Precision of computed tomography and cartilage-reproducing image reconstruction method in generating digital model for potential use in 3D printing of patient-specific radial head prosthesis: a human cadaver study

BACKGROUND: A prosthetic replacement is a standard treatment for an irreparable radial head fracture; however, the surface mismatch of the commercially available designs is concerned for the long-term cartilage wear. The patient-specific implant created from 3D printing technology could be favorable...

Descripción completa

Detalles Bibliográficos
Autores principales: Luenam, Suriya, Bantuchai, Theeraset, Kosiyatrakul, Arkaphat, Chanpoo, Malee, Phakdeewisetkul, Kantapat, Puncreobutr, Chedtha
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7842042/
https://www.ncbi.nlm.nih.gov/pubmed/33507428
http://dx.doi.org/10.1186/s41205-021-00093-w
_version_ 1783643933359734784
author Luenam, Suriya
Bantuchai, Theeraset
Kosiyatrakul, Arkaphat
Chanpoo, Malee
Phakdeewisetkul, Kantapat
Puncreobutr, Chedtha
author_facet Luenam, Suriya
Bantuchai, Theeraset
Kosiyatrakul, Arkaphat
Chanpoo, Malee
Phakdeewisetkul, Kantapat
Puncreobutr, Chedtha
author_sort Luenam, Suriya
collection PubMed
description BACKGROUND: A prosthetic replacement is a standard treatment for an irreparable radial head fracture; however, the surface mismatch of the commercially available designs is concerned for the long-term cartilage wear. The patient-specific implant created from 3D printing technology could be favorable in replicating the normal anatomy and possibly reduce such sequela. Our study aimed to assess the precision of the computed tomography (CT) and cartilage-reproducing image reconstruction method (CIRM) in generating digital models for potentially use in manufacturing the patient-specific prosthesis from 3D printing. METHODS: Eight intact  elbows (3 right and 5 left) from 7 formalin-embalmed cadavers (4 males and 3 females) with mean age of 83 years (range, 79–94 years) were used for this study. Computerized 3D models were generated from CT, and CIRM. The cartilage-reproducing image reconstruction method has compensated the cartilage profile based on the distance between the subchondral surfaces of the radial head and surrounding bones in CT images. The models of actual radial head geometry used as the gold standard was generated from CT arthrography (CTA). All models of each specimen were matched by registering the surface area of radial neck along with the tuberosity. The difference of head diameter, head thickness, and articular disc depth among three models was evaluated and analyzed by Friedman ANOVA and multiple comparison test using Bonferroni method for statistical correction. A p-value of less than 0.01 was considered statistically  significant. The difference of overall 3D geometry was measured with the root mean square of adjacent point pairs. RESULTS: The analysis displayed the difference of diameter, thickness, and disc depth across the models (p< 0.01). Pairwise comparisons revealed statistically significant difference of all parameters between CTA models and CT models (p< 0.01) whereas no difference was found between CTA models and CIRM models. The mean difference of overall 3D geometry between CTA models and CT models was 0.51±0.24 mm, and between CTA models and CIRM models was 0.24±0.10 mm. CONCLUSIONS: CIRM demonstrated encouraging results in reestablish the normal anatomy and could be potentially used in production process of 3D printed patient-specific radial head prosthesis.
format Online
Article
Text
id pubmed-7842042
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-78420422021-01-28 Precision of computed tomography and cartilage-reproducing image reconstruction method in generating digital model for potential use in 3D printing of patient-specific radial head prosthesis: a human cadaver study Luenam, Suriya Bantuchai, Theeraset Kosiyatrakul, Arkaphat Chanpoo, Malee Phakdeewisetkul, Kantapat Puncreobutr, Chedtha 3D Print Med Research BACKGROUND: A prosthetic replacement is a standard treatment for an irreparable radial head fracture; however, the surface mismatch of the commercially available designs is concerned for the long-term cartilage wear. The patient-specific implant created from 3D printing technology could be favorable in replicating the normal anatomy and possibly reduce such sequela. Our study aimed to assess the precision of the computed tomography (CT) and cartilage-reproducing image reconstruction method (CIRM) in generating digital models for potentially use in manufacturing the patient-specific prosthesis from 3D printing. METHODS: Eight intact  elbows (3 right and 5 left) from 7 formalin-embalmed cadavers (4 males and 3 females) with mean age of 83 years (range, 79–94 years) were used for this study. Computerized 3D models were generated from CT, and CIRM. The cartilage-reproducing image reconstruction method has compensated the cartilage profile based on the distance between the subchondral surfaces of the radial head and surrounding bones in CT images. The models of actual radial head geometry used as the gold standard was generated from CT arthrography (CTA). All models of each specimen were matched by registering the surface area of radial neck along with the tuberosity. The difference of head diameter, head thickness, and articular disc depth among three models was evaluated and analyzed by Friedman ANOVA and multiple comparison test using Bonferroni method for statistical correction. A p-value of less than 0.01 was considered statistically  significant. The difference of overall 3D geometry was measured with the root mean square of adjacent point pairs. RESULTS: The analysis displayed the difference of diameter, thickness, and disc depth across the models (p< 0.01). Pairwise comparisons revealed statistically significant difference of all parameters between CTA models and CT models (p< 0.01) whereas no difference was found between CTA models and CIRM models. The mean difference of overall 3D geometry between CTA models and CT models was 0.51±0.24 mm, and between CTA models and CIRM models was 0.24±0.10 mm. CONCLUSIONS: CIRM demonstrated encouraging results in reestablish the normal anatomy and could be potentially used in production process of 3D printed patient-specific radial head prosthesis. Springer International Publishing 2021-01-28 /pmc/articles/PMC7842042/ /pubmed/33507428 http://dx.doi.org/10.1186/s41205-021-00093-w Text en © The Author(s) 2021 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research
Luenam, Suriya
Bantuchai, Theeraset
Kosiyatrakul, Arkaphat
Chanpoo, Malee
Phakdeewisetkul, Kantapat
Puncreobutr, Chedtha
Precision of computed tomography and cartilage-reproducing image reconstruction method in generating digital model for potential use in 3D printing of patient-specific radial head prosthesis: a human cadaver study
title Precision of computed tomography and cartilage-reproducing image reconstruction method in generating digital model for potential use in 3D printing of patient-specific radial head prosthesis: a human cadaver study
title_full Precision of computed tomography and cartilage-reproducing image reconstruction method in generating digital model for potential use in 3D printing of patient-specific radial head prosthesis: a human cadaver study
title_fullStr Precision of computed tomography and cartilage-reproducing image reconstruction method in generating digital model for potential use in 3D printing of patient-specific radial head prosthesis: a human cadaver study
title_full_unstemmed Precision of computed tomography and cartilage-reproducing image reconstruction method in generating digital model for potential use in 3D printing of patient-specific radial head prosthesis: a human cadaver study
title_short Precision of computed tomography and cartilage-reproducing image reconstruction method in generating digital model for potential use in 3D printing of patient-specific radial head prosthesis: a human cadaver study
title_sort precision of computed tomography and cartilage-reproducing image reconstruction method in generating digital model for potential use in 3d printing of patient-specific radial head prosthesis: a human cadaver study
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7842042/
https://www.ncbi.nlm.nih.gov/pubmed/33507428
http://dx.doi.org/10.1186/s41205-021-00093-w
work_keys_str_mv AT luenamsuriya precisionofcomputedtomographyandcartilagereproducingimagereconstructionmethodingeneratingdigitalmodelforpotentialusein3dprintingofpatientspecificradialheadprosthesisahumancadaverstudy
AT bantuchaitheeraset precisionofcomputedtomographyandcartilagereproducingimagereconstructionmethodingeneratingdigitalmodelforpotentialusein3dprintingofpatientspecificradialheadprosthesisahumancadaverstudy
AT kosiyatrakularkaphat precisionofcomputedtomographyandcartilagereproducingimagereconstructionmethodingeneratingdigitalmodelforpotentialusein3dprintingofpatientspecificradialheadprosthesisahumancadaverstudy
AT chanpoomalee precisionofcomputedtomographyandcartilagereproducingimagereconstructionmethodingeneratingdigitalmodelforpotentialusein3dprintingofpatientspecificradialheadprosthesisahumancadaverstudy
AT phakdeewisetkulkantapat precisionofcomputedtomographyandcartilagereproducingimagereconstructionmethodingeneratingdigitalmodelforpotentialusein3dprintingofpatientspecificradialheadprosthesisahumancadaverstudy
AT puncreobutrchedtha precisionofcomputedtomographyandcartilagereproducingimagereconstructionmethodingeneratingdigitalmodelforpotentialusein3dprintingofpatientspecificradialheadprosthesisahumancadaverstudy