High intra- and inter-observer reliability of planning implant size in MRI-based patient-specific instrumentation for total knee arthroplasty

PURPOSE: Patient-specific instrumentation (PSI) in total knee arthroplasty (TKA) uses individually designed disposable guides to determine intraoperative bone cuts. The manufacturer provides the surgeon with proposed planning which can be modified and should be approved by the surgeon before the gui...

Descripción completa

Detalles Bibliográficos
Autores principales: Schoenmakers, Daphne A. L., Theeuwen, Dieuwertje M. J., Schotanus, Martijn G. M., Jansen, Edwin J. P., van Haaren, Emil H., Hendrickx, Roel P. M., Kort, Nanne P.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7892674/
https://www.ncbi.nlm.nih.gov/pubmed/32232539
http://dx.doi.org/10.1007/s00167-020-05946-1
_version_ 1783652894734548992
author Schoenmakers, Daphne A. L.
Theeuwen, Dieuwertje M. J.
Schotanus, Martijn G. M.
Jansen, Edwin J. P.
van Haaren, Emil H.
Hendrickx, Roel P. M.
Kort, Nanne P.
author_facet Schoenmakers, Daphne A. L.
Theeuwen, Dieuwertje M. J.
Schotanus, Martijn G. M.
Jansen, Edwin J. P.
van Haaren, Emil H.
Hendrickx, Roel P. M.
Kort, Nanne P.
author_sort Schoenmakers, Daphne A. L.
collection PubMed
description PURPOSE: Patient-specific instrumentation (PSI) in total knee arthroplasty (TKA) uses individually designed disposable guides to determine intraoperative bone cuts. The manufacturer provides the surgeon with proposed planning which can be modified and should be approved by the surgeon before the guides are produced. This study aims to assess the intra- and inter-observer reliability among preoperative planning by orthopaedic surgeons using PSI. The authors hypothesize a high intra- and inter-observer reliability in planning TKA using PSI. METHODS: Four orthopaedic surgeons modified and approved 40 preoperative MRI-based PSI plannings three times. The surgeons were blinded to their own and each other’s results. Intra- and inter-observer reliability was obtained for planned implant size, resection, and position of the implant. RESULTS: Intra-observer reliability Intraclass Correlation Coefficients (ICC) were excellent for femoral and tibial implant size with a range of 0.948–0.995 and 0.919–0.988, respectively. Inter-observer reliability for femoral and tibial implant size showed an ICC range of 0.953–0.982 and 0.839-0.951, respectively. Next to implant size, intra- and inter-observer reliability demonstrated good to an excellent agreement (ICC > 0.75) for 7 out of 12 remaining parameters and 6 out of 12 remaining parameters, respectively. CONCLUSION: Preoperative planning of TKA implant size using MRI-based PSI showed excellent intra- and inter-observer reliability. Further research on the comparison of predicted implant size preoperatively to intraoperative results is needed.
format Online
Article
Text
id pubmed-7892674
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-78926742021-03-03 High intra- and inter-observer reliability of planning implant size in MRI-based patient-specific instrumentation for total knee arthroplasty Schoenmakers, Daphne A. L. Theeuwen, Dieuwertje M. J. Schotanus, Martijn G. M. Jansen, Edwin J. P. van Haaren, Emil H. Hendrickx, Roel P. M. Kort, Nanne P. Knee Surg Sports Traumatol Arthrosc Knee PURPOSE: Patient-specific instrumentation (PSI) in total knee arthroplasty (TKA) uses individually designed disposable guides to determine intraoperative bone cuts. The manufacturer provides the surgeon with proposed planning which can be modified and should be approved by the surgeon before the guides are produced. This study aims to assess the intra- and inter-observer reliability among preoperative planning by orthopaedic surgeons using PSI. The authors hypothesize a high intra- and inter-observer reliability in planning TKA using PSI. METHODS: Four orthopaedic surgeons modified and approved 40 preoperative MRI-based PSI plannings three times. The surgeons were blinded to their own and each other’s results. Intra- and inter-observer reliability was obtained for planned implant size, resection, and position of the implant. RESULTS: Intra-observer reliability Intraclass Correlation Coefficients (ICC) were excellent for femoral and tibial implant size with a range of 0.948–0.995 and 0.919–0.988, respectively. Inter-observer reliability for femoral and tibial implant size showed an ICC range of 0.953–0.982 and 0.839-0.951, respectively. Next to implant size, intra- and inter-observer reliability demonstrated good to an excellent agreement (ICC > 0.75) for 7 out of 12 remaining parameters and 6 out of 12 remaining parameters, respectively. CONCLUSION: Preoperative planning of TKA implant size using MRI-based PSI showed excellent intra- and inter-observer reliability. Further research on the comparison of predicted implant size preoperatively to intraoperative results is needed. Springer Berlin Heidelberg 2020-03-30 2021 /pmc/articles/PMC7892674/ /pubmed/32232539 http://dx.doi.org/10.1007/s00167-020-05946-1 Text en © The Author(s) 2020 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
spellingShingle Knee
Schoenmakers, Daphne A. L.
Theeuwen, Dieuwertje M. J.
Schotanus, Martijn G. M.
Jansen, Edwin J. P.
van Haaren, Emil H.
Hendrickx, Roel P. M.
Kort, Nanne P.
High intra- and inter-observer reliability of planning implant size in MRI-based patient-specific instrumentation for total knee arthroplasty
title High intra- and inter-observer reliability of planning implant size in MRI-based patient-specific instrumentation for total knee arthroplasty
title_full High intra- and inter-observer reliability of planning implant size in MRI-based patient-specific instrumentation for total knee arthroplasty
title_fullStr High intra- and inter-observer reliability of planning implant size in MRI-based patient-specific instrumentation for total knee arthroplasty
title_full_unstemmed High intra- and inter-observer reliability of planning implant size in MRI-based patient-specific instrumentation for total knee arthroplasty
title_short High intra- and inter-observer reliability of planning implant size in MRI-based patient-specific instrumentation for total knee arthroplasty
title_sort high intra- and inter-observer reliability of planning implant size in mri-based patient-specific instrumentation for total knee arthroplasty
topic Knee
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7892674/
https://www.ncbi.nlm.nih.gov/pubmed/32232539
http://dx.doi.org/10.1007/s00167-020-05946-1
work_keys_str_mv AT schoenmakersdaphneal highintraandinterobserverreliabilityofplanningimplantsizeinmribasedpatientspecificinstrumentationfortotalkneearthroplasty
AT theeuwendieuwertjemj highintraandinterobserverreliabilityofplanningimplantsizeinmribasedpatientspecificinstrumentationfortotalkneearthroplasty
AT schotanusmartijngm highintraandinterobserverreliabilityofplanningimplantsizeinmribasedpatientspecificinstrumentationfortotalkneearthroplasty
AT jansenedwinjp highintraandinterobserverreliabilityofplanningimplantsizeinmribasedpatientspecificinstrumentationfortotalkneearthroplasty
AT vanhaarenemilh highintraandinterobserverreliabilityofplanningimplantsizeinmribasedpatientspecificinstrumentationfortotalkneearthroplasty
AT hendrickxroelpm highintraandinterobserverreliabilityofplanningimplantsizeinmribasedpatientspecificinstrumentationfortotalkneearthroplasty
AT kortnannep highintraandinterobserverreliabilityofplanningimplantsizeinmribasedpatientspecificinstrumentationfortotalkneearthroplasty