Evaluation of diffusion‐weighted MRI and geometric distortion on a 0.35T MR‐LINAC at multiple gantry angles

Diffusion‐weighted imaging (DWI) provides a valuable diagnostic tool for tumor evaluation. Yet, it is difficult to acquire daily MRI data sets in the traditional radiotherapy clinical setting due to patient burden and limited resources. However, integrated MRI radiotherapy treatment systems facilita...

Descripción completa

Detalles Bibliográficos
Autores principales: Lewis, Benjamin, Guta, Anamaria, Mackey, Stacie, Gach, H. Michael, Mutic, Sasa, Green, Olga, Kim, Taeho
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7882099/
https://www.ncbi.nlm.nih.gov/pubmed/33450146
http://dx.doi.org/10.1002/acm2.13135
_version_ 1783650992316743680
author Lewis, Benjamin
Guta, Anamaria
Mackey, Stacie
Gach, H. Michael
Mutic, Sasa
Green, Olga
Kim, Taeho
author_facet Lewis, Benjamin
Guta, Anamaria
Mackey, Stacie
Gach, H. Michael
Mutic, Sasa
Green, Olga
Kim, Taeho
author_sort Lewis, Benjamin
collection PubMed
description Diffusion‐weighted imaging (DWI) provides a valuable diagnostic tool for tumor evaluation. Yet, it is difficult to acquire daily MRI data sets in the traditional radiotherapy clinical setting due to patient burden and limited resources. However, integrated MRI radiotherapy treatment systems facilitate daily functional MRI acquisitions like DWI during treatment exams. Before ADC values from MR‐RT systems can be used clinically their reproducibility and accuracy must be quantified. This study used a NIST traceable DWI phantom to verify ADC values acquired on a 0.35 T MR‐LINAC system at multiple gantry angles. A diffusion‐weighted echo planar imaging sequence was used for all image acquisitions, with b‐values of 0, 500, 900, 2000 s/mm(2) for the 1.5 T and 3.0 T systems and 0, 200, 500, 800 s/mm(2) for the 0.35 T system. Images were acquired at multiple gantry angles on the MR‐LINAC system from 0° to 330° in 30° increments to assess the impact of gantry angle on geometric distortion and ADC values. CT images, and three fiducial markers were used as ground truth for geometric distortion measurements. The distance between fiducial markers increased by as much as 7.2 mm on the MR‐LINAC at gantry angle 60°. ADC values of deionized water vials from the 1.5 T and 3.0 T systems were 8.30 × 10(‐6) mm(2)/s and −0.85 × 10(‐6) mm(2)/s off, respectively, from the expected value of 1127 × 10(‐6) mm(2)/s. The MR‐LINAC system provided an ADC value of the pure water vials that was −116.63 × 10(‐6) mm(2)/s off from the expected value of 1127 × 10(‐6) mm(2)/s. The MR‐LINAC also showed a variation in ADC across all gantry angles of 33.72 × 10(‐6) mm(2)/s and 20.41 × 10(‐6) mm(2)/s for the vials with expected values of 1127 × 10(‐6) mm(2)/s and 248 × 10(‐6) mm(2)/s, respectively. This study showed that variation of the ADC values and geometric information on the 0.35 T MR‐LINAC system was dependent on the gantry angle at acquisition.
format Online
Article
Text
id pubmed-7882099
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-78820992021-02-19 Evaluation of diffusion‐weighted MRI and geometric distortion on a 0.35T MR‐LINAC at multiple gantry angles Lewis, Benjamin Guta, Anamaria Mackey, Stacie Gach, H. Michael Mutic, Sasa Green, Olga Kim, Taeho J Appl Clin Med Phys Medical Imaging Diffusion‐weighted imaging (DWI) provides a valuable diagnostic tool for tumor evaluation. Yet, it is difficult to acquire daily MRI data sets in the traditional radiotherapy clinical setting due to patient burden and limited resources. However, integrated MRI radiotherapy treatment systems facilitate daily functional MRI acquisitions like DWI during treatment exams. Before ADC values from MR‐RT systems can be used clinically their reproducibility and accuracy must be quantified. This study used a NIST traceable DWI phantom to verify ADC values acquired on a 0.35 T MR‐LINAC system at multiple gantry angles. A diffusion‐weighted echo planar imaging sequence was used for all image acquisitions, with b‐values of 0, 500, 900, 2000 s/mm(2) for the 1.5 T and 3.0 T systems and 0, 200, 500, 800 s/mm(2) for the 0.35 T system. Images were acquired at multiple gantry angles on the MR‐LINAC system from 0° to 330° in 30° increments to assess the impact of gantry angle on geometric distortion and ADC values. CT images, and three fiducial markers were used as ground truth for geometric distortion measurements. The distance between fiducial markers increased by as much as 7.2 mm on the MR‐LINAC at gantry angle 60°. ADC values of deionized water vials from the 1.5 T and 3.0 T systems were 8.30 × 10(‐6) mm(2)/s and −0.85 × 10(‐6) mm(2)/s off, respectively, from the expected value of 1127 × 10(‐6) mm(2)/s. The MR‐LINAC system provided an ADC value of the pure water vials that was −116.63 × 10(‐6) mm(2)/s off from the expected value of 1127 × 10(‐6) mm(2)/s. The MR‐LINAC also showed a variation in ADC across all gantry angles of 33.72 × 10(‐6) mm(2)/s and 20.41 × 10(‐6) mm(2)/s for the vials with expected values of 1127 × 10(‐6) mm(2)/s and 248 × 10(‐6) mm(2)/s, respectively. This study showed that variation of the ADC values and geometric information on the 0.35 T MR‐LINAC system was dependent on the gantry angle at acquisition. John Wiley and Sons Inc. 2021-01-15 /pmc/articles/PMC7882099/ /pubmed/33450146 http://dx.doi.org/10.1002/acm2.13135 Text en © 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Imaging
Lewis, Benjamin
Guta, Anamaria
Mackey, Stacie
Gach, H. Michael
Mutic, Sasa
Green, Olga
Kim, Taeho
Evaluation of diffusion‐weighted MRI and geometric distortion on a 0.35T MR‐LINAC at multiple gantry angles
title Evaluation of diffusion‐weighted MRI and geometric distortion on a 0.35T MR‐LINAC at multiple gantry angles
title_full Evaluation of diffusion‐weighted MRI and geometric distortion on a 0.35T MR‐LINAC at multiple gantry angles
title_fullStr Evaluation of diffusion‐weighted MRI and geometric distortion on a 0.35T MR‐LINAC at multiple gantry angles
title_full_unstemmed Evaluation of diffusion‐weighted MRI and geometric distortion on a 0.35T MR‐LINAC at multiple gantry angles
title_short Evaluation of diffusion‐weighted MRI and geometric distortion on a 0.35T MR‐LINAC at multiple gantry angles
title_sort evaluation of diffusion‐weighted mri and geometric distortion on a 0.35t mr‐linac at multiple gantry angles
topic Medical Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7882099/
https://www.ncbi.nlm.nih.gov/pubmed/33450146
http://dx.doi.org/10.1002/acm2.13135
work_keys_str_mv AT lewisbenjamin evaluationofdiffusionweightedmriandgeometricdistortionona035tmrlinacatmultiplegantryangles
AT gutaanamaria evaluationofdiffusionweightedmriandgeometricdistortionona035tmrlinacatmultiplegantryangles
AT mackeystacie evaluationofdiffusionweightedmriandgeometricdistortionona035tmrlinacatmultiplegantryangles
AT gachhmichael evaluationofdiffusionweightedmriandgeometricdistortionona035tmrlinacatmultiplegantryangles
AT muticsasa evaluationofdiffusionweightedmriandgeometricdistortionona035tmrlinacatmultiplegantryangles
AT greenolga evaluationofdiffusionweightedmriandgeometricdistortionona035tmrlinacatmultiplegantryangles
AT kimtaeho evaluationofdiffusionweightedmriandgeometricdistortionona035tmrlinacatmultiplegantryangles