Cargando…

A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System

Although lateral locking plates are often a preferred and successful fixation construct for the treatment of periprosthetic proximal and distal femur fractures, specific complications and modes of failure have been associated and well-described with their use. We present two cases of implant failure...

Descripción completa

Detalles Bibliográficos
Autores principales: Stockwell, Erin, Mormino, Matthew A., Siebler, Justin C.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7861960/
https://www.ncbi.nlm.nih.gov/pubmed/33543880
http://dx.doi.org/10.5435/JAAOSGlobal-D-20-00143
_version_ 1783647183987277824
author Stockwell, Erin
Mormino, Matthew A.
Siebler, Justin C.
author_facet Stockwell, Erin
Mormino, Matthew A.
Siebler, Justin C.
author_sort Stockwell, Erin
collection PubMed
description Although lateral locking plates are often a preferred and successful fixation construct for the treatment of periprosthetic proximal and distal femur fractures, specific complications and modes of failure have been associated and well-described with their use. We present two cases of implant failure in the Non-Contact Bridge Periprosthetic Plating System (Zimmer Biomet) in which a nonlocked screw fretted through the annular seating of the plate. One case demonstrates failure in the setting of a proximal femur periprosthetic fracture, whereas the other demonstrates failure in the setting of a distal femur periprosthetic fracture. This unique mode of failure has not previously been reported in the literature.
format Online
Article
Text
id pubmed-7861960
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Wolters Kluwer
record_format MEDLINE/PubMed
spelling pubmed-78619602021-02-05 A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System Stockwell, Erin Mormino, Matthew A. Siebler, Justin C. J Am Acad Orthop Surg Glob Res Rev Case Report Although lateral locking plates are often a preferred and successful fixation construct for the treatment of periprosthetic proximal and distal femur fractures, specific complications and modes of failure have been associated and well-described with their use. We present two cases of implant failure in the Non-Contact Bridge Periprosthetic Plating System (Zimmer Biomet) in which a nonlocked screw fretted through the annular seating of the plate. One case demonstrates failure in the setting of a proximal femur periprosthetic fracture, whereas the other demonstrates failure in the setting of a distal femur periprosthetic fracture. This unique mode of failure has not previously been reported in the literature. Wolters Kluwer 2021-02-02 /pmc/articles/PMC7861960/ /pubmed/33543880 http://dx.doi.org/10.5435/JAAOSGlobal-D-20-00143 Text en Copyright © 2021 The Authors. Published by Wolters Kluwer Health, Inc. on behalf of the American Academy of Orthopaedic Surgeons. This is an open access article distributed under the Creative Commons Attribution License 4.0 (CCBY) (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Case Report
Stockwell, Erin
Mormino, Matthew A.
Siebler, Justin C.
A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System
title A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System
title_full A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System
title_fullStr A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System
title_full_unstemmed A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System
title_short A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System
title_sort unique mode of failure in the noncontact bridging periprosthetic plating system
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7861960/
https://www.ncbi.nlm.nih.gov/pubmed/33543880
http://dx.doi.org/10.5435/JAAOSGlobal-D-20-00143
work_keys_str_mv AT stockwellerin auniquemodeoffailureinthenoncontactbridgingperiprostheticplatingsystem
AT morminomatthewa auniquemodeoffailureinthenoncontactbridgingperiprostheticplatingsystem
AT sieblerjustinc auniquemodeoffailureinthenoncontactbridgingperiprostheticplatingsystem
AT stockwellerin uniquemodeoffailureinthenoncontactbridgingperiprostheticplatingsystem
AT morminomatthewa uniquemodeoffailureinthenoncontactbridgingperiprostheticplatingsystem
AT sieblerjustinc uniquemodeoffailureinthenoncontactbridgingperiprostheticplatingsystem