Evaluation of Polyethylene Wear in a Brazilian Ultracongruent Knee Prosthesis with a Rotating Platform

Objective  To evaluate the wear of polyethylene in a Brazilian ultracongruent knee prosthesis with a rotating platform (Rotaflex, Víncula, Rio Claro, SP, Brasil). Methods  We used the test method with the loading and preparation parameters mentioned in the standards regulation ISO 14243-1:2009 , and...

Descripción completa

Detalles Bibliográficos
Autores principales: Pécora, José Ricardo, Romero, Valéria
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Thieme Revinter Publicações Ltda. 2021
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7895618/
https://www.ncbi.nlm.nih.gov/pubmed/33627898
http://dx.doi.org/10.1055/s-0040-1712491
_version_ 1783653393106993152
author Pécora, José Ricardo
Romero, Valéria
author_facet Pécora, José Ricardo
Romero, Valéria
author_sort Pécora, José Ricardo
collection PubMed
description Objective  To evaluate the wear of polyethylene in a Brazilian ultracongruent knee prosthesis with a rotating platform (Rotaflex, Víncula, Rio Claro, SP, Brasil). Methods  We used the test method with the loading and preparation parameters mentioned in the standards regulation ISO 14243-1:2009 , and the measurement methods mentioned in the standards regulation ISO 14243-2:2009 , for the evaluation of the wear behavior of a Brazilian prosthesis with a rotating platform. The equipment used for the wear test was the ISO 14243–1 gait simulator (EndoLab, Riedering, Germany). Results  After 10 million cycles, the evaluation of the polyethylene wear showed a regular appearance of surface wear at a mean rate of 2.56 mg per million cycles. Conclusion  The wear of the polyethylene of the evaluated prosthesis was minimal after the tests performed and with safety limits higher than those recommended by biomechanical engineering.
format Online
Article
Text
id pubmed-7895618
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Thieme Revinter Publicações Ltda.
record_format MEDLINE/PubMed
spelling pubmed-78956182021-02-23 Evaluation of Polyethylene Wear in a Brazilian Ultracongruent Knee Prosthesis with a Rotating Platform Pécora, José Ricardo Romero, Valéria Rev Bras Ortop (Sao Paulo) Objective  To evaluate the wear of polyethylene in a Brazilian ultracongruent knee prosthesis with a rotating platform (Rotaflex, Víncula, Rio Claro, SP, Brasil). Methods  We used the test method with the loading and preparation parameters mentioned in the standards regulation ISO 14243-1:2009 , and the measurement methods mentioned in the standards regulation ISO 14243-2:2009 , for the evaluation of the wear behavior of a Brazilian prosthesis with a rotating platform. The equipment used for the wear test was the ISO 14243–1 gait simulator (EndoLab, Riedering, Germany). Results  After 10 million cycles, the evaluation of the polyethylene wear showed a regular appearance of surface wear at a mean rate of 2.56 mg per million cycles. Conclusion  The wear of the polyethylene of the evaluated prosthesis was minimal after the tests performed and with safety limits higher than those recommended by biomechanical engineering. Thieme Revinter Publicações Ltda. 2021-02 2020-06-10 /pmc/articles/PMC7895618/ /pubmed/33627898 http://dx.doi.org/10.1055/s-0040-1712491 Text en Sociedade Brasileira de Ortopedia e Traumatologia. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. ( https://creativecommons.org/licenses/by-nc-nd/4.0/ ) https://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivatives License, which permits unrestricted reproduction and distribution, for non-commercial purposes only; and use and reproduction, but not distribution, of adapted material for non-commercial purposes only, provided the original work is properly cited.
spellingShingle Pécora, José Ricardo
Romero, Valéria
Evaluation of Polyethylene Wear in a Brazilian Ultracongruent Knee Prosthesis with a Rotating Platform
title Evaluation of Polyethylene Wear in a Brazilian Ultracongruent Knee Prosthesis with a Rotating Platform
title_full Evaluation of Polyethylene Wear in a Brazilian Ultracongruent Knee Prosthesis with a Rotating Platform
title_fullStr Evaluation of Polyethylene Wear in a Brazilian Ultracongruent Knee Prosthesis with a Rotating Platform
title_full_unstemmed Evaluation of Polyethylene Wear in a Brazilian Ultracongruent Knee Prosthesis with a Rotating Platform
title_short Evaluation of Polyethylene Wear in a Brazilian Ultracongruent Knee Prosthesis with a Rotating Platform
title_sort evaluation of polyethylene wear in a brazilian ultracongruent knee prosthesis with a rotating platform
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7895618/
https://www.ncbi.nlm.nih.gov/pubmed/33627898
http://dx.doi.org/10.1055/s-0040-1712491
work_keys_str_mv AT pecorajosericardo evaluationofpolyethylenewearinabrazilianultracongruentkneeprosthesiswitharotatingplatform
AT romerovaleria evaluationofpolyethylenewearinabrazilianultracongruentkneeprosthesiswitharotatingplatform