Toward a priori noise characterization for real-time edge-aware denoising in fluoroscopic devices

BACKGROUND: Low-dose X-ray images have become increasingly popular in the last decades, due to the need to guarantee the lowest reasonable patient’s exposure. Dose reduction causes a substantial increase of quantum noise, which needs to be suitably suppressed. In particular, real-time denoising is r...

Descripción completa

Detalles Bibliográficos
Autores principales: Andreozzi, Emilio, Fratini, Antonio, Esposito, Daniele, Cesarelli, Mario, Bifulco, Paolo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8028787/
https://www.ncbi.nlm.nih.gov/pubmed/33827586
http://dx.doi.org/10.1186/s12938-021-00874-8
_version_ 1783676006357270528
author Andreozzi, Emilio
Fratini, Antonio
Esposito, Daniele
Cesarelli, Mario
Bifulco, Paolo
author_facet Andreozzi, Emilio
Fratini, Antonio
Esposito, Daniele
Cesarelli, Mario
Bifulco, Paolo
author_sort Andreozzi, Emilio
collection PubMed
description BACKGROUND: Low-dose X-ray images have become increasingly popular in the last decades, due to the need to guarantee the lowest reasonable patient’s exposure. Dose reduction causes a substantial increase of quantum noise, which needs to be suitably suppressed. In particular, real-time denoising is required to support common interventional fluoroscopy procedures. The knowledge of noise statistics provides precious information that helps to improve denoising performances, thus making noise estimation a crucial task for effective denoising strategies. Noise statistics depend on different factors, but are mainly influenced by the X-ray tube settings, which may vary even within the same procedure. This complicates real-time denoising, because noise estimation should be repeated after any changes in tube settings, which would be hardly feasible in practice. This work investigates the feasibility of an a priori characterization of noise for a single fluoroscopic device, which would obviate the need for inferring noise statics prior to each new images acquisition. The noise estimation algorithm used in this study was tested in silico to assess its accuracy and reliability. Then, real sequences were acquired by imaging two different X-ray phantoms via a commercial fluoroscopic device at various X-ray tube settings. Finally, noise estimation was performed to assess the matching of noise statistics inferred from two different sequences, acquired independently in the same operating conditions. RESULTS: The noise estimation algorithm proved capable of retrieving noise statistics, regardless of the particular imaged scene, also achieving good results even by using only 10 frames (mean percentage error lower than 2%). The tests performed on the real fluoroscopic sequences confirmed that the estimated noise statistics are independent of the particular informational content of the scene from which they have been inferred, as they turned out to be consistent in sequences of the two different phantoms acquired independently with the same X-ray tube settings. CONCLUSIONS: The encouraging results suggest that an a priori characterization of noise for a single fluoroscopic device is feasible and could improve the actual implementation of real-time denoising strategies that take advantage of noise statistics to improve the trade-off between noise reduction and details preservation. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12938-021-00874-8.
format Online
Article
Text
id pubmed-8028787
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-80287872021-04-08 Toward a priori noise characterization for real-time edge-aware denoising in fluoroscopic devices Andreozzi, Emilio Fratini, Antonio Esposito, Daniele Cesarelli, Mario Bifulco, Paolo Biomed Eng Online Research BACKGROUND: Low-dose X-ray images have become increasingly popular in the last decades, due to the need to guarantee the lowest reasonable patient’s exposure. Dose reduction causes a substantial increase of quantum noise, which needs to be suitably suppressed. In particular, real-time denoising is required to support common interventional fluoroscopy procedures. The knowledge of noise statistics provides precious information that helps to improve denoising performances, thus making noise estimation a crucial task for effective denoising strategies. Noise statistics depend on different factors, but are mainly influenced by the X-ray tube settings, which may vary even within the same procedure. This complicates real-time denoising, because noise estimation should be repeated after any changes in tube settings, which would be hardly feasible in practice. This work investigates the feasibility of an a priori characterization of noise for a single fluoroscopic device, which would obviate the need for inferring noise statics prior to each new images acquisition. The noise estimation algorithm used in this study was tested in silico to assess its accuracy and reliability. Then, real sequences were acquired by imaging two different X-ray phantoms via a commercial fluoroscopic device at various X-ray tube settings. Finally, noise estimation was performed to assess the matching of noise statistics inferred from two different sequences, acquired independently in the same operating conditions. RESULTS: The noise estimation algorithm proved capable of retrieving noise statistics, regardless of the particular imaged scene, also achieving good results even by using only 10 frames (mean percentage error lower than 2%). The tests performed on the real fluoroscopic sequences confirmed that the estimated noise statistics are independent of the particular informational content of the scene from which they have been inferred, as they turned out to be consistent in sequences of the two different phantoms acquired independently with the same X-ray tube settings. CONCLUSIONS: The encouraging results suggest that an a priori characterization of noise for a single fluoroscopic device is feasible and could improve the actual implementation of real-time denoising strategies that take advantage of noise statistics to improve the trade-off between noise reduction and details preservation. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12938-021-00874-8. BioMed Central 2021-04-07 /pmc/articles/PMC8028787/ /pubmed/33827586 http://dx.doi.org/10.1186/s12938-021-00874-8 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research
Andreozzi, Emilio
Fratini, Antonio
Esposito, Daniele
Cesarelli, Mario
Bifulco, Paolo
Toward a priori noise characterization for real-time edge-aware denoising in fluoroscopic devices
title Toward a priori noise characterization for real-time edge-aware denoising in fluoroscopic devices
title_full Toward a priori noise characterization for real-time edge-aware denoising in fluoroscopic devices
title_fullStr Toward a priori noise characterization for real-time edge-aware denoising in fluoroscopic devices
title_full_unstemmed Toward a priori noise characterization for real-time edge-aware denoising in fluoroscopic devices
title_short Toward a priori noise characterization for real-time edge-aware denoising in fluoroscopic devices
title_sort toward a priori noise characterization for real-time edge-aware denoising in fluoroscopic devices
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8028787/
https://www.ncbi.nlm.nih.gov/pubmed/33827586
http://dx.doi.org/10.1186/s12938-021-00874-8
work_keys_str_mv AT andreozziemilio towardapriorinoisecharacterizationforrealtimeedgeawaredenoisinginfluoroscopicdevices
AT fratiniantonio towardapriorinoisecharacterizationforrealtimeedgeawaredenoisinginfluoroscopicdevices
AT espositodaniele towardapriorinoisecharacterizationforrealtimeedgeawaredenoisinginfluoroscopicdevices
AT cesarellimario towardapriorinoisecharacterizationforrealtimeedgeawaredenoisinginfluoroscopicdevices
AT bifulcopaolo towardapriorinoisecharacterizationforrealtimeedgeawaredenoisinginfluoroscopicdevices