Accuracy of Guided Implant Surgery in the Edentulous Jaw Using Desktop 3D-Printed Mucosal Supported Guides
Purpose: The aim of this in vitro study is to evaluate the accuracy of implant position using mucosal supported surgical guides, produced by a desktop 3D printer. Methods: Ninety implants (Bone Level Roxolid, 4.1 mm × 10 mm, Straumann, Villerat, Switzerland) were placed in fifteen mandibular casts (...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7864156/ https://www.ncbi.nlm.nih.gov/pubmed/33498516 http://dx.doi.org/10.3390/jcm10030391 |
_version_ | 1783647611285143552 |
---|---|
author | D’haese, Rani Vrombaut, Tom Hommez, Geert De Bruyn, Hugo Vandeweghe, Stefan |
author_facet | D’haese, Rani Vrombaut, Tom Hommez, Geert De Bruyn, Hugo Vandeweghe, Stefan |
author_sort | D’haese, Rani |
collection | PubMed |
description | Purpose: The aim of this in vitro study is to evaluate the accuracy of implant position using mucosal supported surgical guides, produced by a desktop 3D printer. Methods: Ninety implants (Bone Level Roxolid, 4.1 mm × 10 mm, Straumann, Villerat, Switzerland) were placed in fifteen mandibular casts (Bonemodels, Castellón de la Plana, Spain). A mucosa-supported guide was designed and printed for each of the fifteen casts. After placement of the implants, the location was assessed by scanning the cast and scan bodies with an intra-oral scanner (Primescan(®), Dentsply Sirona, York, PA, USA). Two comparisons were performed: one with the mucosa as a reference, and one where only the implants were aligned. Angular, coronal and apical deviations were measured. Results: The mean implant angular deviation for tissue and implant alignment were 3.25° (SD 1.69°) and 2.39° (SD 1.42°) respectively, the coronal deviation 0.82 mm (SD 0.43 mm) and 0.45 mm (SD 0.31 mm) and the apical deviation 0.99 mm (SD 0.45 mm) and 0.71 mm (SD 0.43 mm). All three variables were significantly different between the tissue and implant alignment (p < 0.001). Conclusion: Based on the results of this study, we conclude that guided implant surgery using desktop 3D printed mucosa-supported guides has a clinically acceptable level of accuracy. The resilience of the mucosa has a negative effect on the guide stability and increases the deviation in implant position. |
format | Online Article Text |
id | pubmed-7864156 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-78641562021-02-06 Accuracy of Guided Implant Surgery in the Edentulous Jaw Using Desktop 3D-Printed Mucosal Supported Guides D’haese, Rani Vrombaut, Tom Hommez, Geert De Bruyn, Hugo Vandeweghe, Stefan J Clin Med Article Purpose: The aim of this in vitro study is to evaluate the accuracy of implant position using mucosal supported surgical guides, produced by a desktop 3D printer. Methods: Ninety implants (Bone Level Roxolid, 4.1 mm × 10 mm, Straumann, Villerat, Switzerland) were placed in fifteen mandibular casts (Bonemodels, Castellón de la Plana, Spain). A mucosa-supported guide was designed and printed for each of the fifteen casts. After placement of the implants, the location was assessed by scanning the cast and scan bodies with an intra-oral scanner (Primescan(®), Dentsply Sirona, York, PA, USA). Two comparisons were performed: one with the mucosa as a reference, and one where only the implants were aligned. Angular, coronal and apical deviations were measured. Results: The mean implant angular deviation for tissue and implant alignment were 3.25° (SD 1.69°) and 2.39° (SD 1.42°) respectively, the coronal deviation 0.82 mm (SD 0.43 mm) and 0.45 mm (SD 0.31 mm) and the apical deviation 0.99 mm (SD 0.45 mm) and 0.71 mm (SD 0.43 mm). All three variables were significantly different between the tissue and implant alignment (p < 0.001). Conclusion: Based on the results of this study, we conclude that guided implant surgery using desktop 3D printed mucosa-supported guides has a clinically acceptable level of accuracy. The resilience of the mucosa has a negative effect on the guide stability and increases the deviation in implant position. MDPI 2021-01-20 /pmc/articles/PMC7864156/ /pubmed/33498516 http://dx.doi.org/10.3390/jcm10030391 Text en © 2021 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article D’haese, Rani Vrombaut, Tom Hommez, Geert De Bruyn, Hugo Vandeweghe, Stefan Accuracy of Guided Implant Surgery in the Edentulous Jaw Using Desktop 3D-Printed Mucosal Supported Guides |
title | Accuracy of Guided Implant Surgery in the Edentulous Jaw Using Desktop 3D-Printed Mucosal Supported Guides |
title_full | Accuracy of Guided Implant Surgery in the Edentulous Jaw Using Desktop 3D-Printed Mucosal Supported Guides |
title_fullStr | Accuracy of Guided Implant Surgery in the Edentulous Jaw Using Desktop 3D-Printed Mucosal Supported Guides |
title_full_unstemmed | Accuracy of Guided Implant Surgery in the Edentulous Jaw Using Desktop 3D-Printed Mucosal Supported Guides |
title_short | Accuracy of Guided Implant Surgery in the Edentulous Jaw Using Desktop 3D-Printed Mucosal Supported Guides |
title_sort | accuracy of guided implant surgery in the edentulous jaw using desktop 3d-printed mucosal supported guides |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7864156/ https://www.ncbi.nlm.nih.gov/pubmed/33498516 http://dx.doi.org/10.3390/jcm10030391 |
work_keys_str_mv | AT dhaeserani accuracyofguidedimplantsurgeryintheedentulousjawusingdesktop3dprintedmucosalsupportedguides AT vrombauttom accuracyofguidedimplantsurgeryintheedentulousjawusingdesktop3dprintedmucosalsupportedguides AT hommezgeert accuracyofguidedimplantsurgeryintheedentulousjawusingdesktop3dprintedmucosalsupportedguides AT debruynhugo accuracyofguidedimplantsurgeryintheedentulousjawusingdesktop3dprintedmucosalsupportedguides AT vandeweghestefan accuracyofguidedimplantsurgeryintheedentulousjawusingdesktop3dprintedmucosalsupportedguides |