Good mid-term results with the trident peripheral self-locking cup: a clinical evaluation and migration measurement with EBRA

INTRODUCTION: The most common cause of failure in total hip arthroplasty (THA) is aseptic loosening. Uncemented cup migration analysis by means of Einzel–Bild–Roentgen–Analyse (EBRA) has shown to be a good predictive indicator for early implant failure if the cup migrates more than 2 mm within 4 yea...

Descripción completa

Detalles Bibliográficos
Autores principales: Dammerer, Dietmar, Blum, Philipp, Putzer, David, Tscholl, Andreas, Liebensteiner, Michael C., Thaler, Martin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7886724/
https://www.ncbi.nlm.nih.gov/pubmed/33164141
http://dx.doi.org/10.1007/s00402-020-03639-5
Descripción
Sumario:INTRODUCTION: The most common cause of failure in total hip arthroplasty (THA) is aseptic loosening. Uncemented cup migration analysis by means of Einzel–Bild–Roentgen–Analyse (EBRA) has shown to be a good predictive indicator for early implant failure if the cup migrates more than 2 mm within 4 years after surgery. In this study, we performed a migration analysis of an uncemented peripheral self-locking (PSL) press-fit cup after 4 years follow-up. MATERIALS AND METHODS: We retrospectively reviewed all patients who received a trident PSL press-fit cup at our department between 2004 and 2017. A total of 636 patients were identified. As inclusion criteria for radiological analysis, a minimum follow-up of 2 years was defined. We reviewed medical histories and performed radiological analysis using EBRA software. EBRA measurements and statistical investigations were performed by two independent investigators. RESULTS: A total of 149 cups in 146 patients (female 82; male 64) met our inclusion criteria. Mean age at surgery was 65 years (33–89). We found a significant improvement in the WOMAC score pre- to postoperative (p < 0.0001). EBRA migration analysis showed a mean total migration of 0.6 mm (0.0–8.2) over our follow-up period of 4 years. Of the investigated cups, 69.8% showed a migration rate smaller than 2 mm in the investigated follow-up. CONCLUSION: The acetabular cup used in our study provides low migration at final follow-up. Therefore, a good long-term outcome can be expected for the PSL cup. TRIAL REGISTRATION: Trial registration number is 20181024-1875 and date of registration is 2018-10-24.