The use of trajectory log files for routine patient quality assurance is gaining acceptance. Such use requires the validation of the trajectory log itself. However, the accurate localization of a multileaf collimator (MLC) leaf while it is in motion remains a challenging task. We propose an efficien...

Descripción completa

Detalles Bibliográficos
Autores principales: Lim, Seng Boh, Zwan, Benjamin J., Lee, Danny, Greer, Peter B., Lovelock, Dale Michael
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7984475/
https://www.ncbi.nlm.nih.gov/pubmed/33634952
http://dx.doi.org/10.1002/acm2.13202
_version_ 1783668072318500864
author Lim, Seng Boh
Zwan, Benjamin J.
Lee, Danny
Greer, Peter B.
Lovelock, Dale Michael
author_facet Lim, Seng Boh
Zwan, Benjamin J.
Lee, Danny
Greer, Peter B.
Lovelock, Dale Michael
author_sort Lim, Seng Boh
collection PubMed
description The use of trajectory log files for routine patient quality assurance is gaining acceptance. Such use requires the validation of the trajectory log itself. However, the accurate localization of a multileaf collimator (MLC) leaf while it is in motion remains a challenging task. We propose an efficient phantom‐less technique using the EPID to verify the dynamic MLC positions with high accuracy. Measurements were made on four Varian TrueBeams equipped with M120 MLCs. Two machines were equipped with the S1000 EPID; two were equipped with the S1200 EPID. All EPIDs were geometrically corrected prior to measurements. Dosimetry mode EPID measurements were captured by a frame grabber card directly linked to the linac. All leaf position measurements were corrected both temporally and geometrically. The readout latency of each panel, as a function of pixel row, was determined using a 40 × 1.0 cm(2) sliding window (SW) field moving at 2.5 cm/s orthogonal to the row readout direction. The latency of each panel type was determined by averaging the results of two panels of the same type. Geometric correction was achieved by computing leaf positions with respect to the projected isocenter position as a function of gantry angle. This was determined by averaging the central axis position of fields at two collimator positions of 90° and 270°. The radiological to physical leaf end position was determined by comparison of the measured gap with that determined using a feeler gauge. The radiological to physical leaf position difference was found to be 0.1 mm. With geometric and latency correction, the proposed method was found to be improve the ability to detect dynamic MLC positions from 1.0 to 0.2 mm for all leaves. Latency and panel residual geometric error correction improve EPID‐based MLC position measurement. These improvements provide for the first time a trajectory log QA procedure.
format Online
Article
Text
id pubmed-7984475
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-79844752021-03-25 A novel quality assurance procedure for trajectory log validation using phantom‐less real‐time latency corrected EPID images Lim, Seng Boh Zwan, Benjamin J. Lee, Danny Greer, Peter B. Lovelock, Dale Michael J Appl Clin Med Phys Radiation Oncology Physics The use of trajectory log files for routine patient quality assurance is gaining acceptance. Such use requires the validation of the trajectory log itself. However, the accurate localization of a multileaf collimator (MLC) leaf while it is in motion remains a challenging task. We propose an efficient phantom‐less technique using the EPID to verify the dynamic MLC positions with high accuracy. Measurements were made on four Varian TrueBeams equipped with M120 MLCs. Two machines were equipped with the S1000 EPID; two were equipped with the S1200 EPID. All EPIDs were geometrically corrected prior to measurements. Dosimetry mode EPID measurements were captured by a frame grabber card directly linked to the linac. All leaf position measurements were corrected both temporally and geometrically. The readout latency of each panel, as a function of pixel row, was determined using a 40 × 1.0 cm(2) sliding window (SW) field moving at 2.5 cm/s orthogonal to the row readout direction. The latency of each panel type was determined by averaging the results of two panels of the same type. Geometric correction was achieved by computing leaf positions with respect to the projected isocenter position as a function of gantry angle. This was determined by averaging the central axis position of fields at two collimator positions of 90° and 270°. The radiological to physical leaf end position was determined by comparison of the measured gap with that determined using a feeler gauge. The radiological to physical leaf position difference was found to be 0.1 mm. With geometric and latency correction, the proposed method was found to be improve the ability to detect dynamic MLC positions from 1.0 to 0.2 mm for all leaves. Latency and panel residual geometric error correction improve EPID‐based MLC position measurement. These improvements provide for the first time a trajectory log QA procedure. John Wiley and Sons Inc. 2021-02-26 /pmc/articles/PMC7984475/ /pubmed/33634952 http://dx.doi.org/10.1002/acm2.13202 Text en © 2021 The Authors. Journal of Applied Clinical Medical Physics is published by Wiley Periodicals LLC. on behalf of The American Association of Physicists in Medicine This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Lim, Seng Boh
Zwan, Benjamin J.
Lee, Danny
Greer, Peter B.
Lovelock, Dale Michael
A novel quality assurance procedure for trajectory log validation using phantom‐less real‐time latency corrected EPID images
title A novel quality assurance procedure for trajectory log validation using phantom‐less real‐time latency corrected EPID images
title_full A novel quality assurance procedure for trajectory log validation using phantom‐less real‐time latency corrected EPID images
title_fullStr A novel quality assurance procedure for trajectory log validation using phantom‐less real‐time latency corrected EPID images
title_full_unstemmed A novel quality assurance procedure for trajectory log validation using phantom‐less real‐time latency corrected EPID images
title_short A novel quality assurance procedure for trajectory log validation using phantom‐less real‐time latency corrected EPID images
title_sort novel quality assurance procedure for trajectory log validation using phantom‐less real‐time latency corrected epid images
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7984475/
https://www.ncbi.nlm.nih.gov/pubmed/33634952
http://dx.doi.org/10.1002/acm2.13202
work_keys_str_mv AT limsengboh anovelqualityassuranceprocedurefortrajectorylogvalidationusingphantomlessrealtimelatencycorrectedepidimages
AT zwanbenjaminj anovelqualityassuranceprocedurefortrajectorylogvalidationusingphantomlessrealtimelatencycorrectedepidimages
AT leedanny anovelqualityassuranceprocedurefortrajectorylogvalidationusingphantomlessrealtimelatencycorrectedepidimages
AT greerpeterb anovelqualityassuranceprocedurefortrajectorylogvalidationusingphantomlessrealtimelatencycorrectedepidimages
AT lovelockdalemichael anovelqualityassuranceprocedurefortrajectorylogvalidationusingphantomlessrealtimelatencycorrectedepidimages
AT limsengboh novelqualityassuranceprocedurefortrajectorylogvalidationusingphantomlessrealtimelatencycorrectedepidimages
AT zwanbenjaminj novelqualityassuranceprocedurefortrajectorylogvalidationusingphantomlessrealtimelatencycorrectedepidimages
AT leedanny novelqualityassuranceprocedurefortrajectorylogvalidationusingphantomlessrealtimelatencycorrectedepidimages
AT greerpeterb novelqualityassuranceprocedurefortrajectorylogvalidationusingphantomlessrealtimelatencycorrectedepidimages
AT lovelockdalemichael novelqualityassuranceprocedurefortrajectorylogvalidationusingphantomlessrealtimelatencycorrectedepidimages