Modeling of kyphoplasty cement for accurate dose calculations

We have determined the optimal method for modeling kyphoplasty cement to enable accurate dose calculations in the Eclipse treatment planning system (TPS). The cement studied (Medtronic Kyphon HV‐R®) consists of 30% Barium, 68% polymethylmethacrylate (PMMA), and 2% benzoyl peroxide, formulated to be...

Descripción completa

Detalles Bibliográficos
Autores principales: Halvorsen, Per H., Hariharan, Navneeth, Morelli, Zackary T., Iftimia, Ileana N.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7984498/
https://www.ncbi.nlm.nih.gov/pubmed/33599374
http://dx.doi.org/10.1002/acm2.13174
_version_ 1783668077621149696
author Halvorsen, Per H.
Hariharan, Navneeth
Morelli, Zackary T.
Iftimia, Ileana N.
author_facet Halvorsen, Per H.
Hariharan, Navneeth
Morelli, Zackary T.
Iftimia, Ileana N.
author_sort Halvorsen, Per H.
collection PubMed
description We have determined the optimal method for modeling kyphoplasty cement to enable accurate dose calculations in the Eclipse treatment planning system (TPS). The cement studied (Medtronic Kyphon HV‐R®) consists of 30% Barium, 68% polymethylmethacrylate (PMMA), and 2% benzoyl peroxide, formulated to be radiopaque with kV imaging systems. Neither Barium nor PMMA have a high physical density, resulting in different interaction characteristics for megavoltage treatment beams compared to kV imaging systems. This can lead to significant calculation errors if density mapping is performed using a standard CT number to density curve. To properly characterize the cement for dose calculation, we 3D printed a hemi‐cylindrical container to fit adjacent to a micro‐chamber insert for an anthropomorphic phantom, and filled the container with Kyphon cement. We CT scanned the combination, modeled the cement with multiple material assignments in the TPS, designed plans with different field sizes and beam geometry for five photon modes, and measured the doses for all plans. All photon energies show significant error in calculated dose when the cement is modeled based on the CT number. Of the material assignments we evaluated, polytetrafluoroethylene (PTFE) showed the best overall agreement with measurement. Calculated and measured doses agree within 3.5% for a 340‐degree arc technique (which averages transmission and scatter effects) with the Acuros XB algorithm and PTFE as the assigned material. To confirm that PTFE is a reasonable substitute for kyphoplasty cement, we performed measurements in a slab phantom using rectangular inserts of cement and PTFE, showing average agreement of all photon modes within 2%. Based on these findings, we conclude that the PTFE material assignment provides acceptable dose calculation accuracy for the AAA and Acuros XB photon algorithms in the Eclipse TPS. We recommend that the cement be delineated as a structure and assigned the PTFE material for accurate dose calculation.
format Online
Article
Text
id pubmed-7984498
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-79844982021-03-25 Modeling of kyphoplasty cement for accurate dose calculations Halvorsen, Per H. Hariharan, Navneeth Morelli, Zackary T. Iftimia, Ileana N. J Appl Clin Med Phys Technical Notes We have determined the optimal method for modeling kyphoplasty cement to enable accurate dose calculations in the Eclipse treatment planning system (TPS). The cement studied (Medtronic Kyphon HV‐R®) consists of 30% Barium, 68% polymethylmethacrylate (PMMA), and 2% benzoyl peroxide, formulated to be radiopaque with kV imaging systems. Neither Barium nor PMMA have a high physical density, resulting in different interaction characteristics for megavoltage treatment beams compared to kV imaging systems. This can lead to significant calculation errors if density mapping is performed using a standard CT number to density curve. To properly characterize the cement for dose calculation, we 3D printed a hemi‐cylindrical container to fit adjacent to a micro‐chamber insert for an anthropomorphic phantom, and filled the container with Kyphon cement. We CT scanned the combination, modeled the cement with multiple material assignments in the TPS, designed plans with different field sizes and beam geometry for five photon modes, and measured the doses for all plans. All photon energies show significant error in calculated dose when the cement is modeled based on the CT number. Of the material assignments we evaluated, polytetrafluoroethylene (PTFE) showed the best overall agreement with measurement. Calculated and measured doses agree within 3.5% for a 340‐degree arc technique (which averages transmission and scatter effects) with the Acuros XB algorithm and PTFE as the assigned material. To confirm that PTFE is a reasonable substitute for kyphoplasty cement, we performed measurements in a slab phantom using rectangular inserts of cement and PTFE, showing average agreement of all photon modes within 2%. Based on these findings, we conclude that the PTFE material assignment provides acceptable dose calculation accuracy for the AAA and Acuros XB photon algorithms in the Eclipse TPS. We recommend that the cement be delineated as a structure and assigned the PTFE material for accurate dose calculation. John Wiley and Sons Inc. 2021-02-18 /pmc/articles/PMC7984498/ /pubmed/33599374 http://dx.doi.org/10.1002/acm2.13174 Text en © 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technical Notes
Halvorsen, Per H.
Hariharan, Navneeth
Morelli, Zackary T.
Iftimia, Ileana N.
Modeling of kyphoplasty cement for accurate dose calculations
title Modeling of kyphoplasty cement for accurate dose calculations
title_full Modeling of kyphoplasty cement for accurate dose calculations
title_fullStr Modeling of kyphoplasty cement for accurate dose calculations
title_full_unstemmed Modeling of kyphoplasty cement for accurate dose calculations
title_short Modeling of kyphoplasty cement for accurate dose calculations
title_sort modeling of kyphoplasty cement for accurate dose calculations
topic Technical Notes
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7984498/
https://www.ncbi.nlm.nih.gov/pubmed/33599374
http://dx.doi.org/10.1002/acm2.13174
work_keys_str_mv AT halvorsenperh modelingofkyphoplastycementforaccuratedosecalculations
AT hariharannavneeth modelingofkyphoplastycementforaccuratedosecalculations
AT morellizackaryt modelingofkyphoplastycementforaccuratedosecalculations
AT iftimiaileanan modelingofkyphoplastycementforaccuratedosecalculations