Comparison of effectiveness of the piston-pump method versus the pressure-infusor method for rapid infusion of crystalloids: A bench study

BACKGROUND AND AIMS: The piston-pump method is a simple method for rapid administration of fluids but some problems are unsolved. We compared the effectiveness of using the piston-pump method with that of the pressure-infusor method. METHODS: Twelve anaesthetists were classified randomly into the pi...

Descripción completa

Detalles Bibliográficos
Autores principales: Hashimoto, Wataru, Takenaka, Ichiro, Yasunami, Keisuke, Minami, Tomoko, Sano, Haruhiko
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7852453/
https://www.ncbi.nlm.nih.gov/pubmed/33542570
http://dx.doi.org/10.4103/ija.IJA_864_20
_version_ 1783645826536439808
author Hashimoto, Wataru
Takenaka, Ichiro
Yasunami, Keisuke
Minami, Tomoko
Sano, Haruhiko
author_facet Hashimoto, Wataru
Takenaka, Ichiro
Yasunami, Keisuke
Minami, Tomoko
Sano, Haruhiko
author_sort Hashimoto, Wataru
collection PubMed
description BACKGROUND AND AIMS: The piston-pump method is a simple method for rapid administration of fluids but some problems are unsolved. We compared the effectiveness of using the piston-pump method with that of the pressure-infusor method. METHODS: Twelve anaesthetists were classified randomly into the piston-pump and pressure-infusor groups. They were asked to infuse 500 ml of saline three times successively through a 16-G intravenous cannula as rapidly as possible using a pump with a 50-ml syringe or a pressure-infusor at 300 mmHg. The time taken for infusion and the maximum or minimum pressure in the infusion circuit and substitute vessel were measured. Bacterial culture of the saline infused sterilely was performed to estimate bacterial contamination. RESULTS: The pressure-infusor group led to faster infusion of 500 ml of saline (233 ± 19 s) than the piston-pump group (301 ± 48 s) (P < 0.01). The infusion time at the third attempt (316 ± 43 s) was significantly longer than that at the first attempt (285 ± 53 s) only in the piston-pump group (P < 0.05). The maximum pressure (mmHg) in the circuit was 131 ± 9 and > 200 (P < 0.01) and in the substitute vessel was 5 ± 1 and 17 ± 7 (P < 0.01) in the pressure-infusor and piston-pump groups, respectively. A pressure of <-200 mmHg occurred at all infusion attempts in the piston-pump group. Bacterial contamination was not observed in either group. CONCLUSION: If fluids must be administered rapidly, the pressure-infusor method is more efficient than the piston-pump method because the latter is less effective in infusing fluids rapidly and associated with excessive positive and negative pressure in the infusion circuit.
format Online
Article
Text
id pubmed-7852453
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-78524532021-02-03 Comparison of effectiveness of the piston-pump method versus the pressure-infusor method for rapid infusion of crystalloids: A bench study Hashimoto, Wataru Takenaka, Ichiro Yasunami, Keisuke Minami, Tomoko Sano, Haruhiko Indian J Anaesth Original Article BACKGROUND AND AIMS: The piston-pump method is a simple method for rapid administration of fluids but some problems are unsolved. We compared the effectiveness of using the piston-pump method with that of the pressure-infusor method. METHODS: Twelve anaesthetists were classified randomly into the piston-pump and pressure-infusor groups. They were asked to infuse 500 ml of saline three times successively through a 16-G intravenous cannula as rapidly as possible using a pump with a 50-ml syringe or a pressure-infusor at 300 mmHg. The time taken for infusion and the maximum or minimum pressure in the infusion circuit and substitute vessel were measured. Bacterial culture of the saline infused sterilely was performed to estimate bacterial contamination. RESULTS: The pressure-infusor group led to faster infusion of 500 ml of saline (233 ± 19 s) than the piston-pump group (301 ± 48 s) (P < 0.01). The infusion time at the third attempt (316 ± 43 s) was significantly longer than that at the first attempt (285 ± 53 s) only in the piston-pump group (P < 0.05). The maximum pressure (mmHg) in the circuit was 131 ± 9 and > 200 (P < 0.01) and in the substitute vessel was 5 ± 1 and 17 ± 7 (P < 0.01) in the pressure-infusor and piston-pump groups, respectively. A pressure of <-200 mmHg occurred at all infusion attempts in the piston-pump group. Bacterial contamination was not observed in either group. CONCLUSION: If fluids must be administered rapidly, the pressure-infusor method is more efficient than the piston-pump method because the latter is less effective in infusing fluids rapidly and associated with excessive positive and negative pressure in the infusion circuit. Wolters Kluwer - Medknow 2020-12 2020-12-12 /pmc/articles/PMC7852453/ /pubmed/33542570 http://dx.doi.org/10.4103/ija.IJA_864_20 Text en Copyright: © 2020 Indian Journal of Anaesthesia http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Hashimoto, Wataru
Takenaka, Ichiro
Yasunami, Keisuke
Minami, Tomoko
Sano, Haruhiko
Comparison of effectiveness of the piston-pump method versus the pressure-infusor method for rapid infusion of crystalloids: A bench study
title Comparison of effectiveness of the piston-pump method versus the pressure-infusor method for rapid infusion of crystalloids: A bench study
title_full Comparison of effectiveness of the piston-pump method versus the pressure-infusor method for rapid infusion of crystalloids: A bench study
title_fullStr Comparison of effectiveness of the piston-pump method versus the pressure-infusor method for rapid infusion of crystalloids: A bench study
title_full_unstemmed Comparison of effectiveness of the piston-pump method versus the pressure-infusor method for rapid infusion of crystalloids: A bench study
title_short Comparison of effectiveness of the piston-pump method versus the pressure-infusor method for rapid infusion of crystalloids: A bench study
title_sort comparison of effectiveness of the piston-pump method versus the pressure-infusor method for rapid infusion of crystalloids: a bench study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7852453/
https://www.ncbi.nlm.nih.gov/pubmed/33542570
http://dx.doi.org/10.4103/ija.IJA_864_20
work_keys_str_mv AT hashimotowataru comparisonofeffectivenessofthepistonpumpmethodversusthepressureinfusormethodforrapidinfusionofcrystalloidsabenchstudy
AT takenakaichiro comparisonofeffectivenessofthepistonpumpmethodversusthepressureinfusormethodforrapidinfusionofcrystalloidsabenchstudy
AT yasunamikeisuke comparisonofeffectivenessofthepistonpumpmethodversusthepressureinfusormethodforrapidinfusionofcrystalloidsabenchstudy
AT minamitomoko comparisonofeffectivenessofthepistonpumpmethodversusthepressureinfusormethodforrapidinfusionofcrystalloidsabenchstudy
AT sanoharuhiko comparisonofeffectivenessofthepistonpumpmethodversusthepressureinfusormethodforrapidinfusionofcrystalloidsabenchstudy