Improved reproducibility of diffusion MRI of the human brain with a four‐way blip‐up and down phase‐encoding acquisition approach

PURPOSE: To assess the effects of blip‐up and ‐down echo planar imaging (EPI) acquisition designs, with different choices of phase‐encoding directions (PEDs) on the reproducibility of diffusion MRI (dMRI)‐derived metrics in the human brain. METHODS: Diffusion MRI data in seven subjects were acquired...

Descripción completa

Detalles Bibliográficos
Autores principales: Irfanoglu, M. Okan, Sadeghi, Neda, Sarlls, Joelle, Pierpaoli, Carlo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7898925/
https://www.ncbi.nlm.nih.gov/pubmed/33331068
http://dx.doi.org/10.1002/mrm.28624
Descripción
Sumario:PURPOSE: To assess the effects of blip‐up and ‐down echo planar imaging (EPI) acquisition designs, with different choices of phase‐encoding directions (PEDs) on the reproducibility of diffusion MRI (dMRI)‐derived metrics in the human brain. METHODS: Diffusion MRI data in seven subjects were acquired five times, each with five different protocols. The base design included 64 diffusion directions acquired with anterior‐posterior (AP) PED, the first and second protocols added reverse phase‐encoded [Formula: see text] posterior‐anterior (PA) PED images. The third one included 32 directions all with PED acquisitions with opposite polarity (AP and PA). The fourth protocol, also with 32 unique directions used four PEDs (AP, PA, right‐left (RL), and left‐right (LR)). The scan time was virtually identical for all protocols. The variability of diffusion MRI metrics for each subject and each protocol was computed across the different sessions. RESULTS: The highest reproducibility for all dMRI metrics was obtained with protocol four (AP/PA‐RL/LR, ie, four‐way PED). Protocols that used only [Formula: see text] for distortion correction, which are the most widely used designs, had the lowest reproducibility. CONCLUSIONS: An acquisition design with four PEDs, including all DWIs in addition to [Formula: see text] images should be used to achieve high reproducibility in diffusion MRI studies.