Performance of the hematology analyzer XN-31 prototype in the detection of Plasmodium infections in an endemic region of Colombia

Early and accurate diagnosis is critical in reducing the morbidity and mortality associated with malaria. Microscopy (MI) is the current diagnostic gold standard in the field; however, it requires expert personnel, is time-consuming, and has limited sensitivity. Although rapid diagnostic tests for a...

Descripción completa

Detalles Bibliográficos
Autores principales: Zuluaga-Idárraga, Lina, Rios, Alexandra, Sierra-Cifuentes, Verónica, Garzón, Edwar, Tobón-Castaño, Alberto, Takehara, Ikki, Toya, Yuji, Izuka, Munehisa, Uchihashi, Kinya, Lopera-Mesa, Tatiana M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7933134/
https://www.ncbi.nlm.nih.gov/pubmed/33664389
http://dx.doi.org/10.1038/s41598-021-84594-y
_version_ 1783660542610636800
author Zuluaga-Idárraga, Lina
Rios, Alexandra
Sierra-Cifuentes, Verónica
Garzón, Edwar
Tobón-Castaño, Alberto
Takehara, Ikki
Toya, Yuji
Izuka, Munehisa
Uchihashi, Kinya
Lopera-Mesa, Tatiana M.
author_facet Zuluaga-Idárraga, Lina
Rios, Alexandra
Sierra-Cifuentes, Verónica
Garzón, Edwar
Tobón-Castaño, Alberto
Takehara, Ikki
Toya, Yuji
Izuka, Munehisa
Uchihashi, Kinya
Lopera-Mesa, Tatiana M.
author_sort Zuluaga-Idárraga, Lina
collection PubMed
description Early and accurate diagnosis is critical in reducing the morbidity and mortality associated with malaria. Microscopy (MI) is the current diagnostic gold standard in the field; however, it requires expert personnel, is time-consuming, and has limited sensitivity. Although rapid diagnostic tests for antigen detection (RDTs) are an alternative to diagnosis, they also have limited sensitivity and produce false positive results in detecting recent past infection. The automated hematology analyzer XN-31 prototype (XN-31p) (Sysmex Corporation, Kobe, Japan) is able to identify plasmodium-infected erythrocytes, count parasitemia and perform complete blood-cell counts within one minute. The performance of the XN-31p in diagnosing malaria was evaluated and compared with real-time polymerase chain reaction (qPCR), MI and RDT in an endemic area of Colombia where Plasmodium falciparum and Plasmodium vivax are present. Acute febrile patients were enrolled from July 2018 to April 2019 in Quibdó, Colombia. Malaria diagnoses were obtained from MI and RDT in the field and later confirmed by qPCR. Venous blood samples in EDTA were processed with an XN-31p in the field. Sensitivity, specificity, positive/negative predictive values, and the likelihood ratios of positive and negative tests were calculated with respect to the results from qPCR, MI and RDT. The intraclass correlation coefficient (ICC) and Bland–Altman plot were used to evaluate the concordance in the parasitemia with respect to MI. A total of 1,754 subjects were enrolled. The mean age was 27.0 years (IQR 14–44); 89.6% were Afro-Colombians, 94.3% lived in urban areas and 0.91% were pregnant. With respect to qPCR, the XN-31p showed a sensitivity of 90% (95% CI 87.24–92.34) and a specificity of 99.83% (95% CI 99.38–99.98) in detecting Plasmodium spp.; both parameters were equivalent to those for MI and RDT. Using MI as the reference, the XN-31p showed a sensitivity of 98.09% (95% CI 96.51–99.08), a specificity of 99.83% (95% CI 99.4–99.98), an ICC of 0.85 (95% CI 0.83–0.87) and an average difference of − 3096 parasites/µL when compared with thick-smear MI and an ICC of 0.98 (95% CI 0.97–0.98) and an average difference of − 0.0013% when compared with thin-smear MI. The XN-31p offers a rapid and accurate alternative method for diagnosing malaria in clinical laboratories in areas where P. falciparum and P. vivax cocirculate.
format Online
Article
Text
id pubmed-7933134
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Nature Publishing Group UK
record_format MEDLINE/PubMed
spelling pubmed-79331342021-03-05 Performance of the hematology analyzer XN-31 prototype in the detection of Plasmodium infections in an endemic region of Colombia Zuluaga-Idárraga, Lina Rios, Alexandra Sierra-Cifuentes, Verónica Garzón, Edwar Tobón-Castaño, Alberto Takehara, Ikki Toya, Yuji Izuka, Munehisa Uchihashi, Kinya Lopera-Mesa, Tatiana M. Sci Rep Article Early and accurate diagnosis is critical in reducing the morbidity and mortality associated with malaria. Microscopy (MI) is the current diagnostic gold standard in the field; however, it requires expert personnel, is time-consuming, and has limited sensitivity. Although rapid diagnostic tests for antigen detection (RDTs) are an alternative to diagnosis, they also have limited sensitivity and produce false positive results in detecting recent past infection. The automated hematology analyzer XN-31 prototype (XN-31p) (Sysmex Corporation, Kobe, Japan) is able to identify plasmodium-infected erythrocytes, count parasitemia and perform complete blood-cell counts within one minute. The performance of the XN-31p in diagnosing malaria was evaluated and compared with real-time polymerase chain reaction (qPCR), MI and RDT in an endemic area of Colombia where Plasmodium falciparum and Plasmodium vivax are present. Acute febrile patients were enrolled from July 2018 to April 2019 in Quibdó, Colombia. Malaria diagnoses were obtained from MI and RDT in the field and later confirmed by qPCR. Venous blood samples in EDTA were processed with an XN-31p in the field. Sensitivity, specificity, positive/negative predictive values, and the likelihood ratios of positive and negative tests were calculated with respect to the results from qPCR, MI and RDT. The intraclass correlation coefficient (ICC) and Bland–Altman plot were used to evaluate the concordance in the parasitemia with respect to MI. A total of 1,754 subjects were enrolled. The mean age was 27.0 years (IQR 14–44); 89.6% were Afro-Colombians, 94.3% lived in urban areas and 0.91% were pregnant. With respect to qPCR, the XN-31p showed a sensitivity of 90% (95% CI 87.24–92.34) and a specificity of 99.83% (95% CI 99.38–99.98) in detecting Plasmodium spp.; both parameters were equivalent to those for MI and RDT. Using MI as the reference, the XN-31p showed a sensitivity of 98.09% (95% CI 96.51–99.08), a specificity of 99.83% (95% CI 99.4–99.98), an ICC of 0.85 (95% CI 0.83–0.87) and an average difference of − 3096 parasites/µL when compared with thick-smear MI and an ICC of 0.98 (95% CI 0.97–0.98) and an average difference of − 0.0013% when compared with thin-smear MI. The XN-31p offers a rapid and accurate alternative method for diagnosing malaria in clinical laboratories in areas where P. falciparum and P. vivax cocirculate. Nature Publishing Group UK 2021-03-04 /pmc/articles/PMC7933134/ /pubmed/33664389 http://dx.doi.org/10.1038/s41598-021-84594-y Text en © The Author(s) 2021 Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
spellingShingle Article
Zuluaga-Idárraga, Lina
Rios, Alexandra
Sierra-Cifuentes, Verónica
Garzón, Edwar
Tobón-Castaño, Alberto
Takehara, Ikki
Toya, Yuji
Izuka, Munehisa
Uchihashi, Kinya
Lopera-Mesa, Tatiana M.
Performance of the hematology analyzer XN-31 prototype in the detection of Plasmodium infections in an endemic region of Colombia
title Performance of the hematology analyzer XN-31 prototype in the detection of Plasmodium infections in an endemic region of Colombia
title_full Performance of the hematology analyzer XN-31 prototype in the detection of Plasmodium infections in an endemic region of Colombia
title_fullStr Performance of the hematology analyzer XN-31 prototype in the detection of Plasmodium infections in an endemic region of Colombia
title_full_unstemmed Performance of the hematology analyzer XN-31 prototype in the detection of Plasmodium infections in an endemic region of Colombia
title_short Performance of the hematology analyzer XN-31 prototype in the detection of Plasmodium infections in an endemic region of Colombia
title_sort performance of the hematology analyzer xn-31 prototype in the detection of plasmodium infections in an endemic region of colombia
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7933134/
https://www.ncbi.nlm.nih.gov/pubmed/33664389
http://dx.doi.org/10.1038/s41598-021-84594-y
work_keys_str_mv AT zuluagaidarragalina performanceofthehematologyanalyzerxn31prototypeinthedetectionofplasmodiuminfectionsinanendemicregionofcolombia
AT riosalexandra performanceofthehematologyanalyzerxn31prototypeinthedetectionofplasmodiuminfectionsinanendemicregionofcolombia
AT sierracifuentesveronica performanceofthehematologyanalyzerxn31prototypeinthedetectionofplasmodiuminfectionsinanendemicregionofcolombia
AT garzonedwar performanceofthehematologyanalyzerxn31prototypeinthedetectionofplasmodiuminfectionsinanendemicregionofcolombia
AT toboncastanoalberto performanceofthehematologyanalyzerxn31prototypeinthedetectionofplasmodiuminfectionsinanendemicregionofcolombia
AT takeharaikki performanceofthehematologyanalyzerxn31prototypeinthedetectionofplasmodiuminfectionsinanendemicregionofcolombia
AT toyayuji performanceofthehematologyanalyzerxn31prototypeinthedetectionofplasmodiuminfectionsinanendemicregionofcolombia
AT izukamunehisa performanceofthehematologyanalyzerxn31prototypeinthedetectionofplasmodiuminfectionsinanendemicregionofcolombia
AT uchihashikinya performanceofthehematologyanalyzerxn31prototypeinthedetectionofplasmodiuminfectionsinanendemicregionofcolombia
AT loperamesatatianam performanceofthehematologyanalyzerxn31prototypeinthedetectionofplasmodiuminfectionsinanendemicregionofcolombia