Measurement of T1 Mapping in Patients With Cardiac Devices: Off-Resonance Error Extends Beyond Visual Artifact but Can Be Quantified and Corrected

Background: Measurement of myocardial T1 is increasingly incorporated into standard cardiovascular magnetic resonance (CMR) protocols, however accuracy may be reduced in patients with metallic cardiovascular implants. Measurement is feasible in segments free from visual artifact, but there may still...

Descripción completa

Detalles Bibliográficos
Autores principales: Bhuva, Anish N., Treibel, Thomas A., Seraphim, Andreas, Scully, Paul, Knott, Kristopher D., Augusto, João B., Torlasco, Camilla, Menacho, Katia, Lau, Clement, Patel, Kush, Moon, James C., Kellman, Peter, Manisty, Charlotte H.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7878555/
https://www.ncbi.nlm.nih.gov/pubmed/33585589
http://dx.doi.org/10.3389/fcvm.2021.631366
_version_ 1783650360288608256
author Bhuva, Anish N.
Treibel, Thomas A.
Seraphim, Andreas
Scully, Paul
Knott, Kristopher D.
Augusto, João B.
Torlasco, Camilla
Menacho, Katia
Lau, Clement
Patel, Kush
Moon, James C.
Kellman, Peter
Manisty, Charlotte H.
author_facet Bhuva, Anish N.
Treibel, Thomas A.
Seraphim, Andreas
Scully, Paul
Knott, Kristopher D.
Augusto, João B.
Torlasco, Camilla
Menacho, Katia
Lau, Clement
Patel, Kush
Moon, James C.
Kellman, Peter
Manisty, Charlotte H.
author_sort Bhuva, Anish N.
collection PubMed
description Background: Measurement of myocardial T1 is increasingly incorporated into standard cardiovascular magnetic resonance (CMR) protocols, however accuracy may be reduced in patients with metallic cardiovascular implants. Measurement is feasible in segments free from visual artifact, but there may still be off-resonance induced error. Aim: To quantify off-resonance induced T1 error in patients with metallic cardiovascular implants, and validate a method for error correction for a conventional MOLLI pulse sequence. Methods: Twenty-four patients with cardiac implantable electronic devices (CIEDs: 46% permanent pacemakers, PPMs; 33% implantable loop recorders, ILRs; and 21% implantable cardioverter-defibrillators, ICDs); and 31 patients with aortic valve replacement (AVR) (45% metallic) were studied. Paired mid-myocardial short-axis MOLLI and single breath-hold off-resonance field maps were acquired at 1.5 T. T1 values were measured by AHA segment, and segments with visual artifact were excluded. T1 correction was applied using a published relationship between off-resonance and T1. The accuracy of the correction was assessed in 10 healthy volunteers by measuring T1 before and after external placement of an ICD generator next to the chest to generate off-resonance. Results: T1 values in healthy volunteers with an ICD were underestimated compared to without (967 ± 52 vs. 997 ± 26 ms respectively, p = 0.0001), but were similar after correction (p = 0.57, residual difference 2 ± 27 ms). Artifact was visible in 4 ± 12, 42 ± 31, and 53 ± 27% of AHA segments in patients with ILRs, PPMs, and ICDs, respectively. In segments without artifact, T1 was underestimated by 63 ms (interquartile range: 7–143) per patient. The greatest error for patients with ILRs, PPMs and ICDs were 79, 146, and 191 ms, respectively. The presence of an AVR did not generate T1 error. Conclusion: Even when there is no visual artifact, there is error in T1 in patients with CIEDs, but not AVRs. Off-resonance field map acquisition can detect error in measured T1, and a correction can be applied to quantify T1 MOLLI accurately.
format Online
Article
Text
id pubmed-7878555
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-78785552021-02-13 Measurement of T1 Mapping in Patients With Cardiac Devices: Off-Resonance Error Extends Beyond Visual Artifact but Can Be Quantified and Corrected Bhuva, Anish N. Treibel, Thomas A. Seraphim, Andreas Scully, Paul Knott, Kristopher D. Augusto, João B. Torlasco, Camilla Menacho, Katia Lau, Clement Patel, Kush Moon, James C. Kellman, Peter Manisty, Charlotte H. Front Cardiovasc Med Cardiovascular Medicine Background: Measurement of myocardial T1 is increasingly incorporated into standard cardiovascular magnetic resonance (CMR) protocols, however accuracy may be reduced in patients with metallic cardiovascular implants. Measurement is feasible in segments free from visual artifact, but there may still be off-resonance induced error. Aim: To quantify off-resonance induced T1 error in patients with metallic cardiovascular implants, and validate a method for error correction for a conventional MOLLI pulse sequence. Methods: Twenty-four patients with cardiac implantable electronic devices (CIEDs: 46% permanent pacemakers, PPMs; 33% implantable loop recorders, ILRs; and 21% implantable cardioverter-defibrillators, ICDs); and 31 patients with aortic valve replacement (AVR) (45% metallic) were studied. Paired mid-myocardial short-axis MOLLI and single breath-hold off-resonance field maps were acquired at 1.5 T. T1 values were measured by AHA segment, and segments with visual artifact were excluded. T1 correction was applied using a published relationship between off-resonance and T1. The accuracy of the correction was assessed in 10 healthy volunteers by measuring T1 before and after external placement of an ICD generator next to the chest to generate off-resonance. Results: T1 values in healthy volunteers with an ICD were underestimated compared to without (967 ± 52 vs. 997 ± 26 ms respectively, p = 0.0001), but were similar after correction (p = 0.57, residual difference 2 ± 27 ms). Artifact was visible in 4 ± 12, 42 ± 31, and 53 ± 27% of AHA segments in patients with ILRs, PPMs, and ICDs, respectively. In segments without artifact, T1 was underestimated by 63 ms (interquartile range: 7–143) per patient. The greatest error for patients with ILRs, PPMs and ICDs were 79, 146, and 191 ms, respectively. The presence of an AVR did not generate T1 error. Conclusion: Even when there is no visual artifact, there is error in T1 in patients with CIEDs, but not AVRs. Off-resonance field map acquisition can detect error in measured T1, and a correction can be applied to quantify T1 MOLLI accurately. Frontiers Media S.A. 2021-01-29 /pmc/articles/PMC7878555/ /pubmed/33585589 http://dx.doi.org/10.3389/fcvm.2021.631366 Text en Copyright © 2021 Bhuva, Treibel, Seraphim, Scully, Knott, Augusto, Torlasco, Menacho, Lau, Patel, Moon, Kellman and Manisty. http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Cardiovascular Medicine
Bhuva, Anish N.
Treibel, Thomas A.
Seraphim, Andreas
Scully, Paul
Knott, Kristopher D.
Augusto, João B.
Torlasco, Camilla
Menacho, Katia
Lau, Clement
Patel, Kush
Moon, James C.
Kellman, Peter
Manisty, Charlotte H.
Measurement of T1 Mapping in Patients With Cardiac Devices: Off-Resonance Error Extends Beyond Visual Artifact but Can Be Quantified and Corrected
title Measurement of T1 Mapping in Patients With Cardiac Devices: Off-Resonance Error Extends Beyond Visual Artifact but Can Be Quantified and Corrected
title_full Measurement of T1 Mapping in Patients With Cardiac Devices: Off-Resonance Error Extends Beyond Visual Artifact but Can Be Quantified and Corrected
title_fullStr Measurement of T1 Mapping in Patients With Cardiac Devices: Off-Resonance Error Extends Beyond Visual Artifact but Can Be Quantified and Corrected
title_full_unstemmed Measurement of T1 Mapping in Patients With Cardiac Devices: Off-Resonance Error Extends Beyond Visual Artifact but Can Be Quantified and Corrected
title_short Measurement of T1 Mapping in Patients With Cardiac Devices: Off-Resonance Error Extends Beyond Visual Artifact but Can Be Quantified and Corrected
title_sort measurement of t1 mapping in patients with cardiac devices: off-resonance error extends beyond visual artifact but can be quantified and corrected
topic Cardiovascular Medicine
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7878555/
https://www.ncbi.nlm.nih.gov/pubmed/33585589
http://dx.doi.org/10.3389/fcvm.2021.631366
work_keys_str_mv AT bhuvaanishn measurementoft1mappinginpatientswithcardiacdevicesoffresonanceerrorextendsbeyondvisualartifactbutcanbequantifiedandcorrected
AT treibelthomasa measurementoft1mappinginpatientswithcardiacdevicesoffresonanceerrorextendsbeyondvisualartifactbutcanbequantifiedandcorrected
AT seraphimandreas measurementoft1mappinginpatientswithcardiacdevicesoffresonanceerrorextendsbeyondvisualartifactbutcanbequantifiedandcorrected
AT scullypaul measurementoft1mappinginpatientswithcardiacdevicesoffresonanceerrorextendsbeyondvisualartifactbutcanbequantifiedandcorrected
AT knottkristopherd measurementoft1mappinginpatientswithcardiacdevicesoffresonanceerrorextendsbeyondvisualartifactbutcanbequantifiedandcorrected
AT augustojoaob measurementoft1mappinginpatientswithcardiacdevicesoffresonanceerrorextendsbeyondvisualartifactbutcanbequantifiedandcorrected
AT torlascocamilla measurementoft1mappinginpatientswithcardiacdevicesoffresonanceerrorextendsbeyondvisualartifactbutcanbequantifiedandcorrected
AT menachokatia measurementoft1mappinginpatientswithcardiacdevicesoffresonanceerrorextendsbeyondvisualartifactbutcanbequantifiedandcorrected
AT lauclement measurementoft1mappinginpatientswithcardiacdevicesoffresonanceerrorextendsbeyondvisualartifactbutcanbequantifiedandcorrected
AT patelkush measurementoft1mappinginpatientswithcardiacdevicesoffresonanceerrorextendsbeyondvisualartifactbutcanbequantifiedandcorrected
AT moonjamesc measurementoft1mappinginpatientswithcardiacdevicesoffresonanceerrorextendsbeyondvisualartifactbutcanbequantifiedandcorrected
AT kellmanpeter measurementoft1mappinginpatientswithcardiacdevicesoffresonanceerrorextendsbeyondvisualartifactbutcanbequantifiedandcorrected
AT manistycharlotteh measurementoft1mappinginpatientswithcardiacdevicesoffresonanceerrorextendsbeyondvisualartifactbutcanbequantifiedandcorrected