Micro-Networks for Robust MR-Guided Low Count PET Imaging

Noise suppression is particularly important in low count positron emission tomography (PET) imaging. Post-smoothing (PS) and regularization methods which aim to reduce noise also tend to reduce resolution and introduce bias. Alternatively, anatomical information from another modality such as magneti...

Descripción completa

Detalles Bibliográficos
Formato: Online Artículo Texto
Lenguaje:English
Publicado: IEEE 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7931458/
https://www.ncbi.nlm.nih.gov/pubmed/33681546
http://dx.doi.org/10.1109/TRPMS.2020.2986414
_version_ 1783660297554231296
collection PubMed
description Noise suppression is particularly important in low count positron emission tomography (PET) imaging. Post-smoothing (PS) and regularization methods which aim to reduce noise also tend to reduce resolution and introduce bias. Alternatively, anatomical information from another modality such as magnetic resonance (MR) imaging can be used to improve image quality. Convolutional neural networks (CNNs) are particularly well suited to such joint image processing, but usually require large amounts of training data and have mostly been applied outside the field of medical imaging or focus on classification and segmentation, leaving PET image quality improvement relatively understudied. This article proposes the use of a relatively low-complexity CNN (micro-net) as a post-reconstruction MR-guided image processing step to reduce noise and reconstruction artefacts while also improving resolution in low count PET scans. The CNN is designed to be fully 3-D, robust to very limited amounts of training data, and to accept multiple inputs (including competitive denoising methods). Application of the proposed CNN on simulated low (30 M) count data (trained to produce standard (300 M) count reconstructions) results in a 36% lower normalized root mean squared error (NRMSE, calculated over ten realizations against the ground truth) compared to maximum-likelihood expectation maximization (MLEM) used in clinical practice. In contrast, a decrease of only 25% in NRMSE is obtained when an optimized (using knowledge of the ground truth) PS is performed. A 26% NRMSE decrease is obtained with both RM and optimized PS. Similar improvement is also observed for low count real patient datasets. Overfitting to training data is demonstrated to occur as the network size is increased. In an extreme case, a U-net (which produces better predictions for training data) is shown to completely fail on test data due to overfitting to this case of very limited training data. Meanwhile, the resultant images from the proposed CNN (which has low training data requirements) have lower noise, reduced ringing, and partial volume effects, as well as sharper edges and improved resolution compared to conventional MLEM.
format Online
Article
Text
id pubmed-7931458
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher IEEE
record_format MEDLINE/PubMed
spelling pubmed-79314582021-03-05 Micro-Networks for Robust MR-Guided Low Count PET Imaging IEEE Trans Radiat Plasma Med Sci Article Noise suppression is particularly important in low count positron emission tomography (PET) imaging. Post-smoothing (PS) and regularization methods which aim to reduce noise also tend to reduce resolution and introduce bias. Alternatively, anatomical information from another modality such as magnetic resonance (MR) imaging can be used to improve image quality. Convolutional neural networks (CNNs) are particularly well suited to such joint image processing, but usually require large amounts of training data and have mostly been applied outside the field of medical imaging or focus on classification and segmentation, leaving PET image quality improvement relatively understudied. This article proposes the use of a relatively low-complexity CNN (micro-net) as a post-reconstruction MR-guided image processing step to reduce noise and reconstruction artefacts while also improving resolution in low count PET scans. The CNN is designed to be fully 3-D, robust to very limited amounts of training data, and to accept multiple inputs (including competitive denoising methods). Application of the proposed CNN on simulated low (30 M) count data (trained to produce standard (300 M) count reconstructions) results in a 36% lower normalized root mean squared error (NRMSE, calculated over ten realizations against the ground truth) compared to maximum-likelihood expectation maximization (MLEM) used in clinical practice. In contrast, a decrease of only 25% in NRMSE is obtained when an optimized (using knowledge of the ground truth) PS is performed. A 26% NRMSE decrease is obtained with both RM and optimized PS. Similar improvement is also observed for low count real patient datasets. Overfitting to training data is demonstrated to occur as the network size is increased. In an extreme case, a U-net (which produces better predictions for training data) is shown to completely fail on test data due to overfitting to this case of very limited training data. Meanwhile, the resultant images from the proposed CNN (which has low training data requirements) have lower noise, reduced ringing, and partial volume effects, as well as sharper edges and improved resolution compared to conventional MLEM. IEEE 2020-04-08 /pmc/articles/PMC7931458/ /pubmed/33681546 http://dx.doi.org/10.1109/TRPMS.2020.2986414 Text en https://creativecommons.org/licenses/by/4.0/ This work is licensed under a Creative Commons Attribution 4.0 License. For more information, see https://creativecommons.org/licenses/by/4.0/
spellingShingle Article
Micro-Networks for Robust MR-Guided Low Count PET Imaging
title Micro-Networks for Robust MR-Guided Low Count PET Imaging
title_full Micro-Networks for Robust MR-Guided Low Count PET Imaging
title_fullStr Micro-Networks for Robust MR-Guided Low Count PET Imaging
title_full_unstemmed Micro-Networks for Robust MR-Guided Low Count PET Imaging
title_short Micro-Networks for Robust MR-Guided Low Count PET Imaging
title_sort micro-networks for robust mr-guided low count pet imaging
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7931458/
https://www.ncbi.nlm.nih.gov/pubmed/33681546
http://dx.doi.org/10.1109/TRPMS.2020.2986414
work_keys_str_mv AT micronetworksforrobustmrguidedlowcountpetimaging
AT micronetworksforrobustmrguidedlowcountpetimaging