Performance of a closed-loop glucose control system, comprising a continuous glucose monitoring system and an AI-based controller in swine during severe hypo- and hyperglycemic provocations
Intensive care unit (ICU) patients develop stress induced insulin resistance causing hyperglycemia, large glucose variability and hypoglycemia. These glucose metrics have all been associated with increased rates of morbidity and mortality. The only way to achieve safe glucose control at a lower gluc...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Netherlands
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7943496/ https://www.ncbi.nlm.nih.gov/pubmed/32006145 http://dx.doi.org/10.1007/s10877-020-00474-2 |
_version_ | 1783662505755672576 |
---|---|
author | DeJournett, Jeremy Nekludov, Michael DeJournett, Leon Wallin, Mats |
author_facet | DeJournett, Jeremy Nekludov, Michael DeJournett, Leon Wallin, Mats |
author_sort | DeJournett, Jeremy |
collection | PubMed |
description | Intensive care unit (ICU) patients develop stress induced insulin resistance causing hyperglycemia, large glucose variability and hypoglycemia. These glucose metrics have all been associated with increased rates of morbidity and mortality. The only way to achieve safe glucose control at a lower glucose range (e.g., 4.4–6.6 mmol/L) will be through use of an autonomous closed loop glucose control system (artificial pancreas). Our goal with the present study was to assess the safety and performance of an artificial pancreas system, composed of the EIRUS (Maquet Critical Care AB) continuous glucose monitor (CGM) and novel artificial intelligence-based glucose control software, in a swine model using unannounced hypo- and hyperglycemia challenges. Fourteen piglets (6 control, 8 treated) underwent sequential unannounced hypoglycemic and hyperglycemic challenges with 3 IU of NovoRapid and a glucose infusion at 17 mg/kg/min over the course of 5 h. In the Control animals an experienced ICU physician used every 30-min blood glucose values to maintain control to a range of 4.4–9 mmol/L. In the Treated group the artificial pancreas system attempted to maintain blood glucose control to a range of 4.4–6.6 mmol/L. Five of six Control animals and none of eight Treated animals experienced severe hypoglycemia (< 2.22 mmol/L). The area under the curve 3.5 mmol/L was 28.9 (21.1–54.2) for Control and 4.8 (3.1–5.2) for the Treated animals. The total percent time within tight glucose control range, 4.4–6.6 mmol/L, was 32.8% (32.4–47.1) for Controls and 55.4% (52.9–59.4) for Treated (p < 0.034). Data are median and quartiles. The artificial pancreas system abolished severe hypoglycemia and outperformed the experienced ICU physician in avoiding clinically significant hypoglycemic excursions. |
format | Online Article Text |
id | pubmed-7943496 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Springer Netherlands |
record_format | MEDLINE/PubMed |
spelling | pubmed-79434962021-03-28 Performance of a closed-loop glucose control system, comprising a continuous glucose monitoring system and an AI-based controller in swine during severe hypo- and hyperglycemic provocations DeJournett, Jeremy Nekludov, Michael DeJournett, Leon Wallin, Mats J Clin Monit Comput Original Research Intensive care unit (ICU) patients develop stress induced insulin resistance causing hyperglycemia, large glucose variability and hypoglycemia. These glucose metrics have all been associated with increased rates of morbidity and mortality. The only way to achieve safe glucose control at a lower glucose range (e.g., 4.4–6.6 mmol/L) will be through use of an autonomous closed loop glucose control system (artificial pancreas). Our goal with the present study was to assess the safety and performance of an artificial pancreas system, composed of the EIRUS (Maquet Critical Care AB) continuous glucose monitor (CGM) and novel artificial intelligence-based glucose control software, in a swine model using unannounced hypo- and hyperglycemia challenges. Fourteen piglets (6 control, 8 treated) underwent sequential unannounced hypoglycemic and hyperglycemic challenges with 3 IU of NovoRapid and a glucose infusion at 17 mg/kg/min over the course of 5 h. In the Control animals an experienced ICU physician used every 30-min blood glucose values to maintain control to a range of 4.4–9 mmol/L. In the Treated group the artificial pancreas system attempted to maintain blood glucose control to a range of 4.4–6.6 mmol/L. Five of six Control animals and none of eight Treated animals experienced severe hypoglycemia (< 2.22 mmol/L). The area under the curve 3.5 mmol/L was 28.9 (21.1–54.2) for Control and 4.8 (3.1–5.2) for the Treated animals. The total percent time within tight glucose control range, 4.4–6.6 mmol/L, was 32.8% (32.4–47.1) for Controls and 55.4% (52.9–59.4) for Treated (p < 0.034). Data are median and quartiles. The artificial pancreas system abolished severe hypoglycemia and outperformed the experienced ICU physician in avoiding clinically significant hypoglycemic excursions. Springer Netherlands 2020-01-31 2021 /pmc/articles/PMC7943496/ /pubmed/32006145 http://dx.doi.org/10.1007/s10877-020-00474-2 Text en © The Author(s) 2020 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. |
spellingShingle | Original Research DeJournett, Jeremy Nekludov, Michael DeJournett, Leon Wallin, Mats Performance of a closed-loop glucose control system, comprising a continuous glucose monitoring system and an AI-based controller in swine during severe hypo- and hyperglycemic provocations |
title | Performance of a closed-loop glucose control system, comprising a continuous glucose monitoring system and an AI-based controller in swine during severe hypo- and hyperglycemic provocations |
title_full | Performance of a closed-loop glucose control system, comprising a continuous glucose monitoring system and an AI-based controller in swine during severe hypo- and hyperglycemic provocations |
title_fullStr | Performance of a closed-loop glucose control system, comprising a continuous glucose monitoring system and an AI-based controller in swine during severe hypo- and hyperglycemic provocations |
title_full_unstemmed | Performance of a closed-loop glucose control system, comprising a continuous glucose monitoring system and an AI-based controller in swine during severe hypo- and hyperglycemic provocations |
title_short | Performance of a closed-loop glucose control system, comprising a continuous glucose monitoring system and an AI-based controller in swine during severe hypo- and hyperglycemic provocations |
title_sort | performance of a closed-loop glucose control system, comprising a continuous glucose monitoring system and an ai-based controller in swine during severe hypo- and hyperglycemic provocations |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7943496/ https://www.ncbi.nlm.nih.gov/pubmed/32006145 http://dx.doi.org/10.1007/s10877-020-00474-2 |
work_keys_str_mv | AT dejournettjeremy performanceofaclosedloopglucosecontrolsystemcomprisingacontinuousglucosemonitoringsystemandanaibasedcontrollerinswineduringseverehypoandhyperglycemicprovocations AT nekludovmichael performanceofaclosedloopglucosecontrolsystemcomprisingacontinuousglucosemonitoringsystemandanaibasedcontrollerinswineduringseverehypoandhyperglycemicprovocations AT dejournettleon performanceofaclosedloopglucosecontrolsystemcomprisingacontinuousglucosemonitoringsystemandanaibasedcontrollerinswineduringseverehypoandhyperglycemicprovocations AT wallinmats performanceofaclosedloopglucosecontrolsystemcomprisingacontinuousglucosemonitoringsystemandanaibasedcontrollerinswineduringseverehypoandhyperglycemicprovocations |