“Off-label” Usage of an Oxidized Zirconium Femoral Head in Revision of a Total Hip Arthroplasty with Mechanically Assisted Crevice Corrosion and a Legacy Taper

We present a case of a 72-year-old male with a history of a late 1980s metal-on-polyethylene total hip arthroplasty who presented with unilateral leg vascular compromise, joint pain, and stiffness and subsequently underwent revision for adverse local tissue reaction secondary to mechanically assiste...

Descripción completa

Detalles Bibliográficos
Autores principales: Lachance, Andrew D., McGrory, Brian J., Christman, Robert A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7930501/
https://www.ncbi.nlm.nih.gov/pubmed/33681438
http://dx.doi.org/10.1016/j.artd.2021.01.012
Descripción
Sumario:We present a case of a 72-year-old male with a history of a late 1980s metal-on-polyethylene total hip arthroplasty who presented with unilateral leg vascular compromise, joint pain, and stiffness and subsequently underwent revision for adverse local tissue reaction secondary to mechanically assisted crevice corrosion. His stable and extensively porous coated femoral implant had a legacy taper with no currently manufactured option for a non–Co-alloy femoral head. After shared decision-making with the patient, we opted to use an oxidized zirconium femoral head from another manufacturer with a similar taper during his revision surgery and documented that his vascular compromise resolved and his serum Co was undetectable 3 years after the revision.