Aortic Valve Leaflet Disruption: A Severe Complication of Impella 5.5

A 73-year-old male with a history of severe coronary artery disease and prior coronary artery bypass grafting (CABG) presented with chest pain and elevated troponins. His workup revealed an ejection fraction of 15%, severe native coronary disease, as well as stenosis of his bypass grafts. He underwe...

Descripción completa

Detalles Bibliográficos
Autores principales: Ghannam, Alexander D, Takebe, Manabu, Harmon, Taylor S, Tatum, Scott, Pirris, John
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7948312/
https://www.ncbi.nlm.nih.gov/pubmed/33728184
http://dx.doi.org/10.7759/cureus.13235
Descripción
Sumario:A 73-year-old male with a history of severe coronary artery disease and prior coronary artery bypass grafting (CABG) presented with chest pain and elevated troponins. His workup revealed an ejection fraction of 15%, severe native coronary disease, as well as stenosis of his bypass grafts. He underwent a high-risk redo CABG with an Impella 5.5® (Abiomed, Danvers, MA) placement. The device was removed on postoperative day eight, at which time he went into cardiogenic shock from aortic valve leaflet injury. Given that he had no calcium deposits around the aortic valve annulus and severe aortic insufficiency, a multidisciplinary heart team decided he would be best served by a surgical aortic valve replacement. He was taken back to the operating room for a surgical aortic valve and intra-aortic balloon pump. His postoperative course was complicated by pneumonia, sepsis, and renal failure requiring continuous renal replacement therapy. He was discharged to a rehabilitation facility after 42 days. The following case encompasses the high morbidity risk of acute aortic valve insufficiency after Impella placement, never before documented in an Impella 5.5.