Safety of Red Blood Cell Transfusion Using Small Central Lines in Neonates: An in vitro Non-inferiority Study
Aim: This study aimed to investigate the safety of transfusing red blood cell concentrates (RBCCs) through small [24 gauge (24G)] and extra-small [28 gauge [28G)] peripherally inserted central catheters (PICCs), according to guidelines of transfusion practice in Switzerland. Methods: We performed a...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Frontiers Media S.A.
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7968454/ https://www.ncbi.nlm.nih.gov/pubmed/33748040 http://dx.doi.org/10.3389/fped.2021.606611 |
_version_ | 1783666063161950208 |
---|---|
author | Rosa-Mangeret, Flavia Waldvogel-Abramowski, Sophie Pfister, Riccardo E. Baud, Olivier Fau, Sébastien |
author_facet | Rosa-Mangeret, Flavia Waldvogel-Abramowski, Sophie Pfister, Riccardo E. Baud, Olivier Fau, Sébastien |
author_sort | Rosa-Mangeret, Flavia |
collection | PubMed |
description | Aim: This study aimed to investigate the safety of transfusing red blood cell concentrates (RBCCs) through small [24 gauge (24G)] and extra-small [28 gauge [28G)] peripherally inserted central catheters (PICCs), according to guidelines of transfusion practice in Switzerland. Methods: We performed a non-inferiority in vitro study to assess the safety of transfusing RBCC for 4 h at a 4 ml/h speed through 24G silicone and 28G polyurethane PICC lines, compared with a peripheral 24G short catheter. The primary endpoint was hemolysis percentage. Secondary endpoints were catheter occlusion, inline pressure, and potassium and lactate values. Results: For the primary outcome, hemolysis values were not statistically different among catheter groups (0.06% variation, p = 0.95) or over time (2.75% variation, p = 0.72). The highest hemolysis values in both 24G and 28G PICCs were below the non-inferiority predefined margin. We did not observe catheter occlusion. Inline pressure varied between catheters but followed the same pattern of rapid increase followed by stabilization. Potassium and lactate measurements were not statistically different among tested catheters (0.139% variation, p = 0.98 for potassium and 0.062%, p = 0.96 for lactates). Conclusions: This study shows that RBCC transfusion performed in vitro through 24G silicone and 28G polyurethane PICC lines is feasible without detectable hemolysis or pressure concerns. Also, it adds that, concerning hemolysis, transfusion of RBCC in small and extra-small PICC lines is non-inferior to peripheral short 24G catheters. Clinical prospective assessment in preterm infants is needed to confirm these data further. |
format | Online Article Text |
id | pubmed-7968454 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Frontiers Media S.A. |
record_format | MEDLINE/PubMed |
spelling | pubmed-79684542021-03-18 Safety of Red Blood Cell Transfusion Using Small Central Lines in Neonates: An in vitro Non-inferiority Study Rosa-Mangeret, Flavia Waldvogel-Abramowski, Sophie Pfister, Riccardo E. Baud, Olivier Fau, Sébastien Front Pediatr Pediatrics Aim: This study aimed to investigate the safety of transfusing red blood cell concentrates (RBCCs) through small [24 gauge (24G)] and extra-small [28 gauge [28G)] peripherally inserted central catheters (PICCs), according to guidelines of transfusion practice in Switzerland. Methods: We performed a non-inferiority in vitro study to assess the safety of transfusing RBCC for 4 h at a 4 ml/h speed through 24G silicone and 28G polyurethane PICC lines, compared with a peripheral 24G short catheter. The primary endpoint was hemolysis percentage. Secondary endpoints were catheter occlusion, inline pressure, and potassium and lactate values. Results: For the primary outcome, hemolysis values were not statistically different among catheter groups (0.06% variation, p = 0.95) or over time (2.75% variation, p = 0.72). The highest hemolysis values in both 24G and 28G PICCs were below the non-inferiority predefined margin. We did not observe catheter occlusion. Inline pressure varied between catheters but followed the same pattern of rapid increase followed by stabilization. Potassium and lactate measurements were not statistically different among tested catheters (0.139% variation, p = 0.98 for potassium and 0.062%, p = 0.96 for lactates). Conclusions: This study shows that RBCC transfusion performed in vitro through 24G silicone and 28G polyurethane PICC lines is feasible without detectable hemolysis or pressure concerns. Also, it adds that, concerning hemolysis, transfusion of RBCC in small and extra-small PICC lines is non-inferior to peripheral short 24G catheters. Clinical prospective assessment in preterm infants is needed to confirm these data further. Frontiers Media S.A. 2021-03-03 /pmc/articles/PMC7968454/ /pubmed/33748040 http://dx.doi.org/10.3389/fped.2021.606611 Text en Copyright © 2021 Rosa-Mangeret, Waldvogel-Abramowski, Pfister, Baud and Fau. http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms. |
spellingShingle | Pediatrics Rosa-Mangeret, Flavia Waldvogel-Abramowski, Sophie Pfister, Riccardo E. Baud, Olivier Fau, Sébastien Safety of Red Blood Cell Transfusion Using Small Central Lines in Neonates: An in vitro Non-inferiority Study |
title | Safety of Red Blood Cell Transfusion Using Small Central Lines in Neonates: An in vitro Non-inferiority Study |
title_full | Safety of Red Blood Cell Transfusion Using Small Central Lines in Neonates: An in vitro Non-inferiority Study |
title_fullStr | Safety of Red Blood Cell Transfusion Using Small Central Lines in Neonates: An in vitro Non-inferiority Study |
title_full_unstemmed | Safety of Red Blood Cell Transfusion Using Small Central Lines in Neonates: An in vitro Non-inferiority Study |
title_short | Safety of Red Blood Cell Transfusion Using Small Central Lines in Neonates: An in vitro Non-inferiority Study |
title_sort | safety of red blood cell transfusion using small central lines in neonates: an in vitro non-inferiority study |
topic | Pediatrics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7968454/ https://www.ncbi.nlm.nih.gov/pubmed/33748040 http://dx.doi.org/10.3389/fped.2021.606611 |
work_keys_str_mv | AT rosamangeretflavia safetyofredbloodcelltransfusionusingsmallcentrallinesinneonatesaninvitrononinferioritystudy AT waldvogelabramowskisophie safetyofredbloodcelltransfusionusingsmallcentrallinesinneonatesaninvitrononinferioritystudy AT pfisterriccardoe safetyofredbloodcelltransfusionusingsmallcentrallinesinneonatesaninvitrononinferioritystudy AT baudolivier safetyofredbloodcelltransfusionusingsmallcentrallinesinneonatesaninvitrononinferioritystudy AT fausebastien safetyofredbloodcelltransfusionusingsmallcentrallinesinneonatesaninvitrononinferioritystudy |