Standardized fluoroscopy-guided implantation technique enables optimal electrode placement in sacral neuromodulation: a cadaver study

BACKGROUND: Sacral neuromodulation (SNM) is an established treatment option for patients with faecal incontinence. The location of the stimulating electrode is considered to be essential for treatment success. The purpose of this study was to evaluate the position of SNM electrodes after using a sta...

Descripción completa

Detalles Bibliográficos
Autores principales: Müller, C., Reissig, L. F., Argeny, S., Weninger, W. J., Riss, S.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7884365/
https://www.ncbi.nlm.nih.gov/pubmed/33211205
http://dx.doi.org/10.1007/s10151-020-02364-w
_version_ 1783651399493484544
author Müller, C.
Reissig, L. F.
Argeny, S.
Weninger, W. J.
Riss, S.
author_facet Müller, C.
Reissig, L. F.
Argeny, S.
Weninger, W. J.
Riss, S.
author_sort Müller, C.
collection PubMed
description BACKGROUND: Sacral neuromodulation (SNM) is an established treatment option for patients with faecal incontinence. The location of the stimulating electrode is considered to be essential for treatment success. The purpose of this study was to evaluate the position of SNM electrodes after using a standardized fluoroscopy-guided implantation technique. METHODS: For this cadaver study, SNM electrodes were implanted bilaterally in 5 lower body specimens. The lower edge of the sacroiliac joint and the medial edge of the sacral foramina were marked using fluoroscopy to draw an ‘H’ with the crossing points identifying S3. After electrode placement the pelvis was dissected to describe the exact position of the SNM electrodes. RESULTS: The electrodes were inserted at an angle with a median degree measure of 60° (range 50–65°) to the skin, with a median distance of 9 mm (range 0–13 mm) from the S3 marking. All electrodes entered the third sacral foramen. The median distance of the electrodes to the sacral nerve was 0 mm (range 0–3 mm) for the most proximal, 0.5 mm (range 0–5 mm) for the second, 2.25 mm (range 0–11 mm) for the third and 1.75 mm (range 0–16 mm) for the most distant electrode. There was neither a significant difference in the proximity of the electrodes to the nerve between the right and left side (proximal to distal electrode: p = 0.18, p = 0.16, p = 0.07, p = 0.07) nor between male and female cadavers (p = 0.25, p = 0.21, p = 0.66, p = 0.66). CONCLUSIONS: A standardized fluoroscopy-guided implantation technique enables a close contact between electrode and nerve. This can potentially result in an improved clinical outcome.
format Online
Article
Text
id pubmed-7884365
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-78843652021-02-25 Standardized fluoroscopy-guided implantation technique enables optimal electrode placement in sacral neuromodulation: a cadaver study Müller, C. Reissig, L. F. Argeny, S. Weninger, W. J. Riss, S. Tech Coloproctol Original Article BACKGROUND: Sacral neuromodulation (SNM) is an established treatment option for patients with faecal incontinence. The location of the stimulating electrode is considered to be essential for treatment success. The purpose of this study was to evaluate the position of SNM electrodes after using a standardized fluoroscopy-guided implantation technique. METHODS: For this cadaver study, SNM electrodes were implanted bilaterally in 5 lower body specimens. The lower edge of the sacroiliac joint and the medial edge of the sacral foramina were marked using fluoroscopy to draw an ‘H’ with the crossing points identifying S3. After electrode placement the pelvis was dissected to describe the exact position of the SNM electrodes. RESULTS: The electrodes were inserted at an angle with a median degree measure of 60° (range 50–65°) to the skin, with a median distance of 9 mm (range 0–13 mm) from the S3 marking. All electrodes entered the third sacral foramen. The median distance of the electrodes to the sacral nerve was 0 mm (range 0–3 mm) for the most proximal, 0.5 mm (range 0–5 mm) for the second, 2.25 mm (range 0–11 mm) for the third and 1.75 mm (range 0–16 mm) for the most distant electrode. There was neither a significant difference in the proximity of the electrodes to the nerve between the right and left side (proximal to distal electrode: p = 0.18, p = 0.16, p = 0.07, p = 0.07) nor between male and female cadavers (p = 0.25, p = 0.21, p = 0.66, p = 0.66). CONCLUSIONS: A standardized fluoroscopy-guided implantation technique enables a close contact between electrode and nerve. This can potentially result in an improved clinical outcome. Springer International Publishing 2020-11-19 2021 /pmc/articles/PMC7884365/ /pubmed/33211205 http://dx.doi.org/10.1007/s10151-020-02364-w Text en © The Author(s) 2020 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
spellingShingle Original Article
Müller, C.
Reissig, L. F.
Argeny, S.
Weninger, W. J.
Riss, S.
Standardized fluoroscopy-guided implantation technique enables optimal electrode placement in sacral neuromodulation: a cadaver study
title Standardized fluoroscopy-guided implantation technique enables optimal electrode placement in sacral neuromodulation: a cadaver study
title_full Standardized fluoroscopy-guided implantation technique enables optimal electrode placement in sacral neuromodulation: a cadaver study
title_fullStr Standardized fluoroscopy-guided implantation technique enables optimal electrode placement in sacral neuromodulation: a cadaver study
title_full_unstemmed Standardized fluoroscopy-guided implantation technique enables optimal electrode placement in sacral neuromodulation: a cadaver study
title_short Standardized fluoroscopy-guided implantation technique enables optimal electrode placement in sacral neuromodulation: a cadaver study
title_sort standardized fluoroscopy-guided implantation technique enables optimal electrode placement in sacral neuromodulation: a cadaver study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7884365/
https://www.ncbi.nlm.nih.gov/pubmed/33211205
http://dx.doi.org/10.1007/s10151-020-02364-w
work_keys_str_mv AT mullerc standardizedfluoroscopyguidedimplantationtechniqueenablesoptimalelectrodeplacementinsacralneuromodulationacadaverstudy
AT reissiglf standardizedfluoroscopyguidedimplantationtechniqueenablesoptimalelectrodeplacementinsacralneuromodulationacadaverstudy
AT argenys standardizedfluoroscopyguidedimplantationtechniqueenablesoptimalelectrodeplacementinsacralneuromodulationacadaverstudy
AT weningerwj standardizedfluoroscopyguidedimplantationtechniqueenablesoptimalelectrodeplacementinsacralneuromodulationacadaverstudy
AT risss standardizedfluoroscopyguidedimplantationtechniqueenablesoptimalelectrodeplacementinsacralneuromodulationacadaverstudy