Cargando…
Comparison of seven different reagents of peroxidase method for small and dense low density lipoprotein cholesterol (sdLDL‐C) measurement
BACKGROUND: We validated the performance of seven different reagents of peroxidase method for sdLDL‐C in two automatic analyzers that are common in Chinese laboratories. METHODS: Seven commercially available sdLDL‐C assays were analyzed with the Beckman AU5400 and Mindray BS2000 automatic analyzers....
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7957989/ https://www.ncbi.nlm.nih.gov/pubmed/33377258 http://dx.doi.org/10.1002/jcla.23660 |
_version_ | 1783664763920711680 |
---|---|
author | Xuesong, Fan Enshi, Wang Jianxun, He Lei, Zhang Xiaoli, Zeng Hui, Yuan |
author_facet | Xuesong, Fan Enshi, Wang Jianxun, He Lei, Zhang Xiaoli, Zeng Hui, Yuan |
author_sort | Xuesong, Fan |
collection | PubMed |
description | BACKGROUND: We validated the performance of seven different reagents of peroxidase method for sdLDL‐C in two automatic analyzers that are common in Chinese laboratories. METHODS: Seven commercially available sdLDL‐C assays were analyzed with the Beckman AU5400 and Mindray BS2000 automatic analyzers. A total of 336 blood samples were collected and the reference interval was also validated in 298 apparently healthy individuals. Serum samples were used for method comparison of precision, recovery, lower limit of detection, comparison and concurrence analysis, as well as reference interval for the Mindray reagent. RESULTS: The repeatability CV% of the seven sdLDL‐C assays were 0.81%~3.66% for Mindray BS2000 and 0.76%~3.91% for Beckman AU5400, while Total CVs for Mindray BS2000 sdLDL‐C assay were 1.34%~4.81%, and that of Beckman AU5400 were 2.25%~10.33%. The measured recovery rates of sdLDL‐C assays were within the allowable ±10% deviation range. There was no obvious difference between the reagents in the lower limit detection. There was a difference between the validation results of the reference range and the manufacturer's.BSBE, Mindray, and Dongou had a high degree of association with DENKA SEIKEN on Mindray BS2000, while BSBE, Mindray, Dongou and Merit Choice had a high degree of association with DENKA SEIKEN on Beckman AU5400. Passing–Bablok regression showed excellent linear correlation between BSBE and Mindray and DENKA SEIKEN and on Beckman AU5400. CONCLUSIONS: Our results indicate that the basic performance can meet the testing requirements, but the comparability between them is still insufficient. |
format | Online Article Text |
id | pubmed-7957989 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-79579892021-03-19 Comparison of seven different reagents of peroxidase method for small and dense low density lipoprotein cholesterol (sdLDL‐C) measurement Xuesong, Fan Enshi, Wang Jianxun, He Lei, Zhang Xiaoli, Zeng Hui, Yuan J Clin Lab Anal Research Articles BACKGROUND: We validated the performance of seven different reagents of peroxidase method for sdLDL‐C in two automatic analyzers that are common in Chinese laboratories. METHODS: Seven commercially available sdLDL‐C assays were analyzed with the Beckman AU5400 and Mindray BS2000 automatic analyzers. A total of 336 blood samples were collected and the reference interval was also validated in 298 apparently healthy individuals. Serum samples were used for method comparison of precision, recovery, lower limit of detection, comparison and concurrence analysis, as well as reference interval for the Mindray reagent. RESULTS: The repeatability CV% of the seven sdLDL‐C assays were 0.81%~3.66% for Mindray BS2000 and 0.76%~3.91% for Beckman AU5400, while Total CVs for Mindray BS2000 sdLDL‐C assay were 1.34%~4.81%, and that of Beckman AU5400 were 2.25%~10.33%. The measured recovery rates of sdLDL‐C assays were within the allowable ±10% deviation range. There was no obvious difference between the reagents in the lower limit detection. There was a difference between the validation results of the reference range and the manufacturer's.BSBE, Mindray, and Dongou had a high degree of association with DENKA SEIKEN on Mindray BS2000, while BSBE, Mindray, Dongou and Merit Choice had a high degree of association with DENKA SEIKEN on Beckman AU5400. Passing–Bablok regression showed excellent linear correlation between BSBE and Mindray and DENKA SEIKEN and on Beckman AU5400. CONCLUSIONS: Our results indicate that the basic performance can meet the testing requirements, but the comparability between them is still insufficient. John Wiley and Sons Inc. 2020-12-29 /pmc/articles/PMC7957989/ /pubmed/33377258 http://dx.doi.org/10.1002/jcla.23660 Text en © 2020 The Authors. Journal of Clinical Laboratory Analysis published by Wiley Periodicals LLC. This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc-nd/4.0/ License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made. |
spellingShingle | Research Articles Xuesong, Fan Enshi, Wang Jianxun, He Lei, Zhang Xiaoli, Zeng Hui, Yuan Comparison of seven different reagents of peroxidase method for small and dense low density lipoprotein cholesterol (sdLDL‐C) measurement |
title | Comparison of seven different reagents of peroxidase method for small and dense low density lipoprotein cholesterol (sdLDL‐C) measurement |
title_full | Comparison of seven different reagents of peroxidase method for small and dense low density lipoprotein cholesterol (sdLDL‐C) measurement |
title_fullStr | Comparison of seven different reagents of peroxidase method for small and dense low density lipoprotein cholesterol (sdLDL‐C) measurement |
title_full_unstemmed | Comparison of seven different reagents of peroxidase method for small and dense low density lipoprotein cholesterol (sdLDL‐C) measurement |
title_short | Comparison of seven different reagents of peroxidase method for small and dense low density lipoprotein cholesterol (sdLDL‐C) measurement |
title_sort | comparison of seven different reagents of peroxidase method for small and dense low density lipoprotein cholesterol (sdldl‐c) measurement |
topic | Research Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7957989/ https://www.ncbi.nlm.nih.gov/pubmed/33377258 http://dx.doi.org/10.1002/jcla.23660 |
work_keys_str_mv | AT xuesongfan comparisonofsevendifferentreagentsofperoxidasemethodforsmallanddenselowdensitylipoproteincholesterolsdldlcmeasurement AT enshiwang comparisonofsevendifferentreagentsofperoxidasemethodforsmallanddenselowdensitylipoproteincholesterolsdldlcmeasurement AT jianxunhe comparisonofsevendifferentreagentsofperoxidasemethodforsmallanddenselowdensitylipoproteincholesterolsdldlcmeasurement AT leizhang comparisonofsevendifferentreagentsofperoxidasemethodforsmallanddenselowdensitylipoproteincholesterolsdldlcmeasurement AT xiaolizeng comparisonofsevendifferentreagentsofperoxidasemethodforsmallanddenselowdensitylipoproteincholesterolsdldlcmeasurement AT huiyuan comparisonofsevendifferentreagentsofperoxidasemethodforsmallanddenselowdensitylipoproteincholesterolsdldlcmeasurement |