Cargando…
Atrial tachycardia misinterpreted as ventricular high‐rate event in a patient with the sick sinus disease and a DDD pacemaker
The device misinterpreted atrial high‐rate episode (AHRE) as ventricular high‐rate episode (VHRE), since consecutive atrial beats fell in the postventricular atrial blanking (PVAB) period. This resulted in the failure of auto mode switch (AMS). This was corrected by decreasing the PVAB and the patie...
Autores principales: | Ajit Deshpande, Saurabh, Udyavar, Ameya |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7896469/ https://www.ncbi.nlm.nih.gov/pubmed/33664917 http://dx.doi.org/10.1002/joa3.12497 |
Ejemplares similares
-
Pacemaker‐induced ventricular fibrillation during radiofrequency catheter ablation for ventricular tachycardia
por: Nam, Michael C. Y., et al.
Publicado: (2021) -
Pacemaker implantation for sick sinus syndrome in a pregnant female with situs ambiguus and polysplenia
por: Hayashi, Tatsuya, et al.
Publicado: (2020) -
Monomorphic ventricular tachycardia induced by tilt table testing in a patient with syncope and normal heart
por: Furukawa, Toshiyuki, et al.
Publicado: (2021) -
Successful implantation of a left ventricular lead in an anomalous coronary sinus
por: Elvin Gul, Enes, et al.
Publicado: (2021) -
High‐density mapping and ablation of an incessant ventricular tachycardia with endo‐epicardial involvement in a patient with situs inversus totalis and left ventricular noncompaction
por: Solimene, Francesco, et al.
Publicado: (2021)