A Comparative Study Between the Goldmann Applanation Tonometer and the Non-Contact Air-Puff Tonometer (Huvitz HNT 7000) in Normal Eyes
OBJECTIVE: To assess whether the non-contact air-puff tonometer (NCT) is an appropriate alternative to the Goldmann applanation tonometer (GAT) for measuring intraocular pressure (IOP). PATIENTS AND METHODS: An observational, cross-sectional, and comparative study with a quantitative approach was ca...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Dove
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7872931/ https://www.ncbi.nlm.nih.gov/pubmed/33574653 http://dx.doi.org/10.2147/OPTH.S294710 |
_version_ | 1783649285279055872 |
---|---|
author | Stock, Ricardo Alexandre Ströher, Carine Sampaio, Rodrigo Rosa Mergener, Rafael André Bonamigo, Elcio Luiz |
author_facet | Stock, Ricardo Alexandre Ströher, Carine Sampaio, Rodrigo Rosa Mergener, Rafael André Bonamigo, Elcio Luiz |
author_sort | Stock, Ricardo Alexandre |
collection | PubMed |
description | OBJECTIVE: To assess whether the non-contact air-puff tonometer (NCT) is an appropriate alternative to the Goldmann applanation tonometer (GAT) for measuring intraocular pressure (IOP). PATIENTS AND METHODS: An observational, cross-sectional, and comparative study with a quantitative approach was carried out. Two techniques for IOP measurements using the standard GAT and the NCT were compared. A total of 180 eyes from 90 patients were included in the study. RESULTS: The total mean IOP according to NCT measurements was 14.12 mmHg, and the total mean IOP according to the GAT was 12.98 mmHg; these values were significantly different (p=0.0018). When dividing the participants into three groups according to the measurement range obtained and comparing the mean NCT and GAT measurements in each group, in Group 1 (10–15 mmHg), no statistically significant difference was found between the means of the two tonometers (p=0.3100), a difference was observed between Group 2 (16–19 mmHg) and Group 3 (20 mmHg or more) (p<0.001). When dividing the participants by age group, the means obtained by the two tonometers also differed significantly between Group 4 (40–59 years) and Group 5 (60 years or more) (p<0.0001). In all groups, the mean measurements by the NCT were higher than those by the GAT. CONCLUSION: The NCT presented an approximate mean of the measures with the GAT in group 1 but was overestimated in the measurements of the groups 2 and 3. |
format | Online Article Text |
id | pubmed-7872931 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Dove |
record_format | MEDLINE/PubMed |
spelling | pubmed-78729312021-02-10 A Comparative Study Between the Goldmann Applanation Tonometer and the Non-Contact Air-Puff Tonometer (Huvitz HNT 7000) in Normal Eyes Stock, Ricardo Alexandre Ströher, Carine Sampaio, Rodrigo Rosa Mergener, Rafael André Bonamigo, Elcio Luiz Clin Ophthalmol Original Research OBJECTIVE: To assess whether the non-contact air-puff tonometer (NCT) is an appropriate alternative to the Goldmann applanation tonometer (GAT) for measuring intraocular pressure (IOP). PATIENTS AND METHODS: An observational, cross-sectional, and comparative study with a quantitative approach was carried out. Two techniques for IOP measurements using the standard GAT and the NCT were compared. A total of 180 eyes from 90 patients were included in the study. RESULTS: The total mean IOP according to NCT measurements was 14.12 mmHg, and the total mean IOP according to the GAT was 12.98 mmHg; these values were significantly different (p=0.0018). When dividing the participants into three groups according to the measurement range obtained and comparing the mean NCT and GAT measurements in each group, in Group 1 (10–15 mmHg), no statistically significant difference was found between the means of the two tonometers (p=0.3100), a difference was observed between Group 2 (16–19 mmHg) and Group 3 (20 mmHg or more) (p<0.001). When dividing the participants by age group, the means obtained by the two tonometers also differed significantly between Group 4 (40–59 years) and Group 5 (60 years or more) (p<0.0001). In all groups, the mean measurements by the NCT were higher than those by the GAT. CONCLUSION: The NCT presented an approximate mean of the measures with the GAT in group 1 but was overestimated in the measurements of the groups 2 and 3. Dove 2021-02-05 /pmc/articles/PMC7872931/ /pubmed/33574653 http://dx.doi.org/10.2147/OPTH.S294710 Text en © 2021 Stock et al. http://creativecommons.org/licenses/by-nc/3.0/ This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms (https://www.dovepress.com/terms.php). |
spellingShingle | Original Research Stock, Ricardo Alexandre Ströher, Carine Sampaio, Rodrigo Rosa Mergener, Rafael André Bonamigo, Elcio Luiz A Comparative Study Between the Goldmann Applanation Tonometer and the Non-Contact Air-Puff Tonometer (Huvitz HNT 7000) in Normal Eyes |
title | A Comparative Study Between the Goldmann Applanation Tonometer and the Non-Contact Air-Puff Tonometer (Huvitz HNT 7000) in Normal Eyes |
title_full | A Comparative Study Between the Goldmann Applanation Tonometer and the Non-Contact Air-Puff Tonometer (Huvitz HNT 7000) in Normal Eyes |
title_fullStr | A Comparative Study Between the Goldmann Applanation Tonometer and the Non-Contact Air-Puff Tonometer (Huvitz HNT 7000) in Normal Eyes |
title_full_unstemmed | A Comparative Study Between the Goldmann Applanation Tonometer and the Non-Contact Air-Puff Tonometer (Huvitz HNT 7000) in Normal Eyes |
title_short | A Comparative Study Between the Goldmann Applanation Tonometer and the Non-Contact Air-Puff Tonometer (Huvitz HNT 7000) in Normal Eyes |
title_sort | comparative study between the goldmann applanation tonometer and the non-contact air-puff tonometer (huvitz hnt 7000) in normal eyes |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7872931/ https://www.ncbi.nlm.nih.gov/pubmed/33574653 http://dx.doi.org/10.2147/OPTH.S294710 |
work_keys_str_mv | AT stockricardoalexandre acomparativestudybetweenthegoldmannapplanationtonometerandthenoncontactairpufftonometerhuvitzhnt7000innormaleyes AT strohercarine acomparativestudybetweenthegoldmannapplanationtonometerandthenoncontactairpufftonometerhuvitzhnt7000innormaleyes AT sampaiorodrigorosa acomparativestudybetweenthegoldmannapplanationtonometerandthenoncontactairpufftonometerhuvitzhnt7000innormaleyes AT mergenerrafaelandre acomparativestudybetweenthegoldmannapplanationtonometerandthenoncontactairpufftonometerhuvitzhnt7000innormaleyes AT bonamigoelcioluiz acomparativestudybetweenthegoldmannapplanationtonometerandthenoncontactairpufftonometerhuvitzhnt7000innormaleyes AT stockricardoalexandre comparativestudybetweenthegoldmannapplanationtonometerandthenoncontactairpufftonometerhuvitzhnt7000innormaleyes AT strohercarine comparativestudybetweenthegoldmannapplanationtonometerandthenoncontactairpufftonometerhuvitzhnt7000innormaleyes AT sampaiorodrigorosa comparativestudybetweenthegoldmannapplanationtonometerandthenoncontactairpufftonometerhuvitzhnt7000innormaleyes AT mergenerrafaelandre comparativestudybetweenthegoldmannapplanationtonometerandthenoncontactairpufftonometerhuvitzhnt7000innormaleyes AT bonamigoelcioluiz comparativestudybetweenthegoldmannapplanationtonometerandthenoncontactairpufftonometerhuvitzhnt7000innormaleyes |