Cargando…

A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System

Although lateral locking plates are often a preferred and successful fixation construct for the treatment of periprosthetic proximal and distal femur fractures, specific complications and modes of failure have been associated and well-described with their use. We present two cases of implant failure...

Descripción completa

Detalles Bibliográficos
Autores principales: Stockwell, Erin, Mormino, Matthew A., Siebler, Justin C.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7861960/
https://www.ncbi.nlm.nih.gov/pubmed/33543880
http://dx.doi.org/10.5435/JAAOSGlobal-D-20-00143

Ejemplares similares