Case report of a uniocular topography guided laser-assisted in situ keratomileusis enhancement following an incorrectly treated astigmatic axis
A 23-year-old male presented to us wanting spectacle removal for cosmetic purposes. He underwent bilateral wavefront optimized (WFO) laser-assisted in situ keratomileusis (LASIK) on the Alcon Wavelight(®) EX-500 excimer laser with an incorrectly treated astigmatism axis for left eye due to a manual...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7856993/ https://www.ncbi.nlm.nih.gov/pubmed/33229703 http://dx.doi.org/10.4103/ijo.IJO_1213_20 |
Sumario: | A 23-year-old male presented to us wanting spectacle removal for cosmetic purposes. He underwent bilateral wavefront optimized (WFO) laser-assisted in situ keratomileusis (LASIK) on the Alcon Wavelight(®) EX-500 excimer laser with an incorrectly treated astigmatism axis for left eye due to a manual data entry error in the laser. WFO LASIK treats the sphere and cylinder only. LASIK enhancement with topographic-guided ablation resulted in the elimination of all refractive errors and gave excellent results. Wavelight(®) topographic-guided treatment can perform two separate layers of correction in the same ablation: The first is to treat the corneal irregularities for the higher order aberration (HOA) removal, the second one meant to treat the sphere and cylinder if indicated. |
---|