A Novel Method Based on Headspace-Ion Mobility Spectrometry for the Detection and Discrimination of Different Petroleum Derived Products in Seawater

The objective of the present study is to develop an optimized method where headspace-ion mobility spectrometry is applied for the detection and discrimination between four petroleum-derived products (PDPs) in water. A Box–Behnken design with a response surface methodology was used, and five variable...

Descripción completa

Detalles Bibliográficos
Autores principales: Jaén-González, Lucas, Aliaño-González, Ma José, Ferreiro-González, Marta, Barbero, Gerardo F., Palma, Miguel
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8003363/
https://www.ncbi.nlm.nih.gov/pubmed/33808571
http://dx.doi.org/10.3390/s21062151
Descripción
Sumario:The objective of the present study is to develop an optimized method where headspace-ion mobility spectrometry is applied for the detection and discrimination between four petroleum-derived products (PDPs) in water. A Box–Behnken design with a response surface methodology was used, and five variables (incubation temperature, incubation time, agitation, sample volume, and injection volume) with influences on the ion mobility spectrometry (IMS) response were optimized. An IMS detector was used as a multiple sensor device, in which, each drift time acts as a specific sensor. In this way, the total intensity at each drift time is equivalent to multiple sensor signals. According to our results, 2.5 mL of sample incubated for 5 min at 31 °C, agitated at 750 rpm, and with an injection volume of 0.91 mL were the optimal conditions for successful detection and discrimination of the PDPs. The developed method has exhibited good intermediate precision and repeatability with a coefficient of variation lower than 5%, (RSD (Relative Standard Deviation): 2.35% and 3.09%, respectively). Subsequently, the method was applied in the context of the detection and discrimination of petroleum-derived products added to water samples at low concentration levels (2 µL·L(−1)). Finally, the new method was applied to determine the presence of petroleum-derived products in seawater samples.