A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System
Although lateral locking plates are often a preferred and successful fixation construct for the treatment of periprosthetic proximal and distal femur fractures, specific complications and modes of failure have been associated and well-described with their use. We present two cases of implant failure...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7861960/ https://www.ncbi.nlm.nih.gov/pubmed/33543880 http://dx.doi.org/10.5435/JAAOSGlobal-D-20-00143 |
_version_ | 1783647183987277824 |
---|---|
author | Stockwell, Erin Mormino, Matthew A. Siebler, Justin C. |
author_facet | Stockwell, Erin Mormino, Matthew A. Siebler, Justin C. |
author_sort | Stockwell, Erin |
collection | PubMed |
description | Although lateral locking plates are often a preferred and successful fixation construct for the treatment of periprosthetic proximal and distal femur fractures, specific complications and modes of failure have been associated and well-described with their use. We present two cases of implant failure in the Non-Contact Bridge Periprosthetic Plating System (Zimmer Biomet) in which a nonlocked screw fretted through the annular seating of the plate. One case demonstrates failure in the setting of a proximal femur periprosthetic fracture, whereas the other demonstrates failure in the setting of a distal femur periprosthetic fracture. This unique mode of failure has not previously been reported in the literature. |
format | Online Article Text |
id | pubmed-7861960 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Wolters Kluwer |
record_format | MEDLINE/PubMed |
spelling | pubmed-78619602021-02-05 A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System Stockwell, Erin Mormino, Matthew A. Siebler, Justin C. J Am Acad Orthop Surg Glob Res Rev Case Report Although lateral locking plates are often a preferred and successful fixation construct for the treatment of periprosthetic proximal and distal femur fractures, specific complications and modes of failure have been associated and well-described with their use. We present two cases of implant failure in the Non-Contact Bridge Periprosthetic Plating System (Zimmer Biomet) in which a nonlocked screw fretted through the annular seating of the plate. One case demonstrates failure in the setting of a proximal femur periprosthetic fracture, whereas the other demonstrates failure in the setting of a distal femur periprosthetic fracture. This unique mode of failure has not previously been reported in the literature. Wolters Kluwer 2021-02-02 /pmc/articles/PMC7861960/ /pubmed/33543880 http://dx.doi.org/10.5435/JAAOSGlobal-D-20-00143 Text en Copyright © 2021 The Authors. Published by Wolters Kluwer Health, Inc. on behalf of the American Academy of Orthopaedic Surgeons. This is an open access article distributed under the Creative Commons Attribution License 4.0 (CCBY) (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Case Report Stockwell, Erin Mormino, Matthew A. Siebler, Justin C. A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System |
title | A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System |
title_full | A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System |
title_fullStr | A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System |
title_full_unstemmed | A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System |
title_short | A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System |
title_sort | unique mode of failure in the noncontact bridging periprosthetic plating system |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7861960/ https://www.ncbi.nlm.nih.gov/pubmed/33543880 http://dx.doi.org/10.5435/JAAOSGlobal-D-20-00143 |
work_keys_str_mv | AT stockwellerin auniquemodeoffailureinthenoncontactbridgingperiprostheticplatingsystem AT morminomatthewa auniquemodeoffailureinthenoncontactbridgingperiprostheticplatingsystem AT sieblerjustinc auniquemodeoffailureinthenoncontactbridgingperiprostheticplatingsystem AT stockwellerin uniquemodeoffailureinthenoncontactbridgingperiprostheticplatingsystem AT morminomatthewa uniquemodeoffailureinthenoncontactbridgingperiprostheticplatingsystem AT sieblerjustinc uniquemodeoffailureinthenoncontactbridgingperiprostheticplatingsystem |