In vivo analysis of the origin and characteristics of gaseous microemboli during catheter-mediated irreversible electroporation
AIMS : Irreversible electroporation (IRE) ablation is a non-thermal ablation method based on the application of direct current between a multi-electrode catheter and skin electrode. The delivery of current through blood leads to electrolysis. Some studies suggest that gaseous (micro)emboli might be...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7842095/ https://www.ncbi.nlm.nih.gov/pubmed/33111141 http://dx.doi.org/10.1093/europace/euaa243 |
Sumario: | AIMS : Irreversible electroporation (IRE) ablation is a non-thermal ablation method based on the application of direct current between a multi-electrode catheter and skin electrode. The delivery of current through blood leads to electrolysis. Some studies suggest that gaseous (micro)emboli might be associated with myocardial damage and/or (a)symptomatic cerebral ischaemic events. The aim of this study was to compare the amount of gas generated during IRE ablation and during radiofrequency (RF) ablation. METHODS AND RESULTS: In six 60–75 kg pigs, an extracorporeal femoral shunt was outfitted with a bubble-counter to detect the size and total volume of gas bubbles. Anodal and cathodal 200 J IRE applications were delivered in the left atrium (LA) using a 14-electrode circular catheter. The 30 and 60 s 40 W RF point-by-point ablations were performed. Using transoesophageal echocardiography (TOE), gas formation was visualized. Average gas volumes were 0.6 ± 0.6 and 56.9 ± 19.1 μL (P < 0.01) for each anodal and cathodal IRE application, respectively. Also, qualitative TOE imaging showed significantly less LA bubble contrast with anodal than with cathodal applications. Radiofrequency ablations produced 1.7 ± 2.9 and 6.7 ± 7.4 μL of gas, for 30 and 60 s ablation time, respectively. CONCLUSION : Anodal IRE applications result in significantly less gas formation than both cathodal IRE applications and RF applications. This finding is supported by TOE observations. |
---|