Validation of a portable, non-mydriatic fundus camera compared to gold standard dilated fundus examination using slit lamp biomicroscopy for assessing the optic disc for glaucoma

PURPOSE: To evaluate the sensitivity and specificity of a portable non-mydriatic fundus camera to assess the optic disc for glaucoma. METHODS: We conducted a single-site, cross-sectional, observational, instrument validation study. Non-mydriatic fundus photographs centred at the optic disc were obta...

Descripción completa

Detalles Bibliográficos
Autores principales: Upadhyaya, Swati, Agarwal, Anushri, Rengaraj, Venkatesh, Srinivasan, Kavitha, Newman Casey, Paula Anne, Schehlein, Emily
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7947938/
https://www.ncbi.nlm.nih.gov/pubmed/33707762
http://dx.doi.org/10.1038/s41433-021-01485-2
_version_ 1783663332424679424
author Upadhyaya, Swati
Agarwal, Anushri
Rengaraj, Venkatesh
Srinivasan, Kavitha
Newman Casey, Paula Anne
Schehlein, Emily
author_facet Upadhyaya, Swati
Agarwal, Anushri
Rengaraj, Venkatesh
Srinivasan, Kavitha
Newman Casey, Paula Anne
Schehlein, Emily
author_sort Upadhyaya, Swati
collection PubMed
description PURPOSE: To evaluate the sensitivity and specificity of a portable non-mydriatic fundus camera to assess the optic disc for glaucoma. METHODS: We conducted a single-site, cross-sectional, observational, instrument validation study. Non-mydriatic fundus photographs centred at the optic disc were obtained from 276 eyes of 68 glaucoma and 70 normal patients, using a portable fundus camera (Smartscope, Optomed, Oulu, Finland). A senior Glaucoma consultant, masked to the patient’s study participation, performed a gold standard dilated fundus examination to make the diagnosis of glaucoma. Following this, a mydriatic photograph was taken by a standard table-top fundus camera. All the images were digitalized and de-identified by an independent investigator and presented to two remote graders, masked to the patients, their diagnoses, and photographic modality. Based on individual disc characteristics, a diagnosis of screening positive or negative for glaucoma was made. In the end, the independent investigator re-identified the images. Sensitivity and specificity to detect glaucoma with the undilated Smartscope camera was calculated compared to dilated fundus examination. RESULTS: Grading remote images taken with the portable non-mydriatic fundus camera showed a sensitivity of 96.3% (95% confidence interval (CI): 91.6–98.8%) and 94.8% (95% CI: 89.7–97.9%) and a specificity of 98.5% (95% CI: 94.9–99.8%) and 97.8% (95% CI: 93.9–99.6%) for the two graders respectively as compared to gold standard dilated fundus examination. CONCLUSION: The non-mydriatic Smartscope fundus images have high sensitivity and specificity for diagnosing glaucoma remotely and thus may be an effective tool for use in community outreach programs.
format Online
Article
Text
id pubmed-7947938
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Nature Publishing Group UK
record_format MEDLINE/PubMed
spelling pubmed-79479382021-03-11 Validation of a portable, non-mydriatic fundus camera compared to gold standard dilated fundus examination using slit lamp biomicroscopy for assessing the optic disc for glaucoma Upadhyaya, Swati Agarwal, Anushri Rengaraj, Venkatesh Srinivasan, Kavitha Newman Casey, Paula Anne Schehlein, Emily Eye (Lond) Article PURPOSE: To evaluate the sensitivity and specificity of a portable non-mydriatic fundus camera to assess the optic disc for glaucoma. METHODS: We conducted a single-site, cross-sectional, observational, instrument validation study. Non-mydriatic fundus photographs centred at the optic disc were obtained from 276 eyes of 68 glaucoma and 70 normal patients, using a portable fundus camera (Smartscope, Optomed, Oulu, Finland). A senior Glaucoma consultant, masked to the patient’s study participation, performed a gold standard dilated fundus examination to make the diagnosis of glaucoma. Following this, a mydriatic photograph was taken by a standard table-top fundus camera. All the images were digitalized and de-identified by an independent investigator and presented to two remote graders, masked to the patients, their diagnoses, and photographic modality. Based on individual disc characteristics, a diagnosis of screening positive or negative for glaucoma was made. In the end, the independent investigator re-identified the images. Sensitivity and specificity to detect glaucoma with the undilated Smartscope camera was calculated compared to dilated fundus examination. RESULTS: Grading remote images taken with the portable non-mydriatic fundus camera showed a sensitivity of 96.3% (95% confidence interval (CI): 91.6–98.8%) and 94.8% (95% CI: 89.7–97.9%) and a specificity of 98.5% (95% CI: 94.9–99.8%) and 97.8% (95% CI: 93.9–99.6%) for the two graders respectively as compared to gold standard dilated fundus examination. CONCLUSION: The non-mydriatic Smartscope fundus images have high sensitivity and specificity for diagnosing glaucoma remotely and thus may be an effective tool for use in community outreach programs. Nature Publishing Group UK 2021-03-11 2022-02 /pmc/articles/PMC7947938/ /pubmed/33707762 http://dx.doi.org/10.1038/s41433-021-01485-2 Text en © The Author(s), under exclusive licence to The Royal College of Ophthalmologists 2021
spellingShingle Article
Upadhyaya, Swati
Agarwal, Anushri
Rengaraj, Venkatesh
Srinivasan, Kavitha
Newman Casey, Paula Anne
Schehlein, Emily
Validation of a portable, non-mydriatic fundus camera compared to gold standard dilated fundus examination using slit lamp biomicroscopy for assessing the optic disc for glaucoma
title Validation of a portable, non-mydriatic fundus camera compared to gold standard dilated fundus examination using slit lamp biomicroscopy for assessing the optic disc for glaucoma
title_full Validation of a portable, non-mydriatic fundus camera compared to gold standard dilated fundus examination using slit lamp biomicroscopy for assessing the optic disc for glaucoma
title_fullStr Validation of a portable, non-mydriatic fundus camera compared to gold standard dilated fundus examination using slit lamp biomicroscopy for assessing the optic disc for glaucoma
title_full_unstemmed Validation of a portable, non-mydriatic fundus camera compared to gold standard dilated fundus examination using slit lamp biomicroscopy for assessing the optic disc for glaucoma
title_short Validation of a portable, non-mydriatic fundus camera compared to gold standard dilated fundus examination using slit lamp biomicroscopy for assessing the optic disc for glaucoma
title_sort validation of a portable, non-mydriatic fundus camera compared to gold standard dilated fundus examination using slit lamp biomicroscopy for assessing the optic disc for glaucoma
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7947938/
https://www.ncbi.nlm.nih.gov/pubmed/33707762
http://dx.doi.org/10.1038/s41433-021-01485-2
work_keys_str_mv AT upadhyayaswati validationofaportablenonmydriaticfunduscameracomparedtogoldstandarddilatedfundusexaminationusingslitlampbiomicroscopyforassessingtheopticdiscforglaucoma
AT agarwalanushri validationofaportablenonmydriaticfunduscameracomparedtogoldstandarddilatedfundusexaminationusingslitlampbiomicroscopyforassessingtheopticdiscforglaucoma
AT rengarajvenkatesh validationofaportablenonmydriaticfunduscameracomparedtogoldstandarddilatedfundusexaminationusingslitlampbiomicroscopyforassessingtheopticdiscforglaucoma
AT srinivasankavitha validationofaportablenonmydriaticfunduscameracomparedtogoldstandarddilatedfundusexaminationusingslitlampbiomicroscopyforassessingtheopticdiscforglaucoma
AT newmancaseypaulaanne validationofaportablenonmydriaticfunduscameracomparedtogoldstandarddilatedfundusexaminationusingslitlampbiomicroscopyforassessingtheopticdiscforglaucoma
AT schehleinemily validationofaportablenonmydriaticfunduscameracomparedtogoldstandarddilatedfundusexaminationusingslitlampbiomicroscopyforassessingtheopticdiscforglaucoma