/33504333 http://dx.doi.org/10.1186/s12886-021-01826-5 |
work_keys_str_mv | AT ohrichul comparisonofocularbiometricmeasurementsinpatientswithcataractusingthreesweptsourceopticalcoherencetomographydevices AT ohjooyoun comparisonofocularbiometricmeasurementsinpatientswithcataractusingthreesweptsourceopticalcoherencetomographydevices AT choihyukjin comparisonofocularbiometricmeasurementsinpatientswithcataractusingthreesweptsourceopticalcoherencetomographydevices AT kimmeekum comparisonofocularbiometricmeasurementsinpatientswithcataractusingthreesweptsourceopticalcoherencetomographydevices AT yoonchangho comparisonofocularbiometricmeasurementsinpatientswithcataractusingthreesweptsourceopticalcoherencetomographydevices |