Cargando…

Long-Term Technical Performance of the Osypka QT-5(®) Ventricular Pacemaker Lead

Background: Lead-associated complications and technical issues in patients with cardiac implantable electronic devices are common but underreported in the literature. Methods: All patients undergoing implantation of the Osypka QT-5(®) ventricular lead at the University Clinic St. Pölten between 1 Ja...

Descripción completa

Detalles Bibliográficos
Autores principales: Semmler, Georg, Barbieri, Fabian, Thudt, Karin, Vock, Paul, Mörtl, Deddo, Mayr, Harald, Wollmann, Christian Georg, Adukauskaite, Agne, Pfeifer, Bernhard, Senoner, Thomas, Dichtl, Wolfgang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7915016/
https://www.ncbi.nlm.nih.gov/pubmed/33567486
http://dx.doi.org/10.3390/jcm10040639
Descripción
Sumario:Background: Lead-associated complications and technical issues in patients with cardiac implantable electronic devices are common but underreported in the literature. Methods: All patients undergoing implantation of the Osypka QT-5(®) ventricular lead at the University Clinic St. Pölten between 1 January 2006 and 31 December 2012 were retrospectively analyzed (n = 211). Clinical data including pacemaker follow-up examinations and the need for lead revisions were assessed. Kaplan–Meier analysis to estimate the rate of lead dysfunction during long-term follow-up was conducted. Results: Patients were followed for a median of 5.2 years (interquartile range (IQR) 2.0–8.7). R-wave sensing properties at implantation, compared to last follow-up, remained basically unchanged: 9.9 mV (IQR 6.8–13.4) and 9.6 mV (IQR 5.6–12.0), respectively). Ventricular pacing threshold significantly increased between implantation (0.5 V at 0.4 ms; IQR 0.5–0.8) and the first follow-up visit (1.0 V at 0.4 ms; IQR 0.8–1.3; p < 0.001) and this increase persisted throughout to the last check-up (0.9 V at 0.4 ms; IQR 0.8–1.2). Impedance significantly declined from 1142 Ω (IQR 955–1285) at implantation to 814 Ω (IQR 701–949; p < 0.001) at the first check-up, followed by a further decrease to 450 Ω (IQR 289–652; p < 0.001) at the last check-up. Overall, the Osypka QT-5(®) ventricular lead was replaced in 36 patients (17.1%). Conclusions: This report shows an unexpected high rate of technical issues of the Osypka QT-5(®) ventricular lead during long-term follow-up.