Digital Hemoglobinometers as Point-of-Care Testing Devices for Hemoglobin Estimation: A Validation Study from India
INTRODUCTION: India has launched Anemia Mukt Bharat (AMB) strategy in 2018, to deal with the high burden of anemia in the country. Point-of-care testing (POCT) of anemia using digital hemoglobinometers and treatment is one of the primary interventions under AMB. This study aimed to determine the dia...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7877427/ https://www.ncbi.nlm.nih.gov/pubmed/33623211 http://dx.doi.org/10.4103/ijcm.IJCM_558_19 |
_version_ | 1783650166841016320 |
---|---|
author | Yadav, Kapil Kant, Shashi Ramaswamy, Gomathi Ahamed, Farhad Vohra, Kashish |
author_facet | Yadav, Kapil Kant, Shashi Ramaswamy, Gomathi Ahamed, Farhad Vohra, Kashish |
author_sort | Yadav, Kapil |
collection | PubMed |
description | INTRODUCTION: India has launched Anemia Mukt Bharat (AMB) strategy in 2018, to deal with the high burden of anemia in the country. Point-of-care testing (POCT) of anemia using digital hemoglobinometers and treatment is one of the primary interventions under AMB. This study aimed to determine the diagnostic validity of digital hemoglobinometers (TrueHb and HemoCue 301) for screening of anemia compared to hematology analyzer. METHODOLOGY: A hospital-based, cross-sectional study was conducted among pregnant women attending antenatal clinics of a selected primary health center and subdivisional hospital of Haryana, India, during January 2019. Hemoglobin (Hb) levels of the pregnant women were estimated in digital hemoglobinometers using capillary blood samples and hematology analyzer using venous blood samples. Bias, limits of agreement (LOA), and validity of digital hemoglobinometers were assessed against a hematology analyzer. RESULTS: A total of 110 pregnant women were included. Bias (LOA) in Hb values estimated using digital hemoglobinometers was −0.09 g/dL (−1.97 to 1.80 ) for HemoCue301 and −0.04 g/dL (−1.69 to 1.60) for TrueHb compared to the hematology analyzer. HemoCue(®) 301 (sensitivity: 86% and specificity: 83%) had relatively higher sensitivity and specificity compared to TrueHb (sensitivity: 78.9% and specificity: 81%). CONCLUSIONS: Digital hemoglobinometers have high sensitivity and specificity. Thus, these can be a potential POCTs for screening of anemia in peripheral health facilities. Further studies are required to establish the validity of the digital hemoglobinometers at community settings. |
format | Online Article Text |
id | pubmed-7877427 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Wolters Kluwer - Medknow |
record_format | MEDLINE/PubMed |
spelling | pubmed-78774272021-02-22 Digital Hemoglobinometers as Point-of-Care Testing Devices for Hemoglobin Estimation: A Validation Study from India Yadav, Kapil Kant, Shashi Ramaswamy, Gomathi Ahamed, Farhad Vohra, Kashish Indian J Community Med Original Article INTRODUCTION: India has launched Anemia Mukt Bharat (AMB) strategy in 2018, to deal with the high burden of anemia in the country. Point-of-care testing (POCT) of anemia using digital hemoglobinometers and treatment is one of the primary interventions under AMB. This study aimed to determine the diagnostic validity of digital hemoglobinometers (TrueHb and HemoCue 301) for screening of anemia compared to hematology analyzer. METHODOLOGY: A hospital-based, cross-sectional study was conducted among pregnant women attending antenatal clinics of a selected primary health center and subdivisional hospital of Haryana, India, during January 2019. Hemoglobin (Hb) levels of the pregnant women were estimated in digital hemoglobinometers using capillary blood samples and hematology analyzer using venous blood samples. Bias, limits of agreement (LOA), and validity of digital hemoglobinometers were assessed against a hematology analyzer. RESULTS: A total of 110 pregnant women were included. Bias (LOA) in Hb values estimated using digital hemoglobinometers was −0.09 g/dL (−1.97 to 1.80 ) for HemoCue301 and −0.04 g/dL (−1.69 to 1.60) for TrueHb compared to the hematology analyzer. HemoCue(®) 301 (sensitivity: 86% and specificity: 83%) had relatively higher sensitivity and specificity compared to TrueHb (sensitivity: 78.9% and specificity: 81%). CONCLUSIONS: Digital hemoglobinometers have high sensitivity and specificity. Thus, these can be a potential POCTs for screening of anemia in peripheral health facilities. Further studies are required to establish the validity of the digital hemoglobinometers at community settings. Wolters Kluwer - Medknow 2020 2020-10-28 /pmc/articles/PMC7877427/ /pubmed/33623211 http://dx.doi.org/10.4103/ijcm.IJCM_558_19 Text en Copyright: © 2020 Indian Journal of Community Medicine http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Original Article Yadav, Kapil Kant, Shashi Ramaswamy, Gomathi Ahamed, Farhad Vohra, Kashish Digital Hemoglobinometers as Point-of-Care Testing Devices for Hemoglobin Estimation: A Validation Study from India |
title | Digital Hemoglobinometers as Point-of-Care Testing Devices for Hemoglobin Estimation: A Validation Study from India |
title_full | Digital Hemoglobinometers as Point-of-Care Testing Devices for Hemoglobin Estimation: A Validation Study from India |
title_fullStr | Digital Hemoglobinometers as Point-of-Care Testing Devices for Hemoglobin Estimation: A Validation Study from India |
title_full_unstemmed | Digital Hemoglobinometers as Point-of-Care Testing Devices for Hemoglobin Estimation: A Validation Study from India |
title_short | Digital Hemoglobinometers as Point-of-Care Testing Devices for Hemoglobin Estimation: A Validation Study from India |
title_sort | digital hemoglobinometers as point-of-care testing devices for hemoglobin estimation: a validation study from india |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7877427/ https://www.ncbi.nlm.nih.gov/pubmed/33623211 http://dx.doi.org/10.4103/ijcm.IJCM_558_19 |
work_keys_str_mv | AT yadavkapil digitalhemoglobinometersaspointofcaretestingdevicesforhemoglobinestimationavalidationstudyfromindia AT kantshashi digitalhemoglobinometersaspointofcaretestingdevicesforhemoglobinestimationavalidationstudyfromindia AT ramaswamygomathi digitalhemoglobinometersaspointofcaretestingdevicesforhemoglobinestimationavalidationstudyfromindia AT ahamedfarhad digitalhemoglobinometersaspointofcaretestingdevicesforhemoglobinestimationavalidationstudyfromindia AT vohrakashish digitalhemoglobinometersaspointofcaretestingdevicesforhemoglobinestimationavalidationstudyfromindia |