Cargando…

Atrio‐ventricular synchronous pacing with a single chamber leadless pacemaker: Programming and trouble shooting for common clinical scenarios

Micra leadless pacemaker has progressed from a single chamber pacemaker that can deliver VVIR pacing to a pacing device that can provide atrio‐ventricular (AV) synchrony via a unique pacing algorithm that relies on identifying mechanical atrial contraction. This novel algorithm has its own limitatio...

Descripción completa

Detalles Bibliográficos
Autores principales: El‐Chami, Mikhael F., Bhatia, Neal K., Merchant, Faisal M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7894178/
https://www.ncbi.nlm.nih.gov/pubmed/33179814
http://dx.doi.org/10.1111/jce.14807
Descripción
Sumario:Micra leadless pacemaker has progressed from a single chamber pacemaker that can deliver VVIR pacing to a pacing device that can provide atrio‐ventricular (AV) synchrony via a unique pacing algorithm that relies on identifying mechanical atrial contraction. This novel algorithm has its own limitations and intricacies. In this paper, we review this algorithm, suggest steps for troubleshooting and programming these devices and provide clinical examples of Micra AV cases that required changes in programming for adequate tracking of atrial activity.