Polarimetric Balanced Detection: Background-Free Mid-IR Evanescent Field Laser Spectroscopy for Low-Noise, Long-term Stable Chemical Sensing

[Image: see text] In this work, we introduce polarimetric balanced detection as a new attenuated total reflection (ATR) infrared (IR) sensing scheme, leveraging unequal effective thicknesses achieved with laser light of different polarizations. We combined a monolithic widely tunable Vernier quantum...

Descripción completa

Detalles Bibliográficos
Autores principales: Freitag, Stephan, Baer, Matthias, Buntzoll, Laura, Ramer, Georg, Schwaighofer, Andreas, Schmauss, Bernhard, Lendl, Bernhard
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American Chemical Society 2020
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7872502/
https://www.ncbi.nlm.nih.gov/pubmed/33372759
http://dx.doi.org/10.1021/acssensors.0c01342
_version_ 1783649198934065152
author Freitag, Stephan
Baer, Matthias
Buntzoll, Laura
Ramer, Georg
Schwaighofer, Andreas
Schmauss, Bernhard
Lendl, Bernhard
author_facet Freitag, Stephan
Baer, Matthias
Buntzoll, Laura
Ramer, Georg
Schwaighofer, Andreas
Schmauss, Bernhard
Lendl, Bernhard
author_sort Freitag, Stephan
collection PubMed
description [Image: see text] In this work, we introduce polarimetric balanced detection as a new attenuated total reflection (ATR) infrared (IR) sensing scheme, leveraging unequal effective thicknesses achieved with laser light of different polarizations. We combined a monolithic widely tunable Vernier quantum cascade laser (QCL-XT) and a multibounce ATR IR spectroscopy setup for analysis of liquids in a process analytical setting. Polarimetric balanced detection enables simultaneous recording of background and sample spectra, significantly reducing long-term drifts. The root-mean-square noise could be improved by a factor of 10 in a long-term experiment, compared to conventional absorbance measurements obtained via the single-ended optical channel. The sensing performance of the device was further evaluated by on-site measurements of ethanol in water, leading to an improved limit of detection (LOD) achieved with polarimetric balanced detection. Sequential injection analysis was employed for automated injection of samples into a custom-built ATR flow cell mounted above a zinc sulfide multibounce ATR element. The QCL-XT posed to be suitable for mid-IR-based sensing in liquids due to its wide tunability. Polarimetric balanced detection proved to enhance the robustness and long-term stability of the sensing device, along with improving the LOD by a factor of 5. This demonstrates the potential for new polarimetric QCL-based ATR mid-IR sensing schemes for in-field measurements or process monitoring usually prone to a multitude of interferences.
format Online
Article
Text
id pubmed-7872502
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher American Chemical Society
record_format MEDLINE/PubMed
spelling pubmed-78725022021-02-10 Polarimetric Balanced Detection: Background-Free Mid-IR Evanescent Field Laser Spectroscopy for Low-Noise, Long-term Stable Chemical Sensing Freitag, Stephan Baer, Matthias Buntzoll, Laura Ramer, Georg Schwaighofer, Andreas Schmauss, Bernhard Lendl, Bernhard ACS Sens [Image: see text] In this work, we introduce polarimetric balanced detection as a new attenuated total reflection (ATR) infrared (IR) sensing scheme, leveraging unequal effective thicknesses achieved with laser light of different polarizations. We combined a monolithic widely tunable Vernier quantum cascade laser (QCL-XT) and a multibounce ATR IR spectroscopy setup for analysis of liquids in a process analytical setting. Polarimetric balanced detection enables simultaneous recording of background and sample spectra, significantly reducing long-term drifts. The root-mean-square noise could be improved by a factor of 10 in a long-term experiment, compared to conventional absorbance measurements obtained via the single-ended optical channel. The sensing performance of the device was further evaluated by on-site measurements of ethanol in water, leading to an improved limit of detection (LOD) achieved with polarimetric balanced detection. Sequential injection analysis was employed for automated injection of samples into a custom-built ATR flow cell mounted above a zinc sulfide multibounce ATR element. The QCL-XT posed to be suitable for mid-IR-based sensing in liquids due to its wide tunability. Polarimetric balanced detection proved to enhance the robustness and long-term stability of the sensing device, along with improving the LOD by a factor of 5. This demonstrates the potential for new polarimetric QCL-based ATR mid-IR sensing schemes for in-field measurements or process monitoring usually prone to a multitude of interferences. American Chemical Society 2020-12-29 2021-01-22 /pmc/articles/PMC7872502/ /pubmed/33372759 http://dx.doi.org/10.1021/acssensors.0c01342 Text en © 2020 American Chemical Society This is an open access article published under a Creative Commons Attribution (CC-BY) License (http://pubs.acs.org/page/policy/authorchoice_ccby_termsofuse.html) , which permits unrestricted use, distribution and reproduction in any medium, provided the author and source are cited.
spellingShingle Freitag, Stephan
Baer, Matthias
Buntzoll, Laura
Ramer, Georg
Schwaighofer, Andreas
Schmauss, Bernhard
Lendl, Bernhard
Polarimetric Balanced Detection: Background-Free Mid-IR Evanescent Field Laser Spectroscopy for Low-Noise, Long-term Stable Chemical Sensing
title Polarimetric Balanced Detection: Background-Free Mid-IR Evanescent Field Laser Spectroscopy for Low-Noise, Long-term Stable Chemical Sensing
title_full Polarimetric Balanced Detection: Background-Free Mid-IR Evanescent Field Laser Spectroscopy for Low-Noise, Long-term Stable Chemical Sensing
title_fullStr Polarimetric Balanced Detection: Background-Free Mid-IR Evanescent Field Laser Spectroscopy for Low-Noise, Long-term Stable Chemical Sensing
title_full_unstemmed Polarimetric Balanced Detection: Background-Free Mid-IR Evanescent Field Laser Spectroscopy for Low-Noise, Long-term Stable Chemical Sensing
title_short Polarimetric Balanced Detection: Background-Free Mid-IR Evanescent Field Laser Spectroscopy for Low-Noise, Long-term Stable Chemical Sensing
title_sort polarimetric balanced detection: background-free mid-ir evanescent field laser spectroscopy for low-noise, long-term stable chemical sensing
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7872502/
https://www.ncbi.nlm.nih.gov/pubmed/33372759
http://dx.doi.org/10.1021/acssensors.0c01342
work_keys_str_mv AT freitagstephan polarimetricbalanceddetectionbackgroundfreemidirevanescentfieldlaserspectroscopyforlownoiselongtermstablechemicalsensing
AT baermatthias polarimetricbalanceddetectionbackgroundfreemidirevanescentfieldlaserspectroscopyforlownoiselongtermstablechemicalsensing
AT buntzolllaura polarimetricbalanceddetectionbackgroundfreemidirevanescentfieldlaserspectroscopyforlownoiselongtermstablechemicalsensing
AT ramergeorg polarimetricbalanceddetectionbackgroundfreemidirevanescentfieldlaserspectroscopyforlownoiselongtermstablechemicalsensing
AT schwaighoferandreas polarimetricbalanceddetectionbackgroundfreemidirevanescentfieldlaserspectroscopyforlownoiselongtermstablechemicalsensing
AT schmaussbernhard polarimetricbalanceddetectionbackgroundfreemidirevanescentfieldlaserspectroscopyforlownoiselongtermstablechemicalsensing
AT lendlbernhard polarimetricbalanceddetectionbackgroundfreemidirevanescentfieldlaserspectroscopyforlownoiselongtermstablechemicalsensing