Technical Note: Design and commissioning of a water phantom for proton dosimetry in magnetic fields
PURPOSE: To design and commission a water phantom suitable for constrained environments and magnetic fields for magnetic resonance (MR)‐guided proton therapy. METHODS: A phantom was designed, to enable precise, remote controlled detector positioning in water within the constrained environment of a m...
Autores principales: | Fuchs, Hermann, Padilla‐Cabal, Fatima, Hummel, Andreas, Georg, Dietmar |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7898880/ https://www.ncbi.nlm.nih.gov/pubmed/33222211 http://dx.doi.org/10.1002/mp.14605 |