Cargando…

Custom-made 3D-printed cup-cage implants for complex acetabular revisions: evaluation of pre-planned versus achieved positioning and 1-year migration data in 10 patients

Background and purpose — The use of custom-made 3D-printed titanium implants for the reconstruction of large acetabular defects has been successively introduced in the last decade. In an observational cohort study we evaluated the agreement between preoperatively planned and actually achieved cup-ca...

Descripción completa

Detalles Bibliográficos
Autores principales: Zampelis, Vasileios, Flivik, Gunnar
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Taylor & Francis 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7919918/
https://www.ncbi.nlm.nih.gov/pubmed/32928018
http://dx.doi.org/10.1080/17453674.2020.1819729
_version_ 1783658207075368960
author Zampelis, Vasileios
Flivik, Gunnar
author_facet Zampelis, Vasileios
Flivik, Gunnar
author_sort Zampelis, Vasileios
collection PubMed
description Background and purpose — The use of custom-made 3D-printed titanium implants for the reconstruction of large acetabular defects has been successively introduced in the last decade. In an observational cohort study we evaluated the agreement between preoperatively planned and actually achieved cup-cage position as well as 1-year follow-up migration of the cup-cage component. Patients and methods — 10 patients with Paprosky III defects underwent revision surgery using a custom-made 3D-printed cup-cage. The position of the implant on postoperative CT scan was compared with the preoperative plan and the postoperative CT scan was compared with the 1-year follow-up CT scan. Results — There was a median deviation in postoperative position versus planned in inclination of 3.6° (IQR 1.0–5.4), in anteversion of –2.8° (IQR –7.5 to 1.2), and in rotation of –1.2° (IQR –3.3 to 0.0). The median deviation in position of the center of rotation (COR) was –0.5 mm (IQR 2.9 to 0.7) in the anteroposterior (AP) plane, –0.6 mm (IQR –1.8 to –0.1) in the mediolateral (ML) plane, and 1.1 mm (IQR –1.6 to 2.8) in the superoinferior (SI) plane. The migration between postoperative and 1-year follow-up caused a mean change in inclination of 0.04° (IQR –0.06 to 0.09), in anteversion of –0.13° (IQR –0.23 to –0.06), and in rotation of 0.05° (IQR –0.46 to 1.4). The migration of COR was –0.08 mm (IQR –0.18 to –0.04) in the AP plane, 0.14 mm (IQR –0.08 to 0.22) in the ML plane, and 0.06 mm (IQR –0.02 to 0.35) in the SI plane. There was no re-revision. Interpretation — The early results show good agreement between planned and achieved cup-cage position and small measured migration values of the cup-cage component at the 1-year follow-up.
format Online
Article
Text
id pubmed-7919918
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Taylor & Francis
record_format MEDLINE/PubMed
spelling pubmed-79199182021-03-10 Custom-made 3D-printed cup-cage implants for complex acetabular revisions: evaluation of pre-planned versus achieved positioning and 1-year migration data in 10 patients Zampelis, Vasileios Flivik, Gunnar Acta Orthop Articles Background and purpose — The use of custom-made 3D-printed titanium implants for the reconstruction of large acetabular defects has been successively introduced in the last decade. In an observational cohort study we evaluated the agreement between preoperatively planned and actually achieved cup-cage position as well as 1-year follow-up migration of the cup-cage component. Patients and methods — 10 patients with Paprosky III defects underwent revision surgery using a custom-made 3D-printed cup-cage. The position of the implant on postoperative CT scan was compared with the preoperative plan and the postoperative CT scan was compared with the 1-year follow-up CT scan. Results — There was a median deviation in postoperative position versus planned in inclination of 3.6° (IQR 1.0–5.4), in anteversion of –2.8° (IQR –7.5 to 1.2), and in rotation of –1.2° (IQR –3.3 to 0.0). The median deviation in position of the center of rotation (COR) was –0.5 mm (IQR 2.9 to 0.7) in the anteroposterior (AP) plane, –0.6 mm (IQR –1.8 to –0.1) in the mediolateral (ML) plane, and 1.1 mm (IQR –1.6 to 2.8) in the superoinferior (SI) plane. The migration between postoperative and 1-year follow-up caused a mean change in inclination of 0.04° (IQR –0.06 to 0.09), in anteversion of –0.13° (IQR –0.23 to –0.06), and in rotation of 0.05° (IQR –0.46 to 1.4). The migration of COR was –0.08 mm (IQR –0.18 to –0.04) in the AP plane, 0.14 mm (IQR –0.08 to 0.22) in the ML plane, and 0.06 mm (IQR –0.02 to 0.35) in the SI plane. There was no re-revision. Interpretation — The early results show good agreement between planned and achieved cup-cage position and small measured migration values of the cup-cage component at the 1-year follow-up. Taylor & Francis 2020-09-15 /pmc/articles/PMC7919918/ /pubmed/32928018 http://dx.doi.org/10.1080/17453674.2020.1819729 Text en © 2020 The Author(s). Published by Taylor & Francis on behalf of the Nordic Orthopedic Federation. https://creativecommons.org/licenses/by/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Articles
Zampelis, Vasileios
Flivik, Gunnar
Custom-made 3D-printed cup-cage implants for complex acetabular revisions: evaluation of pre-planned versus achieved positioning and 1-year migration data in 10 patients
title Custom-made 3D-printed cup-cage implants for complex acetabular revisions: evaluation of pre-planned versus achieved positioning and 1-year migration data in 10 patients
title_full Custom-made 3D-printed cup-cage implants for complex acetabular revisions: evaluation of pre-planned versus achieved positioning and 1-year migration data in 10 patients
title_fullStr Custom-made 3D-printed cup-cage implants for complex acetabular revisions: evaluation of pre-planned versus achieved positioning and 1-year migration data in 10 patients
title_full_unstemmed Custom-made 3D-printed cup-cage implants for complex acetabular revisions: evaluation of pre-planned versus achieved positioning and 1-year migration data in 10 patients
title_short Custom-made 3D-printed cup-cage implants for complex acetabular revisions: evaluation of pre-planned versus achieved positioning and 1-year migration data in 10 patients
title_sort custom-made 3d-printed cup-cage implants for complex acetabular revisions: evaluation of pre-planned versus achieved positioning and 1-year migration data in 10 patients
topic Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7919918/
https://www.ncbi.nlm.nih.gov/pubmed/32928018
http://dx.doi.org/10.1080/17453674.2020.1819729
work_keys_str_mv AT zampelisvasileios custommade3dprintedcupcageimplantsforcomplexacetabularrevisionsevaluationofpreplannedversusachievedpositioningand1yearmigrationdatain10patients
AT flivikgunnar custommade3dprintedcupcageimplantsforcomplexacetabularrevisionsevaluationofpreplannedversusachievedpositioningand1yearmigrationdatain10patients