Implant-Free Subpectoral Biceps Tenodesis Is Biomechanically at Higher Risk of Spiral Fracture of the Humerus Compared With Implant-Free Suprapectoral Biceps Tenodesis

PURPOSE: To compare the likelihood of spiral fracture of the humerus using torsional load to failure after intraosseous biceps tenodesis at the position of the arthroscopic suprapectoral tenodesis versus the subpectoral meta-diaphyseal location. METHODS: Eight matched pairs of humeri were dissected....

Descripción completa

Detalles Bibliográficos
Autores principales: Dini, Arash A., Mizels, Joshua E., Sadeghpour, Sohale, O’Brien, Michael J., Savoie, Felix H., Getelman, Mark H.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7879170/
https://www.ncbi.nlm.nih.gov/pubmed/33615250
http://dx.doi.org/10.1016/j.asmr.2020.08.011
_version_ 1783650477358972928
author Dini, Arash A.
Mizels, Joshua E.
Sadeghpour, Sohale
O’Brien, Michael J.
Savoie, Felix H.
Getelman, Mark H.
author_facet Dini, Arash A.
Mizels, Joshua E.
Sadeghpour, Sohale
O’Brien, Michael J.
Savoie, Felix H.
Getelman, Mark H.
author_sort Dini, Arash A.
collection PubMed
description PURPOSE: To compare the likelihood of spiral fracture of the humerus using torsional load to failure after intraosseous biceps tenodesis at the position of the arthroscopic suprapectoral tenodesis versus the subpectoral meta-diaphyseal location. METHODS: Eight matched pairs of humeri were dissected. Unicortical tenodesis holes were drilled, either at the bottom of the bicipital groove (group 1) or just below the pectoralis major tendon insertion (subpectoral) in the humeral diaphysis (group 2). Tenodesis was performed in a 7-mm bone tunnel, with suture fixation distal to this site using 2 separate 2-mm holes, secured with No. 2 polyester suture. Each humerus was potted in plaster and mounted to a hydraulic torsional load frame, consistent with previously validated models for creating humeral spiral fractures. External rotation torque was applied to each humerus distally until fracture occurred. The paired t test was used to compare the 2 groups. RESULTS: Fracture occurred at the subpectoral cortical drill hole in all 8 specimens in group 2. In group 1, only 2 fractures occurred through the tenodesis hole, with spiral fracture resulting in the diaphysis of the humerus in 6 of 8 specimens. Average torque to failure measured 31.35 Nm in group 1 and 25.08 Nm in group 2; the difference was statistically significant (P < .0001). CONCLUSIONS: Subpectoral cortical drill holes for biceps tenodesis were shown to be a stress riser for humeral spiral fracture. Suprapectoral cortical drill holes were shown to be significantly less of a stress riser. The amount of torque required to fracture the humerus through the subpectoral drill holes was less than with the suprapectoral drill holes. Only 2 fractures occurred through the suprapectoral tenodesis holes, and significantly more torque was required to create these fractures. CLINICAL RELEVANCE: Clinically, the difference between suprapectoral and subpectoral tenodesis fracture potential should be considered when selecting a tenodesis location.
format Online
Article
Text
id pubmed-7879170
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-78791702021-02-18 Implant-Free Subpectoral Biceps Tenodesis Is Biomechanically at Higher Risk of Spiral Fracture of the Humerus Compared With Implant-Free Suprapectoral Biceps Tenodesis Dini, Arash A. Mizels, Joshua E. Sadeghpour, Sohale O’Brien, Michael J. Savoie, Felix H. Getelman, Mark H. Arthrosc Sports Med Rehabil Original Article PURPOSE: To compare the likelihood of spiral fracture of the humerus using torsional load to failure after intraosseous biceps tenodesis at the position of the arthroscopic suprapectoral tenodesis versus the subpectoral meta-diaphyseal location. METHODS: Eight matched pairs of humeri were dissected. Unicortical tenodesis holes were drilled, either at the bottom of the bicipital groove (group 1) or just below the pectoralis major tendon insertion (subpectoral) in the humeral diaphysis (group 2). Tenodesis was performed in a 7-mm bone tunnel, with suture fixation distal to this site using 2 separate 2-mm holes, secured with No. 2 polyester suture. Each humerus was potted in plaster and mounted to a hydraulic torsional load frame, consistent with previously validated models for creating humeral spiral fractures. External rotation torque was applied to each humerus distally until fracture occurred. The paired t test was used to compare the 2 groups. RESULTS: Fracture occurred at the subpectoral cortical drill hole in all 8 specimens in group 2. In group 1, only 2 fractures occurred through the tenodesis hole, with spiral fracture resulting in the diaphysis of the humerus in 6 of 8 specimens. Average torque to failure measured 31.35 Nm in group 1 and 25.08 Nm in group 2; the difference was statistically significant (P < .0001). CONCLUSIONS: Subpectoral cortical drill holes for biceps tenodesis were shown to be a stress riser for humeral spiral fracture. Suprapectoral cortical drill holes were shown to be significantly less of a stress riser. The amount of torque required to fracture the humerus through the subpectoral drill holes was less than with the suprapectoral drill holes. Only 2 fractures occurred through the suprapectoral tenodesis holes, and significantly more torque was required to create these fractures. CLINICAL RELEVANCE: Clinically, the difference between suprapectoral and subpectoral tenodesis fracture potential should be considered when selecting a tenodesis location. Elsevier 2020-12-26 /pmc/articles/PMC7879170/ /pubmed/33615250 http://dx.doi.org/10.1016/j.asmr.2020.08.011 Text en © 2020 The Authors https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Original Article
Dini, Arash A.
Mizels, Joshua E.
Sadeghpour, Sohale
O’Brien, Michael J.
Savoie, Felix H.
Getelman, Mark H.
Implant-Free Subpectoral Biceps Tenodesis Is Biomechanically at Higher Risk of Spiral Fracture of the Humerus Compared With Implant-Free Suprapectoral Biceps Tenodesis
title Implant-Free Subpectoral Biceps Tenodesis Is Biomechanically at Higher Risk of Spiral Fracture of the Humerus Compared With Implant-Free Suprapectoral Biceps Tenodesis
title_full Implant-Free Subpectoral Biceps Tenodesis Is Biomechanically at Higher Risk of Spiral Fracture of the Humerus Compared With Implant-Free Suprapectoral Biceps Tenodesis
title_fullStr Implant-Free Subpectoral Biceps Tenodesis Is Biomechanically at Higher Risk of Spiral Fracture of the Humerus Compared With Implant-Free Suprapectoral Biceps Tenodesis
title_full_unstemmed Implant-Free Subpectoral Biceps Tenodesis Is Biomechanically at Higher Risk of Spiral Fracture of the Humerus Compared With Implant-Free Suprapectoral Biceps Tenodesis
title_short Implant-Free Subpectoral Biceps Tenodesis Is Biomechanically at Higher Risk of Spiral Fracture of the Humerus Compared With Implant-Free Suprapectoral Biceps Tenodesis
title_sort implant-free subpectoral biceps tenodesis is biomechanically at higher risk of spiral fracture of the humerus compared with implant-free suprapectoral biceps tenodesis
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7879170/
https://www.ncbi.nlm.nih.gov/pubmed/33615250
http://dx.doi.org/10.1016/j.asmr.2020.08.011
work_keys_str_mv AT diniarasha implantfreesubpectoralbicepstenodesisisbiomechanicallyathigherriskofspiralfractureofthehumeruscomparedwithimplantfreesuprapectoralbicepstenodesis
AT mizelsjoshuae implantfreesubpectoralbicepstenodesisisbiomechanicallyathigherriskofspiralfractureofthehumeruscomparedwithimplantfreesuprapectoralbicepstenodesis
AT sadeghpoursohale implantfreesubpectoralbicepstenodesisisbiomechanicallyathigherriskofspiralfractureofthehumeruscomparedwithimplantfreesuprapectoralbicepstenodesis
AT obrienmichaelj implantfreesubpectoralbicepstenodesisisbiomechanicallyathigherriskofspiralfractureofthehumeruscomparedwithimplantfreesuprapectoralbicepstenodesis
AT savoiefelixh implantfreesubpectoralbicepstenodesisisbiomechanicallyathigherriskofspiralfractureofthehumeruscomparedwithimplantfreesuprapectoralbicepstenodesis
AT getelmanmarkh implantfreesubpectoralbicepstenodesisisbiomechanicallyathigherriskofspiralfractureofthehumeruscomparedwithimplantfreesuprapectoralbicepstenodesis