Prospective evaluation of accuracy of HIV viral load monitoring using the Aptima HIV Quant Dx assay with fingerstick and venous dried blood spots prepared under field conditions in Kenya

Quantification of HIV-1 RNA is essential for clinical management of HIV patients. The limited throughput and significant hands-on time required by most HIV Viral load (VL) tests makes it challenging for laboratories with high test volume, to turn around patient results quickly. The Hologic Aptima HI...

Descripción completa

Detalles Bibliográficos
Autores principales: Mwau, Matilu, Danda, Jeff, Mbugua, Joseph, Handa, Allan, Fortunko, Jacqueline, Worlock, Andrew, Nair, Sangeetha Vijaysri
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8018649/
https://www.ncbi.nlm.nih.gov/pubmed/33798221
http://dx.doi.org/10.1371/journal.pone.0249376
_version_ 1783674235582939136
author Mwau, Matilu
Danda, Jeff
Mbugua, Joseph
Handa, Allan
Fortunko, Jacqueline
Worlock, Andrew
Nair, Sangeetha Vijaysri
author_facet Mwau, Matilu
Danda, Jeff
Mbugua, Joseph
Handa, Allan
Fortunko, Jacqueline
Worlock, Andrew
Nair, Sangeetha Vijaysri
author_sort Mwau, Matilu
collection PubMed
description Quantification of HIV-1 RNA is essential for clinical management of HIV patients. The limited throughput and significant hands-on time required by most HIV Viral load (VL) tests makes it challenging for laboratories with high test volume, to turn around patient results quickly. The Hologic Aptima HIV-1 Quant Dx Assay (Aptima), has the potential to alleviate this burden as it is high throughput and fully automated. This assay is validated for both plasma and dried blood spots (DBS), which are commonly used in resource limited settings. The objective of this study was to compare the performance of Aptima to Abbott RealTime HIV-1 Assay (Abbott RT), which was used as reference. This was a cross-sectional prospective study where HIV VL in finger stick (FS) DBS, venous blood (VB) DBS and plasma, collected from 258 consenting adults visiting 5 medical facilities in Kenya, Africa were tested in Aptima. The results were compared to plasma VL in Abbott RT at the medical decision point (MDP) of 1000 copies/mL and across Aptima assay range. The total agreement at MDP between plasma HIV VL in Abbott RT and plasma, FS and VB DBS tested in Aptima were 97.7%, 92.2% and 95.3% respectively with kappa statistic of 0.95, 0.84 and 0.90. The positive and negative agreement for all 3 sample types were >92%. Regression analysis between VL in Abbott RT plasma and various sample types tested in Aptima had a Pearson’s correlation coefficient ≥0.91 with systematic bias of < 0.20 log copies/mL on Bland-Altman analysis. The high level of agreement in Aptima HIV VL results for all 3 sample types with Abbott RT plasma VL along with the high throughput, complete automation, and ease of use of the Panther platform makes Aptima a good option for HIV VL monitoring for busy laboratories with high volume of testing.
format Online
Article
Text
id pubmed-8018649
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-80186492021-04-13 Prospective evaluation of accuracy of HIV viral load monitoring using the Aptima HIV Quant Dx assay with fingerstick and venous dried blood spots prepared under field conditions in Kenya Mwau, Matilu Danda, Jeff Mbugua, Joseph Handa, Allan Fortunko, Jacqueline Worlock, Andrew Nair, Sangeetha Vijaysri PLoS One Research Article Quantification of HIV-1 RNA is essential for clinical management of HIV patients. The limited throughput and significant hands-on time required by most HIV Viral load (VL) tests makes it challenging for laboratories with high test volume, to turn around patient results quickly. The Hologic Aptima HIV-1 Quant Dx Assay (Aptima), has the potential to alleviate this burden as it is high throughput and fully automated. This assay is validated for both plasma and dried blood spots (DBS), which are commonly used in resource limited settings. The objective of this study was to compare the performance of Aptima to Abbott RealTime HIV-1 Assay (Abbott RT), which was used as reference. This was a cross-sectional prospective study where HIV VL in finger stick (FS) DBS, venous blood (VB) DBS and plasma, collected from 258 consenting adults visiting 5 medical facilities in Kenya, Africa were tested in Aptima. The results were compared to plasma VL in Abbott RT at the medical decision point (MDP) of 1000 copies/mL and across Aptima assay range. The total agreement at MDP between plasma HIV VL in Abbott RT and plasma, FS and VB DBS tested in Aptima were 97.7%, 92.2% and 95.3% respectively with kappa statistic of 0.95, 0.84 and 0.90. The positive and negative agreement for all 3 sample types were >92%. Regression analysis between VL in Abbott RT plasma and various sample types tested in Aptima had a Pearson’s correlation coefficient ≥0.91 with systematic bias of < 0.20 log copies/mL on Bland-Altman analysis. The high level of agreement in Aptima HIV VL results for all 3 sample types with Abbott RT plasma VL along with the high throughput, complete automation, and ease of use of the Panther platform makes Aptima a good option for HIV VL monitoring for busy laboratories with high volume of testing. Public Library of Science 2021-04-02 /pmc/articles/PMC8018649/ /pubmed/33798221 http://dx.doi.org/10.1371/journal.pone.0249376 Text en © 2021 Mwau et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Mwau, Matilu
Danda, Jeff
Mbugua, Joseph
Handa, Allan
Fortunko, Jacqueline
Worlock, Andrew
Nair, Sangeetha Vijaysri
Prospective evaluation of accuracy of HIV viral load monitoring using the Aptima HIV Quant Dx assay with fingerstick and venous dried blood spots prepared under field conditions in Kenya
title Prospective evaluation of accuracy of HIV viral load monitoring using the Aptima HIV Quant Dx assay with fingerstick and venous dried blood spots prepared under field conditions in Kenya
title_full Prospective evaluation of accuracy of HIV viral load monitoring using the Aptima HIV Quant Dx assay with fingerstick and venous dried blood spots prepared under field conditions in Kenya
title_fullStr Prospective evaluation of accuracy of HIV viral load monitoring using the Aptima HIV Quant Dx assay with fingerstick and venous dried blood spots prepared under field conditions in Kenya
title_full_unstemmed Prospective evaluation of accuracy of HIV viral load monitoring using the Aptima HIV Quant Dx assay with fingerstick and venous dried blood spots prepared under field conditions in Kenya
title_short Prospective evaluation of accuracy of HIV viral load monitoring using the Aptima HIV Quant Dx assay with fingerstick and venous dried blood spots prepared under field conditions in Kenya
title_sort prospective evaluation of accuracy of hiv viral load monitoring using the aptima hiv quant dx assay with fingerstick and venous dried blood spots prepared under field conditions in kenya
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8018649/
https://www.ncbi.nlm.nih.gov/pubmed/33798221
http://dx.doi.org/10.1371/journal.pone.0249376
work_keys_str_mv AT mwaumatilu prospectiveevaluationofaccuracyofhivviralloadmonitoringusingtheaptimahivquantdxassaywithfingerstickandvenousdriedbloodspotspreparedunderfieldconditionsinkenya
AT dandajeff prospectiveevaluationofaccuracyofhivviralloadmonitoringusingtheaptimahivquantdxassaywithfingerstickandvenousdriedbloodspotspreparedunderfieldconditionsinkenya
AT mbuguajoseph prospectiveevaluationofaccuracyofhivviralloadmonitoringusingtheaptimahivquantdxassaywithfingerstickandvenousdriedbloodspotspreparedunderfieldconditionsinkenya
AT handaallan prospectiveevaluationofaccuracyofhivviralloadmonitoringusingtheaptimahivquantdxassaywithfingerstickandvenousdriedbloodspotspreparedunderfieldconditionsinkenya
AT fortunkojacqueline prospectiveevaluationofaccuracyofhivviralloadmonitoringusingtheaptimahivquantdxassaywithfingerstickandvenousdriedbloodspotspreparedunderfieldconditionsinkenya
AT worlockandrew prospectiveevaluationofaccuracyofhivviralloadmonitoringusingtheaptimahivquantdxassaywithfingerstickandvenousdriedbloodspotspreparedunderfieldconditionsinkenya
AT nairsangeethavijaysri prospectiveevaluationofaccuracyofhivviralloadmonitoringusingtheaptimahivquantdxassaywithfingerstickandvenousdriedbloodspotspreparedunderfieldconditionsinkenya