Validation of CARE kV automated tube voltage selection for PET-CT: PET quantification and CT radiation dose reduction in phantoms

BACKGROUND: Applied tube voltage (kilovolts, kV) and tube current (milliampere seconds, mAs) affect CT radiation dose and image quality and should be optimised for the individual patient. CARE kV determines the kV and mAs providing the lowest dose to the patient, whilst maintaining user-defined refe...

Descripción completa

Detalles Bibliográficos
Autores principales: Bebbington, Natalie A., Jørgensen, Troels, Dupont, Erik, Micheelsen, Mille A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7981373/
https://www.ncbi.nlm.nih.gov/pubmed/33743091
http://dx.doi.org/10.1186/s40658-021-00373-8
_version_ 1783667536419618816
author Bebbington, Natalie A.
Jørgensen, Troels
Dupont, Erik
Micheelsen, Mille A.
author_facet Bebbington, Natalie A.
Jørgensen, Troels
Dupont, Erik
Micheelsen, Mille A.
author_sort Bebbington, Natalie A.
collection PubMed
description BACKGROUND: Applied tube voltage (kilovolts, kV) and tube current (milliampere seconds, mAs) affect CT radiation dose and image quality and should be optimised for the individual patient. CARE kV determines the kV and mAs providing the lowest dose to the patient, whilst maintaining user-defined reference image quality. Given that kV changes affect CT values which are used to obtain attenuation maps, the aim was to evaluate the effect of kV changes on PET quantification and CT radiation dose using phantoms. METHOD: Four phantoms (‘Lungman’, ‘Lungman plus fat’, ‘Esser’ and ‘NEMA image quality’ (NEMA IQ)) containing F-18 sources underwent 1 PET and 5 CT scans, with CARE kV on (automatic kV selection and mAs modulation) and in semi mode with specified tube voltages of 140, 120, 100 and 80 kV (mAs modulation only). A CARE kV image quality reference of 120 kV/50 mAs was used. Impact on PET quantification was determined by comparing measured activity concentrations for PET reconstructions from different CT scans with the reconstruction using the 120 kV reference, and dose (DLP, CTDI(vol)) differences calculated by comparing doses from all kV settings with the 120 kV reference. RESULTS: CARE kV-determined optimal tube voltage and CARE kV ‘on’ dose (DLP) savings compared with the 120 kV reference were: Lungman, 100 kV, 2.0%; Lungman plus fat, 120 kV, 0%; Esser, 100 kV, 9.3%; NEMA IQ, 100 kV, 3.4%. Using tube voltages in CARE kV ‘semi’ mode which were not advised by CARE kV ‘on’ resulted in dose increases ≤ 65% compared with the 120 kV reference (greatest difference Lungman plus fat, 80 kV). Clinically insignificant differences in PET activity quantification of up to 0.7% (Lungman, 100 kV, mean measured activity concentration) were observed when using the optimal tube voltage advised by CARE kV. Differences in PET quantification of up to 4.0% (Lungman, 140 kV, maximum measured activity concentration) were found over the full selection of tube voltages in semi mode, with the greatest differences seen at the most suboptimal kV for each phantom. However, most differences were within 1%. CONCLUSIONS: CARE kV on can provide CT radiation dose savings without concern over changes in PET quantification.
format Online
Article
Text
id pubmed-7981373
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-79813732021-04-12 Validation of CARE kV automated tube voltage selection for PET-CT: PET quantification and CT radiation dose reduction in phantoms Bebbington, Natalie A. Jørgensen, Troels Dupont, Erik Micheelsen, Mille A. EJNMMI Phys Original Research BACKGROUND: Applied tube voltage (kilovolts, kV) and tube current (milliampere seconds, mAs) affect CT radiation dose and image quality and should be optimised for the individual patient. CARE kV determines the kV and mAs providing the lowest dose to the patient, whilst maintaining user-defined reference image quality. Given that kV changes affect CT values which are used to obtain attenuation maps, the aim was to evaluate the effect of kV changes on PET quantification and CT radiation dose using phantoms. METHOD: Four phantoms (‘Lungman’, ‘Lungman plus fat’, ‘Esser’ and ‘NEMA image quality’ (NEMA IQ)) containing F-18 sources underwent 1 PET and 5 CT scans, with CARE kV on (automatic kV selection and mAs modulation) and in semi mode with specified tube voltages of 140, 120, 100 and 80 kV (mAs modulation only). A CARE kV image quality reference of 120 kV/50 mAs was used. Impact on PET quantification was determined by comparing measured activity concentrations for PET reconstructions from different CT scans with the reconstruction using the 120 kV reference, and dose (DLP, CTDI(vol)) differences calculated by comparing doses from all kV settings with the 120 kV reference. RESULTS: CARE kV-determined optimal tube voltage and CARE kV ‘on’ dose (DLP) savings compared with the 120 kV reference were: Lungman, 100 kV, 2.0%; Lungman plus fat, 120 kV, 0%; Esser, 100 kV, 9.3%; NEMA IQ, 100 kV, 3.4%. Using tube voltages in CARE kV ‘semi’ mode which were not advised by CARE kV ‘on’ resulted in dose increases ≤ 65% compared with the 120 kV reference (greatest difference Lungman plus fat, 80 kV). Clinically insignificant differences in PET activity quantification of up to 0.7% (Lungman, 100 kV, mean measured activity concentration) were observed when using the optimal tube voltage advised by CARE kV. Differences in PET quantification of up to 4.0% (Lungman, 140 kV, maximum measured activity concentration) were found over the full selection of tube voltages in semi mode, with the greatest differences seen at the most suboptimal kV for each phantom. However, most differences were within 1%. CONCLUSIONS: CARE kV on can provide CT radiation dose savings without concern over changes in PET quantification. Springer International Publishing 2021-03-20 /pmc/articles/PMC7981373/ /pubmed/33743091 http://dx.doi.org/10.1186/s40658-021-00373-8 Text en © The Author(s) 2021 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
spellingShingle Original Research
Bebbington, Natalie A.
Jørgensen, Troels
Dupont, Erik
Micheelsen, Mille A.
Validation of CARE kV automated tube voltage selection for PET-CT: PET quantification and CT radiation dose reduction in phantoms
title Validation of CARE kV automated tube voltage selection for PET-CT: PET quantification and CT radiation dose reduction in phantoms
title_full Validation of CARE kV automated tube voltage selection for PET-CT: PET quantification and CT radiation dose reduction in phantoms
title_fullStr Validation of CARE kV automated tube voltage selection for PET-CT: PET quantification and CT radiation dose reduction in phantoms
title_full_unstemmed Validation of CARE kV automated tube voltage selection for PET-CT: PET quantification and CT radiation dose reduction in phantoms
title_short Validation of CARE kV automated tube voltage selection for PET-CT: PET quantification and CT radiation dose reduction in phantoms
title_sort validation of care kv automated tube voltage selection for pet-ct: pet quantification and ct radiation dose reduction in phantoms
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7981373/
https://www.ncbi.nlm.nih.gov/pubmed/33743091
http://dx.doi.org/10.1186/s40658-021-00373-8
work_keys_str_mv AT bebbingtonnataliea validationofcarekvautomatedtubevoltageselectionforpetctpetquantificationandctradiationdosereductioninphantoms
AT jørgensentroels validationofcarekvautomatedtubevoltageselectionforpetctpetquantificationandctradiationdosereductioninphantoms
AT duponterik validationofcarekvautomatedtubevoltageselectionforpetctpetquantificationandctradiationdosereductioninphantoms
AT micheelsenmillea validationofcarekvautomatedtubevoltageselectionforpetctpetquantificationandctradiationdosereductioninphantoms