Intra‐fractional patient motion when using the Qfix Encompass immobilization system during HyperArc treatment of patients with brain metastases

PURPOSE: This study investigated the intra‐fractional motion (IM) of patients immobilized using the QFix Encompass Immobilization System during HyperArc (HA) treatment. METHOD: HA treatment was performed on 89 patients immobilized using the Encompass. The IM during treatment (including megavoltage (...

Descripción completa

Detalles Bibliográficos
Autores principales: Komiyama, Riho, Ohira, Shingo, Ueda, Hikari, Kanayama, Naoyuki, Masaoka, Akira, Isono, Masaru, Ueda, Yoshihiro, Miyazaki, Masayoshi, Teshima, Teruki
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7984469/
https://www.ncbi.nlm.nih.gov/pubmed/33656261
http://dx.doi.org/10.1002/acm2.13143
Descripción
Sumario:PURPOSE: This study investigated the intra‐fractional motion (IM) of patients immobilized using the QFix Encompass Immobilization System during HyperArc (HA) treatment. METHOD: HA treatment was performed on 89 patients immobilized using the Encompass. The IM during treatment (including megavoltage (MV) registration) was analyzed for six degrees of freedom including three axes of translation (anterior‐posterior, superior‐inferior (SI) and left‐right (LR)) and three axes of rotation (pitch, roll, and yaw). Then, the no corrected IM (IM(NC)) was retrospectively simulated (excluding MV registration) in three directions (SI, LR, and yaw). Finally, the correlation between the treatment time and the IM of the 3D vector was assessed. RESULTS: The average IM in terms of the absolute displacement were 0.3 mm (SI), 0.3 mm (LR) and 0.2° (yaw) for Stereotactic radiosurgery (SRS), and 0.3 mm (SI), 0.2 mm (LR), and 0.2° (yaw) for stereotactic radiotherapy (SRT). The absolute maximum values of IM were <1 mm along the SI and LR axes and <1° along the yaw axis. The absolute maximum displacements for IM(NC) were >1 mm along the SI and LR axes and >1° along the yaw axis. In the correlation between the treatment time and the IM, the r‐values were −0.025 and 0.027 for SRS and SRT respectively, along the axes of translation. For the axes of rotation, the r‐values were 0.012 and 0.206 for SRS and SRT, respectively. CONCLUSION: Encompass provided patient immobilization with adequate accuracy during HA treatment. The absolute maximum displacement IM was less than IM(NC) along the translational/rotational axes, and no statistically significant relationship between the treatment time and the IM was observed.