Cargando…
Atrial tachycardia misinterpreted as ventricular high‐rate event in a patient with the sick sinus disease and a DDD pacemaker
The device misinterpreted atrial high‐rate episode (AHRE) as ventricular high‐rate episode (VHRE), since consecutive atrial beats fell in the postventricular atrial blanking (PVAB) period. This resulted in the failure of auto mode switch (AMS). This was corrected by decreasing the PVAB and the patie...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7896469/ https://www.ncbi.nlm.nih.gov/pubmed/33664917 http://dx.doi.org/10.1002/joa3.12497 |
Sumario: | The device misinterpreted atrial high‐rate episode (AHRE) as ventricular high‐rate episode (VHRE), since consecutive atrial beats fell in the postventricular atrial blanking (PVAB) period. This resulted in the failure of auto mode switch (AMS). This was corrected by decreasing the PVAB and the patient was asymptomatic on follow‐up.[Image: see text] |
---|